Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 49
Filter
1.
J Pharm Biomed Anal ; 31(6): 1089-99, 2003 Apr 10.
Article in English | MEDLINE | ID: mdl-12667925

ABSTRACT

Cassette dosing is a method in which multiple drugs are administered to a single animal at the same time, and the plasma concentrations of the individual compounds are simultaneously determined. This method enables high-throughput rapid screening for pharmacokinetic assessment of new drug candidates. An available gradient method was modified for cassette dosing analysis to attain the advantages of high sensitivity and applicability to a wide range of compounds. However, two problems arose; (1). the time-consuming optimization of mobile phases for each compound group, which limited applicability and (2). the remarkable suppression of ionization by polyethyleneglycol, which is commonly used in intravenous administration. To resolve these problems, a new column switching method was established to attain wider applicability and avoid the ionization suppression. This column switching system is very simple because the trap column and the analytical column are specified and the mobile phase is selected from only two species. Method optimization requires only the selection of the mobile phase and takes only a few hours. About 200 compounds, which were administered as about 50 cassettes, were analyzed using this column switching system. Assay validation of one cassette was carried out, and good accuracy and precision were obtained. About 90% of the compounds could be determined within 20% bias. These results showed that this new column switching system for cassette dosing is accurate enough for the screening of drug candidates and offers wide applicability for various compounds. This system was shown to be very useful for the determination of cassette dosing samples, containing multiple compounds.


Subject(s)
Pharmaceutical Preparations/analysis , Technology, Pharmaceutical/methods , Animals , Chromatography, Liquid/instrumentation , Chromatography, Liquid/methods , Female , Gas Chromatography-Mass Spectrometry/instrumentation , Gas Chromatography-Mass Spectrometry/methods , Male , Mice , Mice, Inbred BALB C , Rats , Rats, Sprague-Dawley , Technology, Pharmaceutical/instrumentation
2.
Biosens Bioelectron ; 16(9-12): 621-9, 2001 Dec.
Article in English | MEDLINE | ID: mdl-11679237

ABSTRACT

A device providing frequent, automatic, and non-invasive glucose measurements for persons with diabetes has been developed: the GlucoWatch biographer. This device extracts glucose through intact skin via reverse iontophoresis where it is detected by an amperometric biosensor. The biographer can provide glucose readings every 20 min for 12 h. The performance of this device was evaluated in two large clinical studies in a controlled clinical environment (n=231), and the home environment (n=124). Accuracy of the biographer was evaluated by comparing the automatic biographer readings to serial finger-stick blood glucose (BG) measurements. Biographer performance was comparable in both environments. Mean difference between biographer and finger-stick measurements was -0.01 and 0.26 mmol l(-1) for the clinical and home environments, respectively. The mean absolute value of the relative difference was 1.06 and 1.18 mmol l(-1) for the same studies. Correlation coefficient (r) between biographer and finger-stick measurements was 0.85 and 0.80 for the two studies. In both studies, over 94% of the biographer readings were in the clinically acceptable A+B region of the Clarke Error Grid. A slight positive bias is observed for the biographer readings at low BG levels. Biographer accuracy is relatively constant over all rates of BG changes, except when BG decreases more than 10 mmol l(-1) h(-1), which occurred for only 0.2% of points in the home environment study. Biographer precision, as measured by CV%, is approx. 10%. Skin irritation, characterized by erythema and edema, was either non-existent or mild in >90% of subjects and resolved in virtually all subjects without treatment in several days.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Diabetes Mellitus/blood , Blood Glucose Self-Monitoring/statistics & numerical data , Diabetes Mellitus/metabolism , Equipment Design , Glucose/metabolism , Humans , Iontophoresis , Skin/metabolism
3.
Diabetes Care ; 24(5): 881-5, 2001 May.
Article in English | MEDLINE | ID: mdl-11347748

