Sensors (Basel) ; 13(11): 15846-60, 2013 Nov 21.
Article in English | MEDLINE | ID: mdl-24284775

ABSTRACT

Spirometry is regarded as the only effective method for detecting pulmonary function test (PFT) indices. In this study, a novel impedance pulmonary function measurement system (IPFS) is developed for directly assessing PFT indices. IPFS can obtain high resolution values and remove motion artifacts through real-time base impedance feedback. Feedback enables the detection of PFT indices using only both hands for convenience. IPFS showed no differences in the sitting, supine, and standing postures during the measurements, indicating that patient posture has no effect on IPFS. Mean distance analysis showed good agreement between the volume and flow signal of IPFS (p < 0.05). PFT indices were detected in subjects to differentiate a chronic obstructive pulmonary disease (COPD) patient group from a normal group. The forced vital capacity (FVC), forced expiratory volume in the first second (FEV1), FEV1/FVC, and peak expiratory flow (PEF) in the COPD group were lower than those in the normal group by IPFS (p < 0.05). IPFS is therefore suitable for evaluating pulmonary function in normal and COPD patients. Moreover, IPFS could be useful for periodic monitoring of existing patients diagnosed with obstructive lung disease.


Subject(s)
Pulmonary Disease, Chronic Obstructive/diagnosis , Respiratory Function Tests/methods , Electric Impedance , Forced Expiratory Volume , Humans , Spirometry