1.
Rev Sci Instrum ; 85(1): 014704, 2014 Jan.
Article in English | MEDLINE | ID: mdl-24517795

ABSTRACT

We present a precision, 1-A, digitally interfaced current controller for quantum cascade lasers, with demonstrated temperature coefficients for continuous and 40-kHz full-depth square-wave modulated operation, of 1-2 ppm/ °C and 15 ppm/ °C, respectively. High precision digital to analog converters (DACs) together with an ultra-precision voltage reference produce highly stable, precision voltages, which are selected by a multiplexer (MUX) chip to set output currents via a linear current regulator. The controller is operated in conjunction with a power multiplexing unit, allowing one of three lasers to be driven by the controller, while ensuring protection of controller and all lasers during operation, standby, and switching. Simple ASCII commands sent over a USB connection to a microprocessor located in the current controller operate both the controller (via the DACs and MUX chip) and the power multiplexer.

2.
Analyst ; 139(9): 2047-56, 2014 May 07.
Article in English | MEDLINE | ID: mdl-24384671

ABSTRACT

We present results demonstrating real-time sensing of four different fluorocarbons at low part-per billion (ppb) concentrations using an external cavity quantum cascade laser (ECQCL) designed for infrared vibrational spectroscopy of molecules with broad absorption features. The ECQCL was repeatedly swept at 20 Hz over its full tuning range of 1145-1265 cm(-1) providing a scan rate of 3535 cm(-1) s(-1), and a detailed characterization of the ECQCL scan stability and repeatability is presented. The ECQCL was combined with a 100 meter path length multi-pass cell for direct absorption spectroscopy. A portable sensor system is described, which was deployed on a mobile automotive platform to provide spatially-resolved detection of fluorocarbons in outdoor experiments. Noise-equivalent detection limits of 800-1000 parts-per-trillion (ppt) are demonstrated for 1 s integration times.


Subject(s)
Fluorocarbons/analysis , Gases , Quantum Theory
SELECTION OF CITATIONS
SEARCH DETAIL
...