1.
Radiat Prot Dosimetry ; 147(3): 423-8, 2011 Nov.
Article in English | MEDLINE | ID: mdl-21147790

ABSTRACT

Exposure of children to ionising radiation is considered to carry higher risk than that of adults; therefore a need to suggest diagnostic reference levels (DRLs) for the common paediatric diagnostic X-ray procedures was recognised for the X-ray machines meeting the requirements of the recently implemented Safety Code for Medical Diagnostic X-ray Equipment and Installations in India. Measurements were carried out for entrance surface air kerma (free in air) in conventional paediatric X-ray diagnostic examinations among four age groups: <1, 1-4, 5-9 and 10-15 y. A total of 2240 air kerma measurements at different fixed focus to skin distances were studied for 7 paediatric diagnostic examinations with 11 different projections on 62 X-ray machines installed in 22 selected hospitals in the country. The third quartile values of air kerma per paediatric examination for the age group of 5-9 y were considered as values of paediatric DRLs. The suggested values of DRLs are 0.2 mGy for chest AP/PA, 0.3 mGy for chest LAT, 0.7 mGy for lumber spine AP, 1.3 mGy for lumber spine LAT, 0.3 mGy for thoracic spine AP, 0.6 mGy for thoracic spine LAT, 0.5 mGy for abdomen AP, 0.7 mGy for pelvis AP, 0.6 mGy for skull PA, 0.5 mGy for skull LAT and 0.8 mGy for hip joints AP.


Subject(s)
Physical Examination/standards , Practice Guidelines as Topic , Radiation Dosage , Radiation Protection/statistics & numerical data , Reference Standards , Abdomen/radiation effects , Adolescent , Air , Child , Child, Preschool , Hip Joint/radiation effects , Humans , India , Infant , Pelvis/radiation effects , Radiometry , Reference Values , Relative Biological Effectiveness , Skull/radiation effects , Spine/radiation effects , Thorax/radiation effects , X-Rays
2.
J Med Phys ; 35(4): 229-34, 2010 Oct.
Article in English | MEDLINE | ID: mdl-21170188

ABSTRACT

We conducted a radiological safety and quality assurance (QA) audit of 118 medical X-ray diagnostic machines installed in 45 major hospitals in India. The main objective of the audit was to verify compliance with the regulatory requirements stipulated by the national regulatory body. The audit mainly covered accuracy check of accelerating potential (kVp), linearity of tube current (mA station) and timer, congruence of radiation and optical field, and total filtration; in addition, we also reviewed medical X-ray diagnostic installations with reference to room layout of X-ray machines and conduct of radiological protection survey. A QA kit consisting of a kVp Test-O-Meter (ToM) (Model RAD/FLU-9001), dose Test-O-Meter (ToM) (Model 6001), ionization chamber-based radiation survey meter model Gun Monitor and other standard accessories were used for the required measurements. The important areas where there was noncompliance with the national safety code were: inaccuracy of kVp calibration (23%), lack of congruence of radiation and optical field (23%), nonlinearity of mA station (16%) and timer (9%), improper collimator/diaphragm (19.6%), faulty adjustor knob for alignment of field size (4%), nonavailability of warning light (red light) at the entrance of the X-ray room (29%), and use of mobile protective barriers without lead glass viewing window (14%). The present study on the radiological safety status of diagnostic X-ray installations may be a reasonably good representation of the situation in the country as a whole. The study contributes significantly to the improvement of radiological safety by the way of the steps already taken and by providing a vital feed back to the national regulatory body.