J Cardiovasc Electrophysiol ; 32(1): 67-70, 2021 01.
Article in English | MEDLINE | ID: mdl-33155334

ABSTRACT

Deviation of the esophagus during atrial fibrillation (AF) ablation can reduce esophageal injury. This study reports upon a novel esophageal retractor that utilizes vacuum suction and mechanical deflection to deviate the esophagus. The device was used in seven patients undergoing cryoballoon AF ablation. The esophagus was deviated 31.9 ± 4.4 mm to the right and 28.2 ± 5.9 mm to the left. Endoscopy at 4.4 ± 1.5 days postablation showed no esophageal injury. This study demonstrates the safe and effective deviation of the esophagus without a trailing edge with an esophageal retractor utilizing vacuum suction and mechanical deflection.


Subject(s)
Atrial Fibrillation , Catheter Ablation , Atrial Fibrillation/diagnosis , Atrial Fibrillation/surgery , Catheter Ablation/adverse effects , Esophagus/diagnostic imaging , Esophagus/surgery , Humans , Suction , Vacuum