Int J Radiat Oncol Biol Phys ; 14(5): 963-7, 1988 May.
Article in English | MEDLINE | ID: mdl-3360662

ABSTRACT

We have made initial characterization measurements on a miniature radiation dosimeter which can be used for in vivo radiation measurements. It consists of a radiation sensing field effect transistor (RADFET) mounted in a 0.8 mm OD plastic catheter. The RADFET acts as a dosimeter by storing trapped charge proportional to absorbed dose. The stored charge signal can be differentiated to give dose rate. We report on the techniques for mounting, a circuit for dose readout, drift of the readings, linearity of response, temperature and angular dependence, and unpowered operation of the device.


Subject(s)
Radiometry/instrumentation , Radiotherapy Dosage