1.
Bioanalysis ; 7(18): 2321-2329, 2015 Sep.
Article in English | MEDLINE | ID: mdl-26417882

ABSTRACT

BACKGROUND: An UPLC-MS detection method for the quantification of amikacin, flucloxacillin, meropenem, penicillin G and vancomycin was developed and validated. RESULTS: The calibration curves were found to be linear from 0.47 to 50 mg/l for amikacin, 1.28 to 135 mg/l for flucloxacillin, 0.75 to 80 mg/l for meropenem, 0.38 to 80 mg/l for penicillin G and 0.73 to 80 mg/l for vancomycin. Between- and within-run accuracy was ranged between 85 and 115%. Between and within imprecision expressed as CV was within 15%. CONCLUSION: The validated method was successfully applied to a PK study in very preterm and small for gestational age infants treated for nosocomial sepsis and/or meningitis.