ABSTRACT

OBJECTIVE: Hypoglycemia is a common acute complication of diabetes therapy. The GlucoWatch biographer provides frequent and automatic glucose measurements with an adjustable low-glucose alarm. We have analyzed the performance of the biographer low-glucose alarm relative to hypoglycemia as defined by blood glucose < or = 3.9 mmol/l. RESEARCH DESIGN AND METHODS: The analysis was based on 1,091 biographer uses from four clinical trials. which generated 14,487 paired (biographer and blood glucose) readings. RESULTS: The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose < or = 3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose >3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Blood Glucose/analysis , Hypoglycemia/diagnosis , Monitoring, Ambulatory/instrumentation , Automation , Blood Glucose Self-Monitoring/methods , False Negative Reactions , False Positive Reactions , Humans , Hypoglycemia/blood , Hypoglycemia/prevention & control , Hypoglycemic Agents/adverse effects , Monitoring, Ambulatory/methods , ROC Curve , Reproducibility of Results , Sensitivity and Specificity
4.
Jpn J Thorac Cardiovasc Surg ; 49(11): 669-70, 2001 Nov.
Article in English | MEDLINE | ID: mdl-11757340

ABSTRACT

We describe a case of endobronchial metastasis from prostate cancer originally diagnosed as primary bronchogenic carcinoma. We conducted a sleeve lobectomy and determined a final diagnosis using a serumprostate-specific antigen and immunohistochemical studies.


Subject(s)
Bronchial Neoplasms/secondary , Prostatic Neoplasms/pathology , Aged , Biomarkers, Tumor/blood , Bronchial Neoplasms/diagnosis , Carcinoma, Bronchogenic/diagnosis , Diagnosis, Differential , Humans , Male , Prostate-Specific Antigen/blood
5.
Biomaterials ; 21(18): 1837-45, 2000 Sep.
Article in English | MEDLINE | ID: mdl-10919687

ABSTRACT

This study investigated the in vitro degradation of porous poly(DL-lactic-co-glycolic acid) (PLGA) foams during a 20-week period in pH 7.4 phosphate-buffered saline (PBS) at 37 degrees C and their in vivo degradation following implantation in rat mesentery for up to 8 weeks. Three types of PLGA 85 : 15 and three types of 50 : 50 foams were fabricated using a solvent-casting, particulate-leaching technique. The two types had initial salt weight fraction of 80 and 90%, and a salt particle size of 106-150 microm, while the third type had 90% initial weight fraction of salt in the size range 0-53 microm. The porosities of the resulting foams were 0.82, 0.89, and 0.85 for PLGA 85 : 15, and 0.73, 0.87, and 0.84 for PLGA 50 : 50 foams, respectively. The corresponding median pore diameters were 30, 50, and 17 microm for PLGA 85: 15, and 19, 17, and 17 microm for PLGA 50 : 50. The in vitro and in vivo degradation kinetics of PLGA 85: 15 foams were independent of pore morphology with insignificant variation in foam weight, thickness, pore distribution, compressive creep behavior, and morphology during degradation. The in vitro foam half-lives based on the weight average molecular weight were 11.1 +/- 1.8 (80%, 106-150 microm), 12.0 +/- 2.0 (90%, 106-150 microm), and 11.6 +/- 1.3 (90%, 0-53 microm) weeks, similar to the corresponding values of 9.4 +/- 2.2, 14.3 +/- 1.5, and 13.7 +/- 3.3 weeks for in vivo degradation. In contrast, all PLGA 50 : 50 foams exhibited significant change in foam weight, water absorption, and pore distribution after 6-8 weeks of incubation with PBS. The in vitro foam half-lives were 3.3 +/- 0.3 (80%, 106-150 microm), 3.0 +/- 0.3 (90%, 106-150 microm), and 3.2 +/- 0.1 (90%, 0-53 microm) weeks, and the corresponding in vivo half-lives were 1.9 micro 0.1, 2.2 +/- 0.2, and 2.4 +/- 0.2 weeks. The significantly shorter half-lives of PLGA 50: 50 compared to 85: 15 foams indicated their faster degradation both in vitro and in vivo. In addition, PLGA 50: 50 foams exhibited significantly faster degradation in vivo as compared to in vitro conditions due to an autocatalytic effect of the accumulated acidic degradation products in the medium surrounding the implants. These results suggest that the polymer composition and environmental conditions have significant effects on the degradation rate of porous PLGA foams.


