Radiol Phys Technol ; 6(2): 349-55, 2013 Jul.
Article in English | MEDLINE | ID: mdl-23605696

ABSTRACT

In interventional radiology, dose estimation using the interventional reference point (IRP) is a practical method for obtaining the real-time skin dose of a patient. However, the IRP is defined in terms of adult cardiovascular radiology and is not suitable for dosimetry of the head. In the present study, we defined a new reference point (neuro-IRP) for neuro-interventional procedures. The neuro-IRP was located on the central ray of the X-ray beam, 9 cm from the isocenter, toward the focal spot. To verify whether the neuro-IRP was accurate in dose estimation, we compared calculated doses at the neuro-IRP and actual measured doses at the surface of the head phantom for various directions of the X-ray projection. The resulting calculated doses were fairly consistent with actual measured doses, with the error in this estimation within approximately 15%. These data suggest that dose estimation using the neuro-IRP for the head is valid.


Subject(s)
Cerebrovascular Circulation , Head/diagnostic imaging , Radiology, Interventional/standards , Reference Standards , Safety Management , Tomography, X-Ray Computed/methods , Adult , Aged , Aged, 80 and over , Female , Humans , Male , Middle Aged , Radiation Dosage , Radiation Monitoring , Young Adult