1.
Med Phys ; 51(6): 4389-4401, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38703397

ABSTRACT

BACKGROUND: Biology-guided radiotherapy (BgRT) is a novel radiotherapy delivery technique that utilizes the tumor itself to guide dynamic delivery of treatment dose to the tumor. The RefleXion X1 system is the first radiotherapy system developed to deliver SCINTIX® BgRT. The X1 is characterized by its split arc design, employing two 90-degree positron emission tomography (PET) arcs to guide therapeutic radiation beams in real time, currently cleared by FDA to treat bone and lung tumors. PURPOSE: This study aims to comprehensively evaluate the capabilities of the SCINTIX radiotherapy delivery system by evaluating its sensitivity to changes in PET contrast, its adaptability in the context of patient motion, and its performance across a spectrum of prescription doses. METHODS: A series of experimental scenarios, both static and dynamic, were designed to assess the SCINTIX BgRT system's performance, including an end-to-end test. These experiments involved a range of factors, including changes in PET contrast, motion, and prescription doses. Measurements were performed using a custom-made ArcCHECK insert which included a 2.2 cm spherical target and a c-shape structure that can be filled with a PET tracer with varying concentrations. Sinusoidal and cosine4 motion patterns, simulating patient breathing, was used to test the SCINTIX system's ability to deliver BgRT during motion-induced challenges. Each experiment was evaluated against specific metrics, including Activity Concentration (AC), Normalized Target Signal (NTS), and Biology Tracking Zone (BTZ) bounded dose-volume histogram (bDVH) pass rates. The accuracy of the delivered BgRT doses on ArcCHECK and EBT-XD film were evaluated using gamma 3%/2 mm and 3%/3 mm analysis. RESULTS: In static scenarios, the X1 system consistently demonstrated precision and robustness in SCINTIX dose delivery. The end-to-end delivery to the spherical target yielded good results, with AC and NTS values surpassing the critical thresholds of 5 kBq/mL and 2, respectively. Furthermore, bDVH analysis consistently confirmed 100% pass rates. These results were reaffirmed in scenarios involving changes in PET contrast, emphasizing the system's ability to adapt to varying PET avidities. Gamma analysis with 3%/2 mm (10% dose threshold) criteria consistently achieved pass rates > 91.5% for the static tests. In dynamic SCINTIX delivery scenarios, the X1 system exhibited adaptability under conditions of motion. Sinusoidal and cosine4 motion patterns resulted in 3%/3 mm gamma pass rates > 87%. Moreover, the comparison with gated stereotactic body radiotherapy (SBRT) delivery on a conventional c-arm Linac resulted in 93.9% gamma pass rates and used as comparison to evaluate the interplay effect. The 1 cm step shift tests showed low overall gamma pass rates of 60.3% in ArcCHECK measurements, while the doses in the PTV agreed with the plan with 99.9% for 3%/3 mm measured with film. CONCLUSIONS: The comprehensive evaluation of the X1 radiotherapy delivery system for SCINTIX BgRT demonstrated good agreement for the static tests. The system consistently achieved critical metrics and delivered the BgRT doses per plan. The motion tests demonstrated its ability to co-localize the dose where the PET signal is and deliver acceptable BgRT dose distributions.


Subject(s)
Positron-Emission Tomography , Radiotherapy, Image-Guided , Positron-Emission Tomography/instrumentation , Radiotherapy, Image-Guided/instrumentation , Radiotherapy, Image-Guided/methods , Particle Accelerators , Humans , Radiotherapy Dosage
2.
Article in English | MEDLINE | ID: mdl-38493902

ABSTRACT

PURPOSE: We conducted a multi-institutional dosimetric audit between FLASH and conventional dose rate (CONV) electron irradiations by using an anatomically realistic 3-dimensional (3D) printed mouse phantom. METHODS AND MATERIALS: A computed tomography (CT) scan of a live mouse was used to create a 3D model of bony anatomy, lungs, and soft tissue. A dual-nozzle 3D printer was used to print the mouse phantom using acrylonitrile butadiene styrene (∼1.02 g/cm3) and polylactic acid (∼1.24 g/cm3) simultaneously to simulate soft tissue and bone densities, respectively. The lungs were printed separately using lightweight polylactic acid (∼0.64 g/cm3). Hounsfield units (HU), densities, and print-to-print stability of the phantoms were assessed. Three institutions were each provided a phantom and each institution performed 2 replicates of irradiations at selected anatomic regions. The average dose difference between FLASH and CONV dose distributions and deviation from the prescribed dose were measured with radiochromic film. RESULTS: Compared with the reference CT scan, CT scans of the phantom demonstrated mass density differences of 0.10 g/cm3 for bone, 0.12 g/cm3 for lung, and 0.03 g/cm3 for soft tissue regions. Differences in HU between phantoms were <10 HU for soft tissue and bone, with lung showing the most variation (54 HU), but with minimal effect on dose distribution (<0.5%). Mean differences between FLASH and CONV decreased from the first to the second replicate (4.3%-1.2%), and differences from the prescribed dose decreased for both CONV (3.6%-2.5%) and FLASH (6.4%-2.7%). Total dose accuracy suggests consistent pulse dose and pulse number, although these were not specifically assessed. Positioning variability was observed, likely due to the absence of robust positioning aids or image guidance. CONCLUSIONS: This study marks the first dosimetric audit for FLASH using a nonhomogeneous phantom, challenging conventional calibration practices reliant on homogeneous phantoms. The comparison protocol offers a framework for credentialing multi-institutional studies in FLASH preclinical research to enhance reproducibility of biologic findings.

