ABSTRACT

Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided.


Subject(s)
Artifacts , Consciousness Monitors , Facial Nerve/physiology , Monitoring, Intraoperative/adverse effects , Transcutaneous Electric Nerve Stimulation , Anesthesia, General , Anesthetics, Intravenous/administration & dosage , Craniotomy , Electroencephalography , Electromyography , Female , Fentanyl/administration & dosage , Humans , Meningeal Neoplasms/surgery , Meningioma/surgery , Middle Aged , Monitoring, Intraoperative/instrumentation , Monitoring, Intraoperative/methods , Neuromuscular Nondepolarizing Agents/administration & dosage , Propofol/administration & dosage , Vecuronium Bromide/administration & dosage