Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 10 de 10
Filter
Add more filters










Publication year range
1.
Proc SPIE Int Soc Opt Eng ; 9033: 903321, 2014 Mar 19.
Article in English | MEDLINE | ID: mdl-25301999

ABSTRACT

Neuro-EIGIs require visualization of very small endovascular devices and small vessels. A Microangiographic Fluoroscope (MAF) x-ray detector was developed to improve on the standard flat panel detector's (FPD's) ability to visualize small objects during neuro-EIGIs. To compare the performance of FPD and MAF imaging systems, specific imaging tasks related to those encountered during neuro-EIGIs were used to assess contrast to noise ratio (CNR) of different objects. A bar phantom and a stent were placed at a fixed distance from the x-ray focal spot to mimic a clinical imaging geometry and both objects were imaged by each detector system. Imaging was done without anti-scatter grids and using the same conditions for each system including: the same x-ray beam quality, collimator position, source to imager distance (SID), and source to object distance (SOD). For each object, relative contrasts were found for both imaging systems using the peak and trough signals. The relative noise was found using mean background signal and background noise for varying detector exposures. Next, the CNRs were found for these values for each object imaged and for each imaging system used. A relative CNR metric is defined and used to compare detector imaging performance. The MAF utilizes a temporal filter to reduce the overall image noise. The effects of using this filter with the MAF while imaging the clinical object's CNRs are reported. The relative CNR for the detectors demonstrated that the MAF has superior CNRs for most objects and exposures investigated for this specific imaging task.

2.
Proc SPIE Int Soc Opt Eng ; 9033: 90332F, 2014 Mar 19.
Article in English | MEDLINE | ID: mdl-25302001

ABSTRACT

Endovascular-Image-Guided-Interventional (EIGI) treatment of neuro-vascular conditions such as aneurysms, stenosed arteries, and vessel thrombosis make use of treatment devices such as stents, coils, and balloons which have very small feature sizes, 10's of microns to a few 100's of microns, and hence demand a high resolution imaging system. The current state-of-the-art flat panel detector (FPD) has about a 200-um pixel size with the Nyquist of 2.5 lp/mm. For higher-resolution imaging a charge-coupled device (CCD) based Micro-Angio -Fluoroscope (MAF-CCD) with a pixel size of 35um (Nyquist of 11 lp/mm) was developed and previously reported. Although the detector addresses the high resolution needs, the Field-Of-View (FOV) is limited to 3.5 cm × 3.5 cm, which is much smaller than current FPDs. During the use of the MAF-CCD for delicate parts of the intervention, it may be desirable to have real-time monitoring outside the MAF FOV with a low dose, and lower, but acceptable, quality image. To address this need, a novel imaging technique for biplane imaging systems has been developed, using an MAF-CCD in the frontal plane and a dose-reduced standard large FOV imager in the lateral plane. The dose reduction is achieved by using a combination of ROI fluoroscopy and spatially different temporal filtering, a technique that has been previously presented. In order to evaluate this technique, a simulation using images acquired during an actual EIGI treatment on a patient, followed by an actual implementation on phantoms is presented.

3.
Proc SPIE Int Soc Opt Eng ; 9033: 90335S, 2014 Mar 19.
Article in English | MEDLINE | ID: mdl-25302003

ABSTRACT

Endovascular image-guided intervention (EIGI) has become the primary interventional therapy for the most widespread vascular diseases. These procedures involve the insertion of a catheter into the femoral artery, which is then threaded under fluoroscopic guidance to the site of the pathology to be treated. Flat Panel Detectors (FPDs) are normally used for EIGIs; however, once the catheter is guided to the pathological site, high-resolution imaging capabilities can be used for accurately guiding a successful endovascular treatment. The Micro-Angiographic Fluoroscope (MAF) detector provides needed high-resolution, high-sensitivity, and real-time imaging capabilities. An experimental MAF enabled with a Control, Acquisition, Processing, Image Display and Storage (CAPIDS) system was installed and aligned on a detector changer attached to the C-arm of a clinical angiographic unit. The CAPIDS system was developed and implemented using LabVIEW software and provides a user-friendly interface that enables control of several clinical radiographic imaging modes of the MAF including: fluoroscopy, roadmap, radiography, and digital-subtraction-angiography (DSA). Using the automatic controls, the MAF detector can be moved to the deployed position, in front of a standard FPD, whenever higher resolution is needed during angiographic or interventional vascular imaging procedures. To minimize any possible negative impact to image guidance with the two detector systems, it is essential to have a well-designed workflow that enables smooth deployment of the MAF at critical stages of clinical procedures. For the ultimate success of this new imaging capability, a clear understanding of the workflow design is essential. This presentation provides a detailed description and demonstration of such a workflow design.