3.
Pediatr Cardiol ; 45(5): 986-997, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38509208

ABSTRACT

Despite advancements in postoperative outcomes after Fontan surgery, there remains a risk of suboptimal outcomes and significant morbidity in the early postoperative period. Anatomical obstructions in the Fontan pathway can lead to prolonged pleural effusion or ascites, cyanosis, and low cardiac output syndrome (LCOS). Transcatheter interventions offer an alternative to early re-surgery for treating these complications. Over a 13-year period, early catheter angiography, performed within 30 days post-index procedure, was administered to 41 patients, identifying anatomical issues that necessitated re-intervention in 39 cases. This led to transcatheter interventions in 37 (10.4%) of the 344 Fontan surgery patients. The median age was 4.8 years (IQR: 4-9.4), and the median weight was 16.5 kg (IQR: 15-25.2), with females comprising 51.4% (19/37) of this group. The primary indications for the procedures were persistent pleural effusion or ascites in 27 patients (66%), LCOS in 8 patients (20%), and cyanosis in 6 patients (14%). Among the 37 undergoing transcatheter intervention, 30 were treated solely with this method and discharged, three died in ICU follow-up, and four required early re-surgery. No procedural mortality was observed. Our findings demonstrate that transcatheter interventions, including stent implantation, balloon angioplasty, and fenestration dilation, are safe and effective in the early post-Fontan period. Therefore, they should be considered an integral part of the management strategy for this patient group.


Subject(s)
Cardiac Catheterization , Fontan Procedure , Heart Defects, Congenital , Postoperative Complications , Humans , Female , Male , Child, Preschool , Cardiac Catheterization/methods , Heart Defects, Congenital/surgery , Child , Retrospective Studies , Treatment Outcome , Pleural Effusion/therapy , Follow-Up Studies , Palliative Care/methods
4.
Med Phys ; 2024 Mar 07.
Article in English | MEDLINE | ID: mdl-38452277

ABSTRACT

BACKGROUND: Biology-guided radiotherapy (BgRT) is a novel technology that uses positron emission tomography (PET) data to direct radiotherapy delivery in real-time. BgRT enables the precise delivery of radiation doses based on the PET signals emanating from PET-avid tumors on the fly. In this way, BgRT uniquely utilizes radiotracer uptake as a biological beacon for controlling and adjusting dose delivery in real-time to account for target motion. PURPOSE: To demonstrate using real-time PET for BgRT delivery on the RefleXion X1 radiotherapy machine. The X1 radiotherapy machine is a rotating ring-gantry radiotherapy system that generates a nominal 6MV photon beam, PET, and computed tomography (CT) components. The system utilizes emitted photons from PET-avid targets to deliver effective radiation beamlets or pulses to the tumor in real-time. METHODS: This study demonstrated a real-time PET BgRT delivery experiment under three scenarios. These scenarios included BgRT delivering to (S1 ) a static target in a homogeneous and heterogeneous environment, (S2 ) a static target with a hot avoidance structure and partial PET-avid target, and (S3 ) a moving target. The first step was to create stereotactic body radiotherapy (SBRT) and BgRT plans (offline PET data supported) using RefleXion's custom-built treatment planning system (TPS). Additionally, to create a BgRT plan using PET-guided delivery, the targets were filled with 18F-Fluorodeoxyglucose (FDG), which represents a tumor/target, that is, PET-avid. The background materials were created in the insert with homogeneous water medium (for S1 ) and heterogeneous water with styrofoam mesh medium. A heterogeneous background medium simulated soft tissue surrounding the tumor. The treatment plan was then delivered to the experimental setups using a pre-commercial version of the X1 machine. As a final step, the dosimetric accuracy for S1 and S2 was assessed using the ArcCheck analysis tool-the gamma criteria of 3%/3 mm. For S3 , the delivery dose was quantified using EBT-XD radiochromic film. The accuracy criteria were based on coverage, where 100% of the clinical target volume (CTV) receives at least 97% of the prescription dose, and the maximum dose in the CTV was ≤130% of the maximum planned dose (97 % ≤ CTV ≤ 130%). RESULTS: For the S1, both SBRT and BgRT deliveries had gamma pass rates greater than 95% (SBRT range: 96.9%-100%, BgRT range: 95.2%-98.9%), while in S2 , the gamma pass rate was 98% for SBRT and between 95.2% and 98.9% for BgRT plan delivering. For S3 , both SBRT and BgRT motion deliveries met CTV dose coverage requirements, with BgRT plans delivering a very high dose to the target. The CTV dose ranges were (a) SBRT:100.4%-120.4%, and (b) BgRT: 121.3%-139.9%. CONCLUSIONS: This phantom-based study demonstrated that PET signals from PET-avid tumors can be utilized to direct real-time dose delivery to the tumor accurately, which is comparable to the dosimetric accuracy of SBRT. Furthermore, BgRT delivered a PET-signal controlled dose to the moving target, equivalent to the dose distribution to the static target. A future study will compare the performance of BgRT with conventional image-guided radiotherapy.

