Z Med Phys ; 26(3): 238-50, 2016 Sep.
Article in English | MEDLINE | ID: mdl-26387927

ABSTRACT

The application of various radiation detectors for brachytherapy dosimetry has motivated this study of the energy dependence of radiation quality correction factor kQ,M, the quotient of the detector responses under calibration conditions at a (60)Co unit and under the given non-reference conditions at the point of measurement, M, occurring in photon brachytherapy. The investigated detectors comprise TLD, radiochromic film, ESR, Si diode, plastic scintillator and diamond crystal detectors as well as ionization chambers of various sizes, whose measured response-energy relationships, taken from the literature, served as input data. Brachytherapy photon fields were Monte-Carlo simulated for an ideal isotropic (192)Ir point source, a model spherical (192)Ir source with steel encapsulation and a commercial HDR GammaMed Plus source. The radial source distance was varied within cylindrical water phantoms with outer radii ranging from 10 to 30cm and heights from 20 to 60cm. By application of this semiempirical method - originally developed for teletherapy dosimetry - it has been shown that factor kQ,M is closely correlated with a single variable, the fluence-weighted mean photon energy EF at the point of measurement. The radial profiles of EF obtained with either the commercial (192)Ir source or the two simplified source variants show little variation. The observed correlations between parameters kQ,M and EF are represented by fitting formulae for all investigated detectors, and further variation of the detector type is foreseen. The herewith established close correlation of radiation quality correction factor kQ,M with local mean photon energy EF can be regarded as a simple regularity, facilitating the practical application of correction factor kQ,M for in-phantom dosimetry around (192)Ir brachytherapy sources. EF values can be assessed by Monte Carlo simulation or measurement. A technique describing the local measurement of EF will be published separately.


Subject(s)
Artifacts , Brachytherapy/instrumentation , Iridium Radioisotopes/therapeutic use , Quality Assurance, Health Care/methods , Radiometry/instrumentation , Radiometry/methods , Algorithms , Brachytherapy/methods , Computer Simulation , Equipment Design , Equipment Failure Analysis , Iridium Radioisotopes/analysis , Linear Energy Transfer , Models, Statistical , Monte Carlo Method , Photons , Radiopharmaceuticals/analysis , Radiopharmaceuticals/therapeutic use , Reproducibility of Results , Sensitivity and Specificity
2.
Med Phys ; 42(7): 4285-94, 2015 Jul.
Article in English | MEDLINE | ID: mdl-26133626

ABSTRACT

PURPOSE: The aim of this study was to investigate whether a chamber-type-specific radiation quality correction factor kQ can be determined in order to measure the reference air kerma rate of (60)Co high-dose-rate (HDR) brachytherapy sources with acceptable uncertainty by means of a well-type ionization chamber calibrated for (192)Ir HDR sources. METHODS: The calibration coefficients of 35 well-type ionization chambers of two different chamber types for radiation fields of (60)Co and (192)Ir HDR brachytherapy sources were determined experimentally. A radiation quality correction factor kQ was determined as the ratio of the calibration coefficients for (60)Co and (192)Ir. The dependence on chamber-to-chamber variations, source-to-source variations, and source strength was investigated. RESULTS: For the PTW Tx33004 (Nucletron source dosimetry system (SDS)) well-type chamber, the type-specific radiation quality correction factor kQ is 1.19. Note that this value is valid for chambers with the serial number, SN ≥ 315 (Nucletron SDS SN ≥ 548) onward only. For the Standard Imaging HDR 1000 Plus well-type chambers, the type-specific correction factor kQ is 1.05. Both kQ values are independent of the source strengths in the complete clinically relevant range. The relative expanded uncertainty (k = 2) of kQ is UkQ = 2.1% for both chamber types. CONCLUSIONS: The calibration coefficient of a well-type chamber for radiation fields of (60)Co HDR brachytherapy sources can be calculated from a given calibration coefficient for (192)Ir radiation by using a chamber-type-specific radiation quality correction factor kQ. However, the uncertainty of a (60)Co calibration coefficient calculated via kQ is at least twice as large as that for a direct calibration with a (60)Co source.


Subject(s)
Brachytherapy/instrumentation , Brachytherapy/methods , Cobalt Radioisotopes/therapeutic use , Calibration , Iridium Radioisotopes/therapeutic use , Radiometry , Uncertainty
3.
Radiother Oncol ; 100(3): 442-5, 2011 Sep.
Article in English | MEDLINE | ID: mdl-21924788

ABSTRACT

A large air-filled parallel-plate extrapolation chamber in a phantom of water-equivalent material is used as a primary standard measuring device for low-energy interstitial brachytherapy sources from which the unit of absorbed dose to water can be derived. The chamber is suitable for low-energy photons with energies up to 50 keV. The method to determine the absorbed dose to water was newly developed based on radiation transport theory. It offers a clear analytical expression to determine D(w). A conversion factor C(x(i),x(i)(+1)) has to be applied to the difference of ionization charges measured at two plate separations x(i) and x(i)(+1). The details of the method are presented. The determination of D(w) of an I-125 seed is demonstrated by the measurement of a 'BEBIG Symmetra I25.S16' - seed. Additional measurements of the reference air kerma rate with the PTB primary standard chamber GROVEX I allow to determine experimentally the dose rate constant of an I-125 seed by means of primary standards for the first time. Good agreement was found between the obtained dose rate constant and the published data.


Subject(s)
Brachytherapy/methods , Iodine Radioisotopes/therapeutic use , Radiometry/instrumentation , Absorption , Humans , Monte Carlo Method , Phantoms, Imaging , Photons , Radiotherapy Dosage , Spectrum Analysis/methods , Water
4.
Phys Med Biol ; 54(9): 2915-31, 2009 May 07.
Article in English | MEDLINE | ID: mdl-19384000

ABSTRACT

A phantom study for dosimetry in the urethra using alanine/ESR during (192)Ir HDR brachytherapy of prostate cancer is presented. The measurement method of the secondary standard of the Physikalisch-Technische Bundesanstalt had to be slightly modified in order to be able to measure inside a Foley catheter. The absorbed dose to water response of the alanine dosimetry system to (192)Ir was determined with a reproducibility of 1.8% relative to (60)Co. The resulting uncertainty for measurements inside the urethra was estimated to be 3.6%, excluding the uncertainty of the dose rate constant Lambda. The applied dose calculated by a treatment planning system is compared to the measured dose for a small series of (192)Ir HDR irradiations in a gel phantom. The differences between the measured and applied dose are well within the limits of uncertainty. Therefore, the method is considered to be suitable for measurements in vivo.


Subject(s)
Alanine , Iridium Radioisotopes/therapeutic use , Phantoms, Imaging , Prostatic Neoplasms/radiotherapy , Radiation Dosage , Radiometry/instrumentation , Urethra/radiation effects , Brachytherapy , Electron Spin Resonance Spectroscopy , Humans , Male , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Uncertainty