Rev Sci Instrum ; 87(7): 073503, 2016 Jul.
Article in English | MEDLINE | ID: mdl-27475555

ABSTRACT

An Hα camera has been designed and installed in the U-3M torsatron for spatially and temporally resolved measurements. This device provides fast measurements of the emission brightness profile in the noisy environment of the radio frequency (RF) heated plasma. Unusual topology of diagnostics and the data acquisition system are applied. All the system components, including digitizers, are assembled in a single unit. It allows the suppression of a low-frequency electromagnetic interference by eliminating the ground loops. And the suppression of RF noises is achieved by eliminating the signal interface cables and digital interface cables in the design. The Wi-Fi interface is used to prevent a ground loop in the data transfer stage. The achieved sensitivity of our diagnostics is high enough for measuring the Hα emission from the low-density (ne ≈ (1-2)⋅10(10) cm(-3)) plasma with a temporal resolution of about 20 µs in the noisy environment. Different types of Hα emission fluctuations within the frequency range of 1-5 kHz and poloidal mode numbers m = 0 and m = 5 have been observed in U-3M. A simple technique of the line-of-sight data analysis, based on the U-3M magnetic surface asymmetry, is proposed and used for the spatial localization of the rotating mode and for the determination of mode numbers and its poloidal rotation direction using a single Hα array.