5.
Phys Med ; 119: 103318, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38382210

ABSTRACT

PURPOSE: This study explores the feasibility of employing Generative Adversarial Networks (GANs) to model the RefleXion X1 Linac. The aim is to investigate the accuracy of dose simulation and assess the potential computational benefits. METHODS: The X1 Linac is a new radiotherapy machine with a binary multi-leaf collimation (MLC) system, facilitating innovative biology-guided radiotherapy. A total of 34 GAN generators, each representing a desired MLC aperture, were developed. Each generator was trained using a phase space file generated underneath the corresponding aperture, enabling the generation of particles and serving as a beam source for Monte Carlo simulation. Dose distributions in water were simulated for each aperture using both the GAN and phase space sources. The agreement between dose distributions was evaluated. The computational time reduction from bypassing the collimation simulation and storage space savings were estimated. RESULTS: The percentage depth dose at 10 cm, penumbra, and full-width half maximum of the GAN simulation agree with the phase space simulation, with differences of 0.4 % ± 0.2 %, 0.32 ± 0.66 mm, and 0.26 ± 0.44 mm, respectively. The gamma passing rate (1 %/1mm) for the planar dose exceeded 90 % for all apertures. The estimated time-saving for simulating an plan using 5766 beamlets was 530 CPU hours. The storage usage was reduced by a factor of 102. CONCLUSION: The utilization of the GAN in simulating the X1 Linac demonstrated remarkable accuracy and efficiency. The reductions in both computational time and storage requirements make this approach highly valuable for future dosimetry studies and beam modeling.


Subject(s)
Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Radiotherapy Planning, Computer-Assisted/methods , Monte Carlo Method , Computer Simulation , Radiotherapy Dosage , Radiotherapy, Intensity-Modulated/methods , Particle Accelerators
6.
Pediatr Cardiol ; 45(3): 632-639, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38182891

ABSTRACT

In individuals with a single ventricle undergoing evaluation before Fontan surgery, the presence of excessive pulmonary blood flow can contribute to increased pulmonary artery pressure, notably in those who had a Glenn procedure with antegrade pulmonary flow. 28 patients who had previously undergone Glenn anastomosis with antegrade pulmonary blood flow (APBF) and with elevated mean pulmonary artery (mPAP) pressure > 15 mmHg in diagnostic catheter angiography were included in the study. After addressing other anatomical factors that could affect pulmonary artery pressure, APBF was occluded with semi-compliant, Wedge or sizing balloons to measure pulmonary artery pressure accurately. 23 patients (82% of the cohort) advanced to Fontan completion. In this group, median mPAP dropped from 20.5 (IQR 19-22) mmHg to 13 (IQR 12-14) mmHg post-test (p < 0.001). Median PVR post-test was 1.8 (IQR 1.5-2.1) WU m2. SpO2 levels decreased from a median of 88% (IQR 86%-93%) pre-test to 80% (IQR 75%-84%) post-test (p < 0.001). In five patients, elevated mPAP post-test occlusion on diagnostic catheter angiography led to non-completion of Fontan circulation. In this group, median pre- and post-test mPAP were 23 mmHg (IQR 21.5-23.5) and 19 mmHg (IQR 18.5-20), respectively (p = 0.038). Median post-test PVR was 3.8 (IQR 3.6-4.5) WU m2. SpO2 levels decreased from a median of 79% (IQR 76%-81%) pre-test to 77% (IQR 73.5%-80%) post-test (p = 0.039). Our study presents a specialized approach for patients initially deemed unsuitable for Fontan due to elevated pulmonary artery pressures. We were able to successfully complete the Fontan procedure in the majority of these high-risk cases after temporary balloon occlusion test.


Subject(s)
Fontan Procedure , Heart Defects, Congenital , Humans , Pulmonary Artery/diagnostic imaging , Pulmonary Artery/surgery , Pulmonary Circulation/physiology , Heart Defects, Congenital/surgery , Retrospective Studies , Treatment Outcome , Fontan Procedure/methods
7.
Adv Radiat Oncol ; 9(1): 101300, 2024 Jan.
Article in English | MEDLINE | ID: mdl-38260216

