1.
Opt Express ; 29(14): 22533-22552, 2021 Jul 05.
Article in English | MEDLINE | ID: mdl-34266014

ABSTRACT

We have developed a planar absolute radiometer for room temperature (PARRoT) that will replace the legacy C-series calorimeter as the free-space continuous-wave laser power detector standard at the National Institute of Standards and Technology (NIST). This instrument will lower the combined relative expanded measurement uncertainty (k = 2) from 0.84 % to 0.13 %. PARRoT's performance was validated by comparing its response against a transfer standard silicon trap detector traceable to NIST's primary standard laser optimized cryogenic radiometer (LOCR) and against the C-series calorimeter. On average, these comparisons agreed to better than 0.008 % and 0.05 %, respectively.

2.
J Res Natl Inst Stand Technol ; 126: 126011, 2021.
Article in English | MEDLINE | ID: mdl-38469451

ABSTRACT

To calibrate laser power and energy meters, the National Institute of Standards and Technology (NIST) uses several detector-based realizations of the scale for optical radiant flux; these realizations are appropriate for specific laser power/energy ranges and optical coupling configurations. Calibrations from 1 µW to 2 W are currently based upon calorimeters. Validation by comparisons against other primary representations of the optical watt over the last two decades suggests the instruments operate well within their typical reported uncertainty level of 0.86 % with 95 % confidence. The dominant uncertainty contribution in the instrument is attributable to light scattered by the legacy window, which was not previously recognized. The inherent electro-optical inequivalence in the calorimeter's response was reassessed by thermal modeling to be 0.03 %. The principal contributions to the overall inequivalence were corrected, yielding a shift in scale representation under 0.2 % for typical calibrations. With updates in several uncertainty contributions resulting from this reassessment, the resulting combined expanded uncertainty (k = 2) is 0.84 %, which is essentially unchanged from the previous result provided to calibration customers.

3.
Appl Opt ; 58(2): 257-263, 2019 Jan 10.
Article in English | MEDLINE | ID: mdl-30645302

ABSTRACT

We demonstrate improved manufacturability of spectrally flat detectors for visible to mid-infrared wavelengths by characterizing a carbon nanotube spray coating compatible with lithium tantalate and other thermal sensors. Compared against previous spray coatings, it demonstrated the highest responsivity yet attained due to both higher absorptivity and thermal diffusivity, while also being matured to a commercially available product. It demonstrated spectral nonuniformity from 300 nm to 12 µm less than 1% with uncertainty (k=2) under 0.4%. The spatial nonuniformity of the assembled sensor was less than 0.5% over the central half (4 mm) of the absorber. As with previous developments employing isotropic carbon nanotube coatings, the absorber surface was sufficiently robust to withstand cleaning by compressed air blast and survived repeated vacuum cycling without measurable impact upon responsivity.