Vasc Endovascular Surg ; 53(7): 529-535, 2019 Oct.
Article in English | MEDLINE | ID: mdl-31230589

ABSTRACT

OBJECTIVE: We developed a duplex ultrasound simulator and used it to assess accuracy of volume flow measurements in dialysis access fistula (DAF) models. METHODS: The simulator consists of a mannequin, computer, and mock transducer. Each case is built from a patient's B-mode images that are used to create a 3-dimensional surface model of the DAF. Computational fluid dynamics is used to determine blood flow velocities based on model vessel geometry. The simulator displays real-time B-mode and color-flow images, and Doppler spectral waveforms are generated according to user-defined settings. Accuracy was assessed by scanning each case and measuring volume flow in the inflow artery and outflow vein for comparison with true volume flow values. RESULTS: Four examiners made 96 volume flow measurements on four DAF models. Measured volume flow deviated from the true value by 35 ± 36%. Mean absolute deviation from true volume flow was lower for arteries than veins (22 ± 19%, N = 48 vs. 58 ± 33%, N = 48, p < 0.0001). This finding is attributed to eccentricity of outflow veins which resulted in underestimating true cross-sectional area. Regression analysis indicated that error in measuring cross-sectional area was a predictor of error in volume flow measurement (ß = 0.948, p < 0.001). Volume flow error was reduced from 35 ± 36% to 9 ± 8% (p < 0.000001) by calculating vessel area as an ellipse. CONCLUSIONS: Duplex volume flow measurements are based on a circular vessel shape. DAF inflow arteries are circular, but outflow veins can be elliptical. Simulation-based analysis showed that error in measuring volume flow is mainly due to assumption of a circular vessel.


Subject(s)
Arteriovenous Shunt, Surgical , Computer Simulation , Manikins , Models, Cardiovascular , Renal Dialysis , Upper Extremity/blood supply , Blood Flow Velocity , Humans , Hydrodynamics , Observer Variation , Predictive Value of Tests , Regional Blood Flow , Reproducibility of Results , Transducers , Ultrasonography, Doppler, Color/instrumentation
2.
J Clin Neurosci ; 61: 153-159, 2019 Mar.
Article in English | MEDLINE | ID: mdl-30470652

ABSTRACT

Perianeurysmal hemodynamics play a vital role in the initiation, growth and rupture of intracranial aneurysms. In vitro investigations of aneurysmal hemodynamics are helpful to visualize and measure blood flow, and aiding surgical planning approaches. Improving in vitro model creation can improve the feasibility and accuracy of hemodynamic investigations and surgical planning, improving clinical value. In this study, in vitro models were created from three-dimensional rotational angiography (3DRA) of six patients harboring intracranial aneurysms using a multi-step process involving 3D printing, index of refraction matching and silicone casting that renders the models transparent for flow visualization. Each model was treated with the same commercially-available, patient-specific, endovascular devices (coils and/or stents). All models were scanned by synchrotron X-ray microtomography to obtain high-resolution imaging of the vessel lumen, aneurysmal sac and endovascular devices. Dimensional accuracy was compared by quantifying the differences between the microtomographic reconstructions of the fabricated phantoms and the original 3DRA obtained during patient treatment. True-scale in vitro flow phantoms were successfully created for all six patients. Optical transparency was verified by using an index of refraction matched working fluid that replicated the mechanical behavior of blood. Synchrotron imaging of vessel lumen, aneurysmal sac and endovascular devices was successfully obtained, and dimensional errors were found to be O(100 µm). The creation of dimensionally-accurate, optically-transparent flow phantoms of patient-specific intracranial aneurysms is feasible using 3D printing technology. Such models may enable in vitro investigations of aneurysmal hemodynamics to aid in treatment planning and outcome prediction to devise optimal patient-specific neurointerventional strategies.


