Radiol Phys Technol ; 6(1): 157-61, 2013 Jan.
Article in English | MEDLINE | ID: mdl-23086629

ABSTRACT

We investigated the neonatal entrance-surface dose (ESD) and doses of scattered radiation emitted by a digital mobile X-ray system. The system is equipped with a novel flat-panel detector and is used in the neonatal intensive care unit. In the present study, the following three experiments were performed on frequently used X-ray-imaging condition: (1) the digital characteristics of the FPD were evaluated; (2) the ESD to a water-equivalent phantom was measured with a patient skin dosimeter (PSD); and (3) the scattered radiation around the incubator was measured with an ionization chamber survey meter. The digital characteristic curve showed that the system had excellent linearity and that the contrast characteristics were not affected by the tube voltage in the range of 50-110 kV. The ESD was 51-52 µGy with an 8-cm-thick phantom and 33-34 µGy with a 4.5-cm phantom, for one exposure. The doses measured around the incubator were 0.1-0.6 µSv or below measurable limits. Use of the new device demonstrates the potential of reducing the ESD to the patient and operator.


Subject(s)
Radiation Dosage , Radiation Protection/instrumentation , Radiographic Image Enhancement/instrumentation , Tomography, X-Ray Computed/instrumentation , Feasibility Studies , Humans , Infant, Newborn , Occupational Exposure/prevention & control , Phantoms, Imaging , Radiography, Thoracic , Scattering, Radiation , Tomography, X-Ray Computed/adverse effects
SELECTION OF CITATIONS
SEARCH DETAIL
...