ABSTRACT

Purpose: The aim of this study was to present the first-year experience of treating patients using intensity modulated radiation therapy (IMRT) and stereotactic body radiation therapy (SBRT) with a biology-guided radiation therapy machine, the RefleXion X1 system, installed in a clinical setting. Methods and Materials: A total of 78 patients were treated on the X1 system using IMRT and SBRT from May 2021 to May 2022. Clinical and technical data including treatment sites, number of pretreatment kilovoltage computed tomography (kVCT) scans, beam-on time, patient setup time, and imaging time were collected and analyzed. Machine quality assurance (QA) results, machine performance, and user satisfactory survey were also collected and reported. Results: The most commonly treated site was the head and neck (63%), followed by the pelvis (23%), abdomen (8%), and thorax (6%). Except for 5 patients (6%) who received SBRT treatments for bony metastases in the pelvis, all treatments were conventionally fractionated IMRT. The number of kVCT scans per fraction was 1.2 ± 0.5 (mean ± standard deviation). The beam-on time was 9.2 ± 3.5 minutes. The patient setup time and imaging time per kVCT was 4.8 ± 2.6 minutes and 4.6 ± 1.5 minutes, respectively. The daily machine output deviation was 0.4 ± 1.2% from the baseline. The patient QA had a passing rate of 97.4 ± 2.8% at 3%/2 mm gamma criteria. The machine uptime was 92% of the total treatment time. The daily QA and kVCT image quality received the highest level of satisfaction. The treatment workflow for therapists received the lowest level of satisfaction. Conclusions: One year after the installation, 78 patients were successfully treated with the X1 system using IMRT and/or SBRT. With the recent Food and Drug Administration clearance of biology-guided radiation therapy, our department is preparing to treat patients using positron emission tomography-guidance via a new product release, which will address deficiencies in the current image-guided radiation therapy workflow.

8.
Article in English | MEDLINE | ID: mdl-38171387

ABSTRACT

PURPOSE: Ultrahigh-dose-rate (FLASH) irradiation has been reported to reduce normal tissue damage compared with conventional dose rate (CONV) irradiation without compromising tumor control. This proof-of-concept study aims to develop a deep learning (DL) approach to quantify the FLASH isoeffective dose (dose of CONV that would be required to produce the same effect as the given physical FLASH dose) with postirradiation mouse intestinal histology images. METHODS AND MATERIALS: Eighty-four healthy C57BL/6J female mice underwent 16 MeV electron CONV (0.12 Gy/s; n = 41) or FLASH (200 Gy/s; n = 43) single fraction whole abdominal irradiation. Physical dose ranged from 12 to 16 Gy for FLASH and 11 to 15 Gy for CONV in 1 Gy increments. Four days after irradiation, 9 jejunum cross-sections from each mouse were hematoxylin and eosin stained and digitized for histological analysis. CONV data set was randomly split into training (n = 33) and testing (n = 8) data sets. ResNet101-based DL models were retrained using the CONV training data set to estimate the dose based on histological features. The classical manual crypt counting (CC) approach was implemented for model comparison. Cross-section-wise mean squared error was computed to evaluate the dose estimation accuracy of both approaches. The validated DL model was applied to the FLASH data set to map the physical FLASH dose into the isoeffective dose. RESULTS: The DL model achieved a cross-section-wise mean squared error of 0.20 Gy2 on the CONV testing data set compared with 0.40 Gy2 of the CC approach. Isoeffective doses estimated by the DL model for FLASH doses of 12, 13, 14, 15, and 16 Gy were 12.19 ± 0.46, 12.54 ± 0.37, 12.69 ± 0.26, 12.84 ± 0.26, and 13.03 ± 0.28 Gy, respectively. CONCLUSIONS: Our proposed DL model achieved accurate CONV dose estimation. The DL model results indicate that in the physical dose range of 13 to 16 Gy, the biologic dose response of small intestinal tissue to FLASH irradiation is represented by a lower isoeffective dose compared with the physical dose. Our DL approach can be a tool for studying isoeffective doses of other radiation dose modifying interventions.

9.
Int J Radiat Oncol Biol Phys ; 118(5): 1172-1180, 2024 Apr 01.
Article in English | MEDLINE | ID: mdl-38147912

ABSTRACT

PURPOSE: Positron emission tomography (PET)-guided radiation therapy is a novel tracked dose delivery modality that uses real-time PET to guide radiation therapy beamlets. The BIOGUIDE-X study was performed with sequential cohorts of participants to (1) identify the fluorodeoxyglucose (FDG) dose for PET-guided therapy and (2) confirm that the emulated dose distribution was consistent with a physician-approved radiation therapy plan. METHODS AND MATERIALS: This prospective study included participants with at least 1 FDG-avid targetable primary or metastatic tumor (2-5 cm) in the lung or bone. For cohort I, a modified 3 + 3 design was used to determine the FDG dose that would result in adequate signal for PET-guided therapy. For cohort II, PET imaging data were collected on the X1 system before the first and last fractions among patients undergoing conventional stereotactic body radiation therapy. PET-guided therapy dose distributions were modeled on the patient's computed tomography anatomy using the collected PET data at each fraction as input to an "emulated delivery" and compared with the physician-approved plan. RESULTS: Cohort I demonstrated adequate FDG activity in 6 of 6 evaluable participants (100.0%) with the first injected dose level of 15 mCi FDG. In cohort II, 4 patients with lung tumors and 5 with bone tumors were enrolled, and evaluable emulated delivery data points were collected for 17 treatment fractions. Sixteen of the 17 emulated deliveries resulted in dose distributions that were accurate with respect to the approved PET-guided therapy plan. The 17th data point was just below the 95% threshold for accuracy (dose-volume histogram score = 94.6%). All emulated fluences were physically deliverable. No toxicities were attributed to multiple FDG administrations. CONCLUSIONS: PET-guided therapy is a novel radiation therapy modality in which a radiolabeled tumor can act as its own fiducial for radiation therapy targeting. Emulated therapy dose distributions calculated from continuously acquired real-time PET data were accurate and machine-deliverable in tumors that were 2 to 5 cm in size with adequate FDG signal characteristics.


