Anal Chim Acta ; 839: 26-33, 2014 Aug 11.
Article in English | MEDLINE | ID: mdl-25066715

ABSTRACT

Nickel hydroxide nanoparticles were successfully electrodeposited on graphite electrode (Gr/NiONP) and employed as a robust non-enzymatic glucose sensor. The results of cyclic voltammetry (CV) and chronoamperometry demonstrated that the Gr/NiONP electrode displayed high electrocatalytic activity toward glucose. The oxidation current is directly related to the glucose concentration from 1 µM to 15 mM. Besides, the glucose sensor displayed high sensitivity (2400 µA mM(-1) cm(-2)) with a detection limit of 0.53 µM (S/N=3) in basic solution. Moreover, the sensor showed excellent selectivity, reproducibility and stability properties. The relative standard deviation is 1.2% for 10 successive measurements in 16 µM glucose. Interestingly, the signal for glucose was maintained at 95% of its initial value even after 6 months of storage under ambient conditions. Gr/NiONP electrode has also been tested to detect glucose in human serum with satisfactory results.


Subject(s)
Biosensing Techniques , Glucose/analysis , Hydroxides/chemistry , Nanoparticles , Nickel/chemistry , Microscopy, Electron, Scanning , Oxidation-Reduction , Reproducibility of Results , X-Ray Diffraction