Subject(s)
Biocompatible Materials/chemistry , Biocompatible Materials/pharmacokinetics , Lactic Acid/chemistry , Lactic Acid/pharmacokinetics , Polyglycolic Acid/chemistry , Polyglycolic Acid/pharmacokinetics , Polymers/chemistry , Polymers/pharmacokinetics , Animals , Biodegradation, Environmental , Half-Life , Hot Temperature , Hydrogen-Ion Concentration , Kinetics , Mercury , Mesentery/cytology , Mesentery/pathology , Polylactic Acid-Polyglycolic Acid Copolymer , Prostheses and Implants , Rats , Thermodynamics
6.
Biomaterials ; 21(15): 1595-605, 2000 Aug.
Article in English | MEDLINE | ID: mdl-10885732

ABSTRACT

This study investigated the in vitro degradation of porous poly(L-lactic acid) (PLLA) foams during a 46-week period in pH 7.4 phosphate-buffered saline at 37 degrees C. Four types of PLLA foams were fabricated using a solvent-casting, particulate-leaching technique. The three types had initial salt weight fraction of 70, 80, and 90%, and a salt particle size of 106-150 microm, while the fourth type had 90% initial weight fraction of salt in the size range 0-53 microm. The porosities of the resulting foams were 0.67, 0.79, 0.91, and 0.84, respectively. The corresponding median pore diameters were 33, 52, 91, and 34 microm. The macroscopic degradation of PLLA foams was independent of pore morphology with insignificant variation in foam weight, thickness, pore distribution, compressive creep behavior, and morphology during degradation. However, decrease in melting temperature and slight increase in crystallinity were observed at the end of degradation. The foam half-lives based on the weight average molecular weight were 11.6+/-0.7 (70%, 106-150 microm), 15.8+/-1.2 (80%, 106-150 microm), 21.5+/-1.5 (90%, 106-150 microm), and 43.0+/-2.7 (90%, 0-53 microm) weeks. The thicker pore walls of foams prepared with 70 or 80% salt weight fraction as compared to those with 90% salt weight fraction contributed to an autocatalytic effect resulting in faster foam degradation. Also, the increased pore surface/volume ratio of foams prepared with salt in the range 0-53 microm enhanced the release of degradation products thus diminishing the autocatalytic effect and resulting in slower foam degradation compared to those with salt in the range 106-150 microm. Formation and release of crystalline PLLA particulates occurred for foams fabricated with 90% salt weight fraction at early stages of degradation. These results suggest that the degradation rate of porous foams can be engineered by varying the pore wall thickness and pore surface/volume ratio.


Subject(s)
Biocompatible Materials/chemistry , Lactic Acid/chemistry , Polymers/chemistry , Calorimetry, Differential Scanning , Chromatography, Gel , Drug Delivery Systems , Drug Stability , Microscopy, Electron, Scanning , Molecular Weight , Polyesters
7.
Ann Med ; 32(9): 632-41, 2000 Dec.
Article in English | MEDLINE | ID: mdl-11209971

ABSTRACT

The GlucoWatch (Cygnus, Inc, Redwood City, CA, USA) biographer provides automatic, frequent and noninvasive blood glucose measurements for up to 12 h. The device extracts glucose through intact skin where it is measured by an amperometric biosensor. Clinical trials in a variety of environments have shown that the biographer provides accurate and precise glucose measurements when compared with serial fingerstick blood glucose measurements. Mean difference between these measurements was 0.26 mmol/L in the home environment (r = 0.80). Over 94% of biographer readings were in the clinically acceptable A+B region of the Clarke Error Grid. A slight positive bias is observed for the biographer readings at low glucose levels. Biographer precision, as measured by coefficient of variation (CV)%, is approximately 10%. The low glucose alert function of the biographer was able to detect up to 75% of hypoglycaemic episodes with a low false alert level. Skin irritation, characterized by erythema and oedema was either nonexistent or mild in over 87% of subjects and resolved in virtually all subjects without treatment in several days. The GlucoWatch biographer has been shown to be a safe and effective method to track glucose level trends and patterns, which should enable improved glycaemic control for many patients.


