Appl Radiat Isot ; 113: 89-95, 2016 07.
Article in English | MEDLINE | ID: mdl-27152914

ABSTRACT

For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.


Subject(s)
Positron-Emission Tomography/statistics & numerical data , Positron-Emission Tomography/standards , Radioisotopes/analysis , Radiometry/statistics & numerical data , Radiometry/standards , Bromine Radioisotopes/analysis , Calibration , Copper Radioisotopes/analysis , Humans , Iodine Radioisotopes/analysis , Manganese/analysis , Yttrium Radioisotopes/analysis , Zirconium/analysis