1.
Med Phys ; 29(4): 538-43, 2002 Apr.
Article in English | MEDLINE | ID: mdl-11991125

ABSTRACT

The low dose rate CDC-type miniature cylindrical 137Cs sources are available, with one or three active beads, for use in source trains in automatic and manual afterloading systems for gynecological brachytherapy. Absolute dose rate distributions in water have been calculated around these sources using the Monte Carlo GEANT3 code and they are presented as conventional two-dimensional Cartesian lookup tables. The AAPM Task Group 43 formalism for dose calculation has been also applied. The dose rate constant obtained for the one bead source is lambda = 1.113 +/- 0.003cGyh(-1) U(-1), and the value for the three bead source is A= 1.103 +/- 0.003cGyh(-1) U(-1). Finally, for the treatment planning systems based on Sievert-type algorithms, the attenuation coefficients that best reproduce Monte Carlo dose rate distribution are given.


Subject(s)
Brachytherapy/instrumentation , Brachytherapy/methods , Radiometry/methods , Air , Algorithms , Monte Carlo Method , Radiometry/standards , Radiotherapy Planning, Computer-Assisted
2.
Phys Med Biol ; 46(7): 2029-40, 2001 Jul.
Article in English | MEDLINE | ID: mdl-11474942

ABSTRACT

Basic dosimetric data for the Walstam CDC.K-type low dose rate 137Cs sources in water have been calculated using Monte Carlo techniques. These sources, CDC.K1 -K3 and CDC.K4, are widely used in a range of applicators and moulds for the treatment of intracavitary and superficial cancers. Our purpose is to improve existing data about these sources using the Monte Carlo simulation code GEANT3. Absolute dose rate distributions in water have been calculated around these sources and are presented as conventional 2D Cartesian look-up tables. Also the AAPM Task Group 43 formalism for dose calculation has been applied. The calculated dose rate constant for the CDC.K1-K3 source is A = 1.106 +/- 0.001 cGy h(-1) U(-1), and for the CDC.K4 source, A = 1.092 +/- 0.001 cGy h(-1) U(-1). The anisotropy of the sources are accurately studied and F(r, theta) tables are given. Also phi an(r) factors are presented. The radial dose functions are given as a polynomial fit to the calculated data up to 15 cm. Best-fit values of coefficients suitable for use in Sievert integral calculations have been derived.


Subject(s)
Cesium Radioisotopes , Radiotherapy Planning, Computer-Assisted/methods , Algorithms , Anisotropy , Humans , Models, Statistical , Monte Carlo Method , Radiometry , Software
3.
Phys Med Biol ; 46(3): N79-90, 2001 Mar.
Article in English | MEDLINE | ID: mdl-11277236

ABSTRACT

In this study a complete set of dosimetric data is presented for the high dose rate (HDR) source from Amersham used in the Buchler remote afterloading HDR unit. These data have been calculated by means of the Monte Carlo simulation code GEANT taking into account the detailed geometry of the source. Absolute dose rate distributions in water were calculated around this source and are presented as conventional 2D Cartesian look-up tables. All dosimetric quantities recommended by the AAPM Task Group 43 report have been calculated. Quantities determined are: dose rate constant, radial dose function, anisotropy function, anisotropy factor and anisotropy constant. The dose rate distributions of the Buchler HDR source are compared with those of other HDR sources used in brachytherapy, showing that the differences are large in zones near the long source axis due to oblique filtration. These Monte Carlo simulated data in water can be used for clinical applications.


Subject(s)
Brachytherapy/instrumentation , Brachytherapy/methods , Iridium Radioisotopes , Anisotropy , Equipment Design , Monte Carlo Method , Radiotherapy Dosage
4.
Med Phys ; 28(12): 2576-85, 2001 Dec.
Article in English | MEDLINE | ID: mdl-11797963

ABSTRACT

In this study a complete set of dosimetric data for the Plus and 12i Gammamed PDR (pulsed dose rate) 192Ir sources is presented. These data have been calculated using the Monte Carlo simulation code GEANT3. Absolute dose rate distributions in water around these sources were calculated and are presented in form of conventional two dimensional (2D) Cartesian look-up tables. All dosimetric quantities recommended by the AAPM Task Group 43 report have been also calculated. These quantities are dose rate constant, radial dose function, anisotropy function and anisotropy factor. The dose rate distribution of the 12i source was compared with the corresponding data for the microselectron PDR source showing large differences between both sources.


Subject(s)
Iridium Radioisotopes/therapeutic use , Radiometry/methods , Air , Anisotropy , Monte Carlo Method , Particle Accelerators , Water
5.
Med Phys ; 28(12): 2586-91, 2001 Dec.
Article in English | MEDLINE | ID: mdl-11797964

ABSTRACT

In this study a complete set of dosimetric data for the GammaMed high dose rate (HDR) 12i and Plus 192Ir sources are presented. These data have been calculated by means of the Monte Carlo simulation code GEANT3. Absolute dose rate distributions in water are presented as conventional two dimensional (2D) Cartesian look-up tables, and in the TG43 formalism.


Subject(s)
Brachytherapy/instrumentation , Iridium Radioisotopes/therapeutic use , Radiometry/methods , Anisotropy , Biophysical Phenomena , Biophysics , Monte Carlo Method , Water
SELECTION OF CITATIONS