J Res Natl Inst Stand Technol ; 100(3): 227-239, 1995.
Article in English | MEDLINE | ID: mdl-29151737

ABSTRACT

A bilateral intercomparison of photometric units between NIST, USA and PTB, Germany has been conducted to update the knowledge of the relationship between the photometric units disseminated in each country. The luminous intensity unit (cd) and the luminous flux unit (lm) maintained at both laboratories are compared by circulating transfer standard lamps. Also, the photometric responsivity sv is compared by circulating a V(λ)-corrected detector with a built-in current-to-voltage converter. The results show that the difference of luminous intensity unit between NIST and PTB, (PTB-NIST)/NIST, is 0.2 % with a relative expanded uncertainty (coverage factor k = 2) of 0.24 %. The difference is reduced significantly from that at the 1985 CCPR intercomparison (0.9 %). The difference in luminous flux unit, (PTB - NIST)/NIST, is found to be 1.5 % with a relative expanded uncertainty (coverage factor k =2) of 0.15 %. The difference remained nearly the same as that at the 1985 intercomparison (1.6 %). These results agree with what is predicted from the history of maintaining the units at each laboratory.