Radiat Prot Dosimetry ; 188(1): 117-122, 2020 Jun 12.
Article in English | MEDLINE | ID: mdl-31747040

ABSTRACT

A current-mode neutron detector with a pair of 6Li- and 7Li-glass scintillators has been developed to measure high-flux neutrons in a boron neutron capture therapy field. Neutrons are basically measured by subtracting gamma-ray component using current outputs from the 7Li-glass scintillator. In the present study, the difference in the gamma-ray sensitivity between the 6Li- and 7Li-glass scintillators and the neutron sensitivity for the 7Li-glass scintillator due to the 6Li contamination were also considered to improve the gamma-ray subtraction precision. The gamma-ray subtraction procedure was experimentally investigated in thermal neutron fields with 252Cf and 241Am-Be neutron sources, which have different gamma-ray intensities per unit neutron fluence. A linear relation between neutron fluence and current output was obtained for the neutron detector in the two types of thermal neutron fields with different gamma-ray intensities. It was found that the gamma-ray subtraction procedure is useful for current-mode neutron detectors.


Subject(s)
Boron Neutron Capture Therapy , Glass , Isotopes , Lithium , Scintillation Counting/instrumentation , Americium , Californium , Equipment Design , Gamma Rays , Humans , Neutrons
SELECTION OF CITATIONS