Subject(s)
Fluorodeoxyglucose F18 , Lung Neoplasms , Humans , Prospective Studies , Positron-Emission Tomography , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/radiotherapy , Lung Neoplasms/pathology , Tomography, X-Ray Computed/methods , Radiopharmaceuticals
10.
ArXiv ; 2023 Sep 28.
Article in English | MEDLINE | ID: mdl-37808098

ABSTRACT

We conducted a multi-institutional audit of dosimetric variability between FLASH and conventional dose rate (CONV) electron irradiations by using an anatomically realistic 3D-printed mouse phantom. A CT scan of a live mouse was used to create a 3D model of bony anatomy, lungs, and soft tissue. A dual-nozzle 3D printer was used to print the mouse phantom using acrylonitrile butadiene styrene ($~1.02 g/cm^3$) and polylactic acid ($~1.24 g/cm^3$) simultaneously to simulate soft tissue and bone densities, respectively. The lungs were printed separately using lightweight polylactic acid ($~0.64 g/cm^3$). Hounsfield units (HU) and densities were compared with the reference CT scan of the live mouse. Print-to-print reproducibility of the phantom was assessed. Three institutions were each provided a phantom, and each institution performed two replicates of irradiations at selected mouse anatomic regions. The average dose difference between FLASH and CONV dose distributions and deviation from the prescribed dose were measured with radiochromic film. Compared to the reference CT scan, CT scans of the phantom demonstrated mass density differences of $0.10 g/cm^3$ for bone, $0.12 g/cm^3$ for lung, and $0.03 g/cm^3$ for soft tissue regions. Between phantoms, the difference in HU for soft tissue and bone was <10 HU from print to print. Lung exhibited the most variation (54 HU) but minimally affected dose distribution (<0.5% dose differences between phantoms). The mean difference between FLASH and CONV from the first replicate to the second decreased from 4.3% to 1.2%, and the mean difference from the prescribed dose decreased from 3.6% to 2.5% for CONV and 6.4% to 2.7% for FLASH. The framework presented here is promising for credentialing of multi-institutional studies of FLASH preclinical research to maximize the reproducibility of biological findings.

11.
Radiother Oncol ; 188: 109906, 2023 Nov.
Article in English | MEDLINE | ID: mdl-37690668

ABSTRACT

BACKGROUND AND PURPOSE: The impact of radiotherapy (RT) at ultra high vs conventional dose rate (FLASH vs CONV) on the generation and repair of DNA double strand breaks (DSBs) is an important question that remains to be investigated. Here, we tested the hypothesis as to whether FLASH-RT generates decreased chromosomal translocations compared to CONV-RT. MATERIALS AND METHODS: We used two FLASH validated electron beams and high-throughput rejoin and genome-wide translocation sequencing (HTGTS-JoinT-seq), employing S. aureus and S. pyogenes Cas9 "bait" DNA double strand breaks (DSBs) in HEK239T cells, to measure differences in bait-proximal repair and their genome-wide translocations to "prey" DSBs generated after various irradiation doses, dose rates and oxygen tensions (normoxic, 21% O2; physiological, 4% O2; hypoxic, 2% and 0.5% O2). Electron irradiation was delivered using a FLASH capable Varian Trilogy and the eRT6/Oriatron at CONV (0.08-0.13 Gy/s) and FLASH (1x102-5x106 Gy/s) dose rates. Related experiments using clonogenic survival and γH2AX foci in the 293T and the U87 glioblastoma lines were also performed to discern FLASH-RT vs CONV-RT DSB effects. RESULTS: Normoxic and physioxic irradiation of HEK293T cells increased translocations at the cost of decreasing bait-proximal repair but were indistinguishable between CONV-RT and FLASH-RT. Although no apparent increase in chromosome translocations was observed with hypoxia-induced apoptosis, the combined decrease in oxygen tension with IR dose-rate modulation did not reveal significant differences in the level of translocations nor in their junction structures. Furthermore, RT dose rate modality on U87 cells did not change γH2AX foci numbers at 1- and 24-hours post-irradiation nor did this affect 293T clonogenic survival. CONCLUSION: Irrespective of oxygen tension, FLASH-RT produces translocations and junction structures at levels and proportions that are indistinguishable from CONV-RT.

