ABSTRACT

With the advancements in electronics technology, high-density (HD) EMG sensing systems have become available and have been investigated for their feasibility and performance in neural-machine interface (NMI) applications. Comparing to the traditional single channel-based targeted muscle sensing method, HD EMG sensing performs a sampling of the electrical activity over a larger surface area and has the promise of 1) providing richer neural information from one temporal and two spatial dimensions and 2) ease of wear in real life without the need of anatomically targeted electrode placement. To use HD EMG in real-time NMI applications, challenges including high computational burden and unreliability of EMG recordings over time need to be addressed. This paper presented an HD EMG PR based NMI which seamlessly integrates HD EMG PR with a Sensor Fault-Tolerant Module (SFTM) which aimed to provide robust PR in real time. Experimental results showed that the SFTM was able to recover the PR accuracies by 6%-22% from disturbances including contact artifacts and loose contacts. A Python-based implementation of the proposed HD EMG SFTM was developed and was demonstrated to be computationally efficient for real-time performance. These results have demonstrated the feasibility of a robust real-time HD EMG PR-based NMI.


Subject(s)
Algorithms , Artifacts , Electrodes , Electromyography
SELECTION OF CITATIONS