Subject(s)
Hemodynamics/physiology , Intracranial Aneurysm/physiopathology , Models, Cardiovascular , Phantoms, Imaging , Printing, Three-Dimensional , Cerebral Angiography/methods , Humans , Intracranial Aneurysm/surgery
3.
Ultrasound Med Biol ; 44(8): 1712-1720, 2018 08.
Article in English | MEDLINE | ID: mdl-29793851

ABSTRACT

We developed a duplex ultrasound simulator for training and assessment of scanning skills. We used the simulator to test examiner performance in the measurement of flow velocities in dialysis access fistulas. Test cases were created from 3-D ultrasound scans of two dialysis access fistulas by reconstructing 3-D blood vessel models and simulating blood flow velocity fields within the lumens. The simulator displays a 2-D B-mode or color Doppler image corresponding to transducer position on a mannequin; a spectral waveform is generated according to Doppler sample volume location and system settings. Examiner performance was assessed by comparing the measured peak systolic velocity (PSV) with the true PSV provided by the computational flow model. The PSV measured by four expert examiners deviated from the true value by 7.8 ± 6.1%. The results indicate the ability of the simulator to objectively assess an examiner's measurement accuracy in complex vascular targets.


Subject(s)
Arteriovenous Shunt, Surgical , Clinical Competence/statistics & numerical data , Computer Simulation , Kidney/diagnostic imaging , Renal Dialysis/instrumentation , Ultrasonography, Doppler, Duplex/methods , Blood Flow Velocity , Graft Occlusion, Vascular/diagnostic imaging , Imaging, Three-Dimensional/methods , Kidney/blood supply , Manikins , Models, Biological , Regional Blood Flow , Reproducibility of Results , Ultrasonics/education
4.
Vasc Endovascular Surg ; 50(5): 309-16, 2016 Jul.
Article in English | MEDLINE | ID: mdl-27206747

ABSTRACT

OBJECTIVE: Duplex ultrasound scanning with B-mode imaging and both color Doppler and Doppler spectral waveforms is relied upon for diagnosis of vascular pathology and selection of patients for further evaluation and treatment. In most duplex ultrasound applications, classification of disease severity is based primarily on alterations in blood flow velocities, particularly the peak systolic velocity (PSV) obtained from Doppler spectral waveforms. We developed a duplex ultrasound simulator for training and assessment of scanning skills. METHODS: Duplex ultrasound cases were prepared from 2-dimensional (2D) images of normal and stenotic carotid arteries by reconstructing the common carotid, internal carotid, and external carotid arteries in 3 dimensions and computationally simulating blood flow velocity fields within the lumen. The simulator displays a 2D B-mode image corresponding to transducer position on a mannequin, overlaid by color coding of velocity data. A spectral waveform is generated according to examiner-defined settings (depth and size of the Doppler sample volume, beam steering, Doppler beam angle, and pulse repetition frequency or scale). The accuracy of the simulator was assessed by comparing the PSV measured from the spectral waveforms with the true PSV which was derived from the computational flow model based on the size and location of the sample volume within the artery. RESULTS: Three expert examiners made a total of 36 carotid artery PSV measurements based on the simulated cases. The PSV measured by the examiners deviated from true PSV by 8% ± 5% (N = 36). The deviation in PSV did not differ significantly between artery segments, normal and stenotic arteries, or examiners. CONCLUSION: To our knowledge, this is the first simulation of duplex ultrasound that can create and display real-time color Doppler images and Doppler spectral waveforms. The results demonstrate that an examiner can measure PSV from the spectral waveforms using the settings on the simulator with a mean absolute error in the velocity measurement of less than 10%. With the addition of cases with a range of pathologies, this duplex ultrasound simulator will be a useful tool for training health-care providers in vascular ultrasound applications and for assessing their skills in an objective and quantitative manner.


Subject(s)
Cardiology/education , Carotid Artery, External/diagnostic imaging , Carotid Artery, Internal/diagnostic imaging , Carotid Stenosis/diagnostic imaging , Computer Simulation , Computer-Assisted Instruction , Education, Medical/methods , Manikins , Ultrasonography, Doppler, Duplex , Blood Flow Velocity , Carotid Artery, External/physiopathology , Carotid Artery, Internal/physiopathology , Carotid Stenosis/physiopathology , Case-Control Studies , Clinical Competence , Humans , Learning Curve , Observer Variation , Predictive Value of Tests , Regional Blood Flow , Reproducibility of Results