12.
Pediatr Cardiol ; 2023 Jul 20.
Article in English | MEDLINE | ID: mdl-37474608

ABSTRACT

Transcatheter closure of the tubular ducts remains the most challenging procedure, with higher complication rates than other types. This study evaluates the characteristics of transcatheter closure of tubular ducts with pulmonary hypertension. 73 patients with tubular ducts who underwent cardiac catheterization for transcatheter PDA closure were analyzed. The mean age and weight were 1.93 ± 2.68 years and 8.83 ± 6.14 kg, respectively. Transcatheter closure was attempted in 72 patients. Four cases (5.5%) were referred to surgery, while the procedure was completed in the remaining (94.5%). Amplatzer duct occluder (ADO) I or Cardiofix duct occluder (CDO) was the most commonly used devices. However, the use of Amplatzer vascular plug (AVP) II raised in recent years. The most common concern was aortic protrusion/stenosis in ADO I/CDO devices, but most regressed during follow-up. Iatrogenic coarctation of the aorta was observed in two with ADO I/CDO. Embolization of the device to the pulmonary artery was observed in three with CDO, AVP II, and AVP I. Significant left pulmonary artery stenosis requiring stenting developed in one after closure with an MVSDO device. Tubular ducts are highly associated with pulmonary arterial hypertension, and transcatheter closure of them is still challenging despite the developing device armamentarium. Although ADO I or similar devices are widely used, off-label devices are usually needed at increasing rates. The AVP II device is unsuitable for short tubular ducts but seems the best option for long ones.

13.
Int J Radiat Oncol Biol Phys ; 117(2): 505-514, 2023 10 01.
Article in English | MEDLINE | ID: mdl-37141982

ABSTRACT

PURPOSE: This study explored deep-learning-based patient-specific auto-segmentation using transfer learning on daily RefleXion kilovoltage computed tomography (kVCT) images to facilitate adaptive radiation therapy, based on data from the first group of patients treated with the innovative RefleXion system. METHODS AND MATERIALS: For head and neck (HaN) and pelvic cancers, a deep convolutional segmentation network was initially trained on a population data set that contained 67 and 56 patient cases, respectively. Then the pretrained population network was adapted to the specific RefleXion patient by fine-tuning the network weights with a transfer learning method. For each of the 6 collected RefleXion HaN cases and 4 pelvic cases, initial planning computed tomography (CT) scans and 5 to 26 sets of daily kVCT images were used for the patient-specific learning and evaluation separately. The performance of the patient-specific network was compared with the population network and the clinical rigid registration method and evaluated by the Dice similarity coefficient (DSC) with manual contours being the reference. The corresponding dosimetric effects resulting from different auto-segmentation and registration methods were also investigated. RESULTS: The proposed patient-specific network achieved mean DSC results of 0.88 for 3 HaN organs at risk (OARs) of interest and 0.90 for 8 pelvic target and OARs, outperforming the population network (0.70 and 0.63) and the registration method (0.72 and 0.72). The DSC of the patient-specific network gradually increased with the increment of longitudinal training cases and approached saturation with more than 6 training cases. Compared with using the registration contour, the target and OAR mean doses and dose-volume histograms obtained using the patient-specific auto-segmentation were closer to the results using the manual contour. CONCLUSIONS: Auto-segmentation of RefleXion kVCT images based on the patient-specific transfer learning could achieve higher accuracy, outperforming a common population network and clinical registration-based method. This approach shows promise in improving dose evaluation accuracy in RefleXion adaptive radiation therapy.


Subject(s)
Image Processing, Computer-Assisted , Radiotherapy Planning, Computer-Assisted , Humans , Radiotherapy Planning, Computer-Assisted/methods , Image Processing, Computer-Assisted/methods , Organs at Risk/diagnostic imaging , Organs at Risk/radiation effects , Radiometry , Tomography, X-Ray Computed
14.
Allergol Immunopathol (Madr) ; 51(3): 8-14, 2023.
Article in English | MEDLINE | ID: mdl-37169554

ABSTRACT

BACKGROUND: Food allergy (FA), hence the incidence of food anaphylaxis, is a public health problem that has increased in recent years. There are still no biomarkers for patients with FA to predict severe allergic reactions such as anaphylaxis. OBJECTIVE: There is limited information on whether regulatory T (Treg) cell levels are a biomarker that predicts clinical severity in cases presenting with FA, and which patients are at a greater risk for anaphylaxis. METHODS: A total of 70 children were included in the study: 25 who had IgE-mediated cow's milk protein allergy (CMPA) and presented with non-anaphylactic symptoms (FA/A-), 16 who had IgE-mediated CMPA and presented with anaphylaxis (FA/A+) (a total of 41 FA cases), and a control group consisting of 29 children without FA. The study was conducted by performing CD4+CD25+CD127loFOXP3+ cell flow cytometric analysis during resting at least 2 weeks after the elimination diet to FA subjects. RESULTS: When the FA group was compared with healthy control subjects, CD4+CD25+CD127loFOXP3+ cell rates were found to be significantly lower in the FA group (p < 0.001). When the FA/A- and FA/A+ groups and the control group were compared in terms of CD4+CD25+CD127loFOXP3+ cell ratios, they were significantly lower in the FA/A- and FA/A+ groups compared to the control group (p < 0.001). CONCLUSIONS: Although there was no significant difference between the FA/A+ group and the FA/A- group in terms of CD4+CD25+CD127loFOXP3+ cells, our study is important, as it is the first pediatric study we know to investigate whether CD4+CD25+CD127loFOXP3+ cells in FA predict anaphylaxis.


