Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Rofo ; 173(3): 181-6, 2001 Mar.
Article in German | MEDLINE | ID: mdl-11293856

ABSTRACT

PURPOSE: Clinical and radiographic findings are of little value in the diagnosis of hip endoprothesis loosening. Refined contrast and radionuclide arthrographic techniques are more precise in predicting endoprothesis loosening. Here we compare both techniques with respect to the surgical result. MATERIAL AND METHODS: 68 patients with hip pain after various arthroplasty procedure were investigated by clinical examination, X-ray, DSAr and AS. Surgical proof was obtained in 73.5%. RESULTS: An acetabular component loosening was detected in 89% by DSAr and in 86% by AS with a specificity for both methods of 93%. The femoral component loosening was seen with a sensitivity of 89% by DSAr and 100% by AS, respectively. Specificity of DSAr in the femoral compartment was 94%, of AS 88%. CONCLUSIONS: DSAr and AS are useful tools in the diagnosis of hip endoprothesis loosening and in planning surgical treatment. Both methods are comparable concerning the acetabular component, whereas AS is more sensitive in detecting femoral stem loosening. Combination of both methods is not recommended for routine diagnosis.


Subject(s)
Arthrography/methods , Arthroplasty, Replacement, Hip , Hip Prosthesis , Prosthesis Failure , Radiographic Image Enhancement , Aged , Bone and Bones/diagnostic imaging , Female , Humans , Male , Middle Aged , Radionuclide Imaging , Reoperation , Sensitivity and Specificity
SELECTION OF CITATIONS
SEARCH DETAIL
...