Am Rev Respir Dis ; 148(2): 358-64, 1993 Aug.
Article in English | MEDLINE | ID: mdl-8342899

ABSTRACT

We studied the response time (RT) and reliability of three neonatal patient-triggered ventilator (PTV) systems: the Draeger Babylog 8000, the Bear Cub enhancement module (CEM), and the Infrasonics Star Sync. In 10 adult rabbits, airway flow and pressure recordings showed the RT of the Star Sync to be shorter than that of the Bear CEM (53 +/- 13 versus 65 +/- 15 ms, p < 0.05), and both were shorter than that of the Babylog (95 +/- 24 ms, p < 0.01) by ANOVA. The RT of the Bear CEM and the Babylog increased significantly at decreased trigger sensitivity settings. All ventilators triggered successfully on assist-control (A/C). However, the Babylog had a higher rate of asynchrony on SIMV (30 +/- 25%) than the Bear CEM (1.1 +/- 0.3%) and the Star Sync (1.2 +/- 0.4%), p < 0.01. In 10 infants with respiratory failure, recordings of airway flow and pressure were made at ventilator inspiratory time (Ti) settings of 0.3, 0.4, and 0.5 s on assist-control and on SIMV at rates of 15, 30, 45, and 60 breaths/min. The Star Sync and Bear CEM triggered successfully on A/C (100%) and had low rates of asynchrony on SIMV (1 to 3%). The Babylog had a lower success rate on A/C (70 +/- 12%) and a higher rate of asynchrony on SIMV (29 +/- 30%) than the other two ventilators; p < 0.01. The lower reliability of the Babylog was due to its variable refractory period (0.2 to 0.5 s, to equal the set Ti).(ABSTRACT TRUNCATED AT 250 WORDS)


Subject(s)
Respiration, Artificial , Ventilators, Mechanical , Animals , Equipment Design , Equipment Safety , Humans , Infant, Newborn , Inhalation/physiology , Intermittent Positive-Pressure Ventilation , Intubation, Intratracheal/instrumentation , Positive-Pressure Respiration , Pressure , Rabbits , Respiration/physiology , Respiration, Artificial/methods , Respiratory Insufficiency/physiopathology , Respiratory Insufficiency/therapy , Tidal Volume/physiology , Time Factors
SELECTION OF CITATIONS
SEARCH DETAIL
...