Subject(s)
Anaphylaxis , Food Hypersensitivity , Child , Humans , Anaphylaxis/diagnosis , Anaphylaxis/metabolism , Biomarkers/metabolism , Food Hypersensitivity/immunology , Forkhead Transcription Factors/metabolism , Immunoglobulin E/metabolism , Milk Hypersensitivity/diagnosis , Milk Hypersensitivity/immunology , T-Lymphocytes, Regulatory
15.
Allergol. immunopatol ; 51(3): 8-14, 01 mayo 2023. tab
Article in English | IBECS | ID: ibc-219808

ABSTRACT

Background: Food allergy (FA), hence the incidence of food anaphylaxis, is a public health problem that has increased in recent years. There are still no biomarkers for patients with FA to predict severe allergic reactions such as anaphylaxis. Objective: There is limited information on whether regulatory T (Treg) cell levels are a biomarker that predicts clinical severity in cases presenting with FA, and which patients are at a greater risk for anaphylaxis. Methods: A total of 70 children were included in the study: 25 who had IgE-mediated cow’s milk protein allergy (CMPA) and presented with non-anaphylactic symptoms (FA/A−), 16 who had IgE-mediated CMPA and presented with anaphylaxis (FA/A+) (a total of 41 FA cases), and a control group consisting of 29 children without FA. The study was conducted by performing CD4+CD25+CD127loFOXP3+ cell flow cytometric analysis during resting at least 2 weeks after the elimination diet to FA subjects. Results: When the FA group was compared with healthy control subjects, CD4+CD25+CD127loFOXP3+ cell rates were found to be significantly lower in the FA group (p < 0.001). When the FA/A− and FA/A+ groups and the control group were compared in terms of CD4+CD25+CD127loFOXP3+ cell ratios, they were significantly lower in the FA/A− and FA/A+ groups compared to the control group (p < 0.001). Conclusions: Although there was no significant difference between the FA/A+ group and the FA/A− group in terms of CD4+CD25+CD127loFOXP3+ cells, our study is important, as it is the first pediatric study we know to investigate whether CD4+CD25+CD127loFOXP3+ cells in FA predict anaphylaxis (AU)


Subject(s)
Humans , Male , Female , Infant , Child, Preschool , Forkhead Box Protein O3/blood , Milk Hypersensitivity/diagnosis , Milk Hypersensitivity/immunology , Anaphylaxis/etiology , Anaphylaxis/immunology , /enzymology , Milk Hypersensitivity/complications , Biomarkers/blood , CD4 Antigens/immunology , Interleukin-2 Receptor alpha Subunit/immunology
16.
bioRxiv ; 2023 Mar 27.
Article in English | MEDLINE | ID: mdl-37034651

ABSTRACT

The molecular and cellular mechanisms driving the enhanced therapeutic ratio of ultra-high dose-rate radiotherapy (FLASH-RT) over slower conventional (CONV-RT) radiotherapy dose-rate remain to be elucidated. However, attenuated DNA damage and transient oxygen depletion are among several proposed models. Here, we tested whether FLASH-RT under physioxic (4% O 2 ) and hypoxic conditions (≤2% O 2 ) reduces genome-wide translocations relative to CONV-RT and whether any differences identified revert under normoxic (21% O 2 ) conditions. We employed high-throughput rejoin and genome-wide translocation sequencing ( HTGTS-JoinT-seq ), using S. aureus and S. pyogenes Cas9 "bait" DNA double strand breaks (DSBs), to measure differences in bait-proximal repair and their genome-wide translocations to "prey" DSBs generated by electron beam CONV-RT (0.08-0.13Gy/s) and FLASH-RT (1×10 2 -5×10 6 Gy/s), under varying ionizing radiation (IR) doses and oxygen tensions. Normoxic and physioxic irradiation of HEK293T cells increased translocations at the cost of decreasing bait-proximal repair but were indistinguishable between CONV-RT and FLASH-RT. Although no apparent increase in chromosome translocations was observed with hypoxia-induced apoptosis, the combined decrease in oxygen tension with IR dose-rate modulation did not reveal significant differences in the level of translocations nor in their junction structures. Thus, Irrespective of oxygen tension, FLASH-RT produces translocations and junction structures at levels and proportions that are indistinguishable from CONV-RT.