4.
Proc SPIE Int Soc Opt Eng ; 86682013 Mar 06.
Article in English | MEDLINE | ID: mdl-24353386

ABSTRACT

Present day treatment for neurovascular pathological conditions involves the use of devices with very small features such as stents, coils, and balloons; hence, these interventional procedures demand high resolution x-ray imaging under fluoroscopic conditions to provide the capability to guide the deployment of these fine endovascular devices. To address this issue, a high resolution x-ray detector based on EMCCD technology is being developed. The EMCCD field-of-view is enlarged using a fiber-optic taper so that the detector features an effective pixel size of 37 µm giving it a Nyquist frequency of 13.5 lp/mm, which is significantly higher than that of the state of the art Flat Panel Detectors (FPD). Quantitative analysis of the detector, including gain calibration, instrumentation noise equivalent exposure (INEE) and modulation transfer function (MTF) determination, are presented in this work. The gain of the detector is a function of the detector temperature; with the detector cooled to 5° C, the highest relative gain that could be achieved was calculated to be 116 times. At this gain setting, the lowest INEE was measured to be 0.6 µR/frame. The MTF, measured using the edge method, was over 2% up to 7 cycles/ mm. To evaluate the performance of the detector under clinical conditions, an aneurysm model was placed over an anthropomorphic head phantom and a coil was guided into the aneurysm under fluoroscopic guidance using the detector. Image sequences from the procedure are presented demonstrating the high resolution of this SSXII.

5.
Proc SPIE Int Soc Opt Eng ; 86682013 Mar 06.
Article in English | MEDLINE | ID: mdl-24353387

ABSTRACT

The new Solid State X-ray Image Intensifier (SSXII) is a high-resolution, high-sensitivity, real-time region-of-interest (ROI) x-ray imaging detector. Evaluations were made of both standard linear systems metrics (MTF, DQE) and total system performance with generalized linear systems metrics (GMTF, GDQE) including scatter and geometric un-sharpness for simulated clinical conditions. The SSXII is based on a 1k × 1k EMCCD sensor coupled to a 300 µm thick CsI(Tl) phosphor through a 2.88:1 fiber optic taper resulting in a 37 µm effective pixel size and an effective 3.7 cm × 3.7 cm square field-of-view (FOV). Standard methods were used to calculate MTF, NNPS and DQE. Generalized metrics were calculated and compared for three different magnifications (1.03, 1.11 and 1.2) and three different focal spots (0.3 mm, 0.5 mm and 0.8 mm) for a scatter fraction of 0.28. For an RQA5 spectrum, at 5 cycles/mm the MTF was found to be 0.06 and DQE was 0.04, while the DQE(0) was 0.60. Focal spot un-sharpness and scatter significantly degrades the GMTF and GDQE performance of the detector. A low frequency drop is caused by scatter and is almost independent of focal spot size and magnification. The degradation for middle range frequencies is caused by geometric un-sharpness and increases with focal spot size and magnification. This degradation was least in the case of the small focal spot and almost independent of magnification. In spite of this degradation, the high resolution SSXII with a small FOV may have a significant impact on ROI image-guided neuro-interventions since it demonstrates far better performance than standard current detectors.

6.
Proc SPIE Int Soc Opt Eng ; 86682013 Mar 06.
Article in English | MEDLINE | ID: mdl-24353389

ABSTRACT

The detectors that are used for endovascular image-guided interventions (EIGI), particularly for neurovascular interventions, do not provide clinicians with adequate visualization to ensure the best possible treatment outcomes. Developing an improved x-ray imaging detector requires the determination of estimated clinical x-ray entrance exposures to the detector. The range of exposures to the detector in clinical studies was found for the three modes of operation: fluoroscopic mode, high frame-rate digital angiographic mode (HD fluoroscopic mode), and DSA mode. Using these estimated detector exposure ranges and available CMOS detector technical specifications, design requirements were developed to pursue a quantum limited, high resolution, dynamic x-ray detector based on a CMOS sensor with 50 µm pixel size. For the proposed MAF-CMOS, the estimated charge collected within the full exposure range was found to be within the estimated full well capacity of the pixels. Expected instrumentation noise for the proposed detector was estimated to be 50-1,300 electrons. Adding a gain stage such as a light image intensifier would minimize the effect of the estimated instrumentation noise on total image noise but may not be necessary to ensure quantum limited detector operation at low exposure levels. A recursive temporal filter may decrease the effective total noise by 2 to 3 times, allowing for the improved signal to noise ratios at the lowest estimated exposures despite consequent loss in temporal resolution. This work can serve as a guide for further development of dynamic x-ray imaging prototypes or improvements for existing dynamic x-ray imaging systems.

7.
Proc SPIE Int Soc Opt Eng ; 8313: 831355, 2012.
Article in English | MEDLINE | ID: mdl-22866212

ABSTRACT

Region-of-interest (ROI) fluoroscopy takes advantage of the fact that most neurovascular interventional activity is performed in only a small portion of an x-ray imaging field of view (FOV). The ROI beam filter is an attenuating material that reduces patient dose in the area peripheral to the object of interest. This project explores a method of moving the beam-attenuator aperture with the object of interest such that it always remains in the ROI. In this study, the ROI attenuator, which reduces the dose by 80% in the peripheral region, is mounted on a linear stage placed near the x-ray tube. Fluoroscopy is performed using the Microangiographic Fluoroscope (MAF) which is a high-resolution, CCD-based x-ray detector. A stainless-steel stent is selected as the object of interest, and is moved across the FOV and localized using an object-detection algorithm available in the IMAQ Vision package of LabVIEW. The ROI is moved to follow the stent motion. The pixel intensities are equalized in both FOV regions and an adaptive temporal filter dependent on the motion of the object of interest is implemented inside the ROI. With a temporal filter weight of 5% for the current image in the peripheral region, the SNR measured is 47.8. The weights inside the ROI vary between 10% and 33% with a measured SNR of 57.9 and 35.3 when the object is stationary and moving, respectively. This method allows patient dose reduction as well as maintenance of superior image quality in the ROI while tracking the object.

8.
Proc SPIE Int Soc Opt Eng ; 8313: 83130Q, 2012 Feb 23.
Article in English | MEDLINE | ID: mdl-24027618

ABSTRACT

Phantom equivalents of different human anatomical parts are routinely used for imaging system evaluation or dose calculations. The various recommendations on the generic phantom structure given by organizations such as the AAPM, are not always accurate when evaluating a very specific task. When we compared the AAPM head phantom containing 3 mm of aluminum to actual neuro-endovascular image guided interventions (neuro-EIGI) occurring in the Circle of Willis, we found that the system automatic exposure rate control (AERC) significantly underestimated the x-ray parameter selection. To build a more accurate phantom for neuro-EIGI, we reevaluated the amount of aluminum which must be included in the phantom. Human skulls were imaged at different angles, using various angiographic exposures, at kV's relevant to neuro-angiography. An aluminum step wedge was also imaged under identical conditions, and a correlation between the gray values of the imaged skulls and those of the aluminum step thicknesses was established. The average equivalent aluminum thickness for the skull samples for frontal projections in the Circle of Willis region was found to be about 13 mm. The results showed no significant changes in the average equivalent aluminum thickness with kV or mAs variation. When a uniform phantom using 13 mm aluminum and 15 cm acrylic was compared with an anthropomorphic head phantom the x-ray parameters selected by the AERC system were practically identical. These new findings indicate that for this specific task, the amount of aluminum included in the head equivalent must be increased substantially from 3 mm to a value of 13 mm.

9.
Article in English | MEDLINE | ID: mdl-23367291

ABSTRACT

We describe and demonstrate for the first time the use of the novel Multiple Module Multiplexer (MMMIC) for a 2×2 array of new electron multiplying charge coupled device (EMCCD) based x-ray detectors. It is highly desirable for x-ray imaging systems to have larger fields of view (FOV) extensible in two directions yet to still be capable of doing high resolution imaging over regions-of-interest (ROI). The MMMIC achieves these goals by acquiring and multiplexing data from an array of imaging modules thereby enabling a larger FOV, and at the same time allowing high resolution ROI imaging through selection of a subset of modules in the array. MMMIC also supports different binning modes. This paper describes how a specific two stage configuration connecting three identical MMMICs is used to acquire and multiplex data from a 2×2 array of EMCCD based detectors. The first stage contains two MMMICs wherein each MMMIC is getting data from two EMCCD detectors. The multiplexed data from these MMMICs is then forwarded to the second stage MMMIC in the similar fashion. The second stage that has only one MMMIC gives the final 12 bit multiplexed data from four modules. This data is then sent over a high speed Camera Link interface to the image processing computer. X-ray images taken through the 2×2 array of EMCCD based detectors using this two stage configuration of MMMICs are shown successfully demonstrating the concept.


Subject(s)
Radiometry/instrumentation , X-Rays
10.
Article in English | MEDLINE | ID: mdl-23367292

ABSTRACT

We have designed and developed a new solid-state x-ray imaging system that consists of a 2×2 array of electron multiplying charge coupled devices (EMCCDs). This system is intended for fluoroscopic and angiographic medical imaging. The key components are the four 1024 × 1024 pixel EMCCDs with a pixel size of 13 × 13 µm(2). Each EMCCD is bonded to a fiber optic plate (FOP), and optically coupled to a 350 µm thick micro-columnar CsI(TI) scintillator via a 3.22∶1 fiber optic taper (FOT). The detector provides x-ray images of 9 line pairs/mm resolution at 15 frames/sec and real-time live video at 30 frames/sec with binning at a lower resolution, independent of the electronic gain applied to the EMCCD. The total field of view (FOV) of the array is 8.45 cm × 8.45 cm. The system is designed to also provide the ability to do region-of- interest imaging (ROI) by selectively enabling individual modules of the array.


Subject(s)
Diagnostic Imaging/instrumentation , Radiometry/instrumentation , X-Rays
SELECTION OF CITATIONS
SEARCH DETAIL
...