17.
Adv Radiat Oncol ; 8(5): 101186, 2023.
Article in English | MEDLINE | ID: mdl-37035034

ABSTRACT

Purpose: The aim of this study was to apply the Six Sigma methodology and failure mode and effect analysis (FMEA) to mitigate errors in intensity modulated radiation therapy (IMRT) and stereotactic body radiation therapy (SBRT) treatment planning with the first clinical installation of RefleXion X1. Methods and Materials: The Six Sigma approach consisted of 5 phases: define, measure, analyze, improve, and control. The define, measure, and analyze phases consisted of process mapping and an FMEA of IMRT and SBRT treatment planning on the X1. The multidisciplinary team outlined the workflow process and identified and ranked the failure modes associated with the plan check items using the American Association of Physicists in Medicine Task Group 100 recommendations. Items with the highest average risk priority numbers (RPNs) and severity ≥7 were prioritized for automation using the Eclipse Scripting Application Programming Interface (ESAPI). The "improve" phase consisted of developing ESAPI scripts before the clinical launch of X1 to improve efficiency and safety. In the "control" phase, the FMEA ranking was re-evaluated 1 year after clinical launch. Results: Overall, 100 plan check items were identified in which the RPN values ranged from 10.2 to 429.0. Fifty of these items (50%) were suitable for automation within ESAPI. Of the 10 highest-risk items, 8 were suitable for automation. Based on the results of the FMEA, 2 scripts were developed: Planning Assistant, used by the planner during preparation for planning, and Automated Plan Check, used by the planner and the plan checker during plan preparation for treatment. After 12 months of clinical use of the X1 and developed scripts, only 3 errors were reported. The average prescript RPN was 138.0, compared with the average postscript RPN of 47.8 (P < .05), signifying a safer process. Conclusions: Implementing new technology in the clinic can be an error-prone process in which the likelihood of errors increases with increasing pressure to implement the technology quickly. To limit errors in clinical implementation of the novel RefleXion X1 system, the Six Sigma method was used to identify failure modes, establish quality control checks, and re-evaluate these checks 1 year after clinical implementation.

18.
Int J Radiat Oncol Biol Phys ; 116(5): 1202-1217, 2023 Aug 01.
Article in English | MEDLINE | ID: mdl-37121362

ABSTRACT

FLASH radiation therapy (FLASH-RT), delivered with ultrahigh dose rate (UHDR), may allow patients to be treated with less normal tissue toxicity for a given tumor dose compared with currently used conventional dose rate. Clinical trials are being carried out and are needed to test whether this improved therapeutic ratio can be achieved clinically. During the clinical trials, quality assurance and credentialing of equipment and participating sites, particularly pertaining to UHDR-specific aspects, will be crucial for the validity of the outcomes of such trials. This report represents an initial framework proposed by the NRG Oncology Center for Innovation in Radiation Oncology FLASH working group on quality assurance of potential UHDR clinical trials and reviews current technology gaps to overcome. An important but separate consideration is the appropriate design of trials to most effectively answer clinical and scientific questions about FLASH. This paper begins with an overview of UHDR RT delivery methods. UHDR beam delivery parameters are then covered, with a focus on electron and proton modalities. The definition and control of safe UHDR beam delivery and current and needed dosimetry technologies are reviewed and discussed. System and site credentialing for large, multi-institution trials are reviewed. Quality assurance is then discussed, and new requirements are presented for treatment system standard analysis, patient positioning, and treatment planning. The tables and figures in this paper are meant to serve as reference points as we move toward FLASH-RT clinical trial performance. Some major questions regarding FLASH-RT are discussed, and next steps in this field are proposed. FLASH-RT has potential but is associated with significant risks and complexities. We need to redefine optimization to focus not only on the dose but also on the dose rate in a manner that is robust and understandable and that can be prescribed, validated, and confirmed in real time. Robust patient safety systems and access to treatment data will be critical as FLASH-RT moves into the clinical trials.


Subject(s)
Credentialing , Electrons , Humans , Health Facilities , Patient Positioning , Technology , Radiotherapy Dosage
19.
Int J Radiat Oncol Biol Phys ; 117(2): 482-492, 2023 10 01.
Article in English | MEDLINE | ID: mdl-37105403

ABSTRACT

PURPOSE: Ultrahigh-dose-rate (UHDR) radiation therapy (RT) has produced the FLASH effect in preclinical models: reduced toxicity with comparable tumor control compared with conventional-dose-rate RT. Early clinical trials focused on UHDR RT feasibility using specialized devices. We explore the technical feasibility of practical electron UHDR RT on a standard clinical linear accelerator (LINAC). METHODS AND MATERIALS: We tuned the program board of a decommissioned electron energy for UHDR electron delivery on a clinical LINAC without hardware modification. Pulse delivery was controlled using the respiratory gating interface. A short source-to-surface distance (SSD) electron setup with a standard scattering foil was configured and tested on an anthropomorphic phantom using circular blocks with 3- to 20-cm field sizes. Dosimetry was evaluated using radiochromic film and an ion chamber profiler. RESULTS: UHDR open-field mean dose rates at 100, 80, 70, and 59 cm SSD were 36.82, 59.52, 82.01, and 112.83 Gy/s, respectively. At 80 cm SSD, mean dose rate was ∼60 Gy/s for all collimated field sizes, with an R80 depth of 6.1 cm corresponding to an energy of 17.5 MeV. Heterogeneity was <5.0% with asymmetry of 2.2% to 6.2%. The short SSD setup was feasible under realistic treatment conditions simulating broad clinical indications on an anthropomorphic phantom. CONCLUSIONS: Short SSD and tuning for high electron beam current on a standard clinical LINAC can deliver flat, homogenous UHDR electrons over a broad, clinically relevant range of field sizes and depths with practical working distances in a configuration easily reversible to standard clinical use.


Subject(s)
Electrons , Neoplasms , Humans , Radiometry/methods , Particle Accelerators , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy Dosage