Subject(s)
Blood Glucose Self-Monitoring , Diabetes Mellitus/blood , Analysis of Variance , Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Clinical Trials as Topic , Equipment Design , Female , Humans , Hypoglycemia/diagnosis , Male , Middle Aged
8.
Diabetes Technol Ther ; 2(2): 199-207, 2000.
Article in English | MEDLINE | ID: mdl-11469259

ABSTRACT

BACKGROUND: Improved glycemic control significantly reduces long-term microvascular complications of diabetes mellitus associated with chronic hyperglycemia. The GlucoWatch biographer is designed to facilitate intensive diabetes management by providing automatic, frequent, and noninvasive glucose readings up to three times per hour for as long as 12 hours. METHODS: The device extracts glucose through intact skin using reverse iontophoresis and measures the extracted glucose with an electrochemical biosensor. A clinical trial was performed to assess the effect of acetaminophen, a potential interference for traditional blood glucose meters, on the accuracy of the GlucoWatch biographer in adult subjects with diabetes (n = 18). One thousand milligram doses of acetaminophen were administered to subjects in two groups: one to achieve Cmax (peak acetominophen concentration) at the time of biographer calibration and the other to achieve Cmax during the measurement period. The biographer readings were compared to serial fingerstick blood glucose measurements. RESULTS: Time profiles over 9 hours show close tracking of the biographer glucose results with fingerstick blood glucose measurements for all groups. The mean difference between the two measurements is between 8 and 12 mg/dL for all groups. The mean absolute value of the relative difference is less than 20%, and more than 93% of the points were in the clinically acceptable (A+B) region of the Clarke Error Grid. No statistically significant differences were found for any accuracy measurement across all groups. CONCLUSIONS: The GlucoWatch Biographer provides frequent measurements of glucose over a 12-hour period with high accuracy. No effect of therapeutic dosage of acetaminophen on the accuracy of the glucose readings was found.


Subject(s)
Acetaminophen/pharmacology , Blood Glucose Self-Monitoring/methods , Blood Glucose/analysis , Monitoring, Ambulatory/instrumentation , Monitoring, Ambulatory/methods , Adult , Analysis of Variance , Automation , Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Calibration , Electrochemistry , Equipment Design , Humans , Reproducibility of Results , Time Factors , United States , White People
9.
JAMA ; 282(19): 1839-44, 1999 Nov 17.
Article in English | MEDLINE | ID: mdl-10573275

ABSTRACT

CONTEXT: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management. OBJECTIVE: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements. DESIGN: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study. SETTING: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States. PARTICIPANTS: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin. MAIN OUTCOME MEASURES: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid. RESULTS: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis. CONCLUSION: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.


Subject(s)
Biosensing Techniques , Blood Glucose Self-Monitoring/instrumentation , Adult , Blood Glucose/analysis , Diabetes Mellitus/prevention & control , Female , Humans , Male
10.
Diabetes Care ; 22(10): 1708-14, 1999 Oct.
Article in English | MEDLINE | ID: mdl-10526740

ABSTRACT

OBJECTIVE: The purpose of this study was to compare measurements of glucose obtained via iontophoretic extraction with the GlucoWatch automatic glucose biographer (Cygnus, Inc., Redwood City, CA) with capillary blood glucose values that were determined 1) in a controlled outpatient clinic setting and 2) in a home setting. RESEARCH DESIGN AND METHODS: There were 76 GlucoWatch biographers used on 28 different young adults (21 women and 7 men) with type 1 diabetes (age 30.9 +/- 6.9 years and duration of diabetes 18.4 +/- 8.1 years [mean +/- SD]) in a controlled outpatient clinic setting. Some subjects participated on multiple days. Subjects wore two GlucoWatch biographers, each on the forearm (ventral aspect). Comparisons were made to HemoCue blood glucose analyzer (Aktiebolgat Leo, Helsingborg, Sweden) capillary blood glucose measurements. In addition, GlucoWatch biographers (one each day for 3 consecutive days) were used by 12 subjects (8 women, 4 men) in a home setting. Comparisons were made to capillary blood glucose values determined using the One Touch Profile meter (Johnson & Johnson, New Brunswick, NJ). RESULTS: GlucoWatch biographer glucose values correlated well with capillary blood glucose values determined using the HemoCue analyzer in the clinic setting (r = 0.90, 1,554 paired data points) and using the One Touch Profile meter in the home setting (r = 0.85, 204 paired data points). When 36 subjects wore two biographers simultaneously, the correlation between the two biographers was r = 0.94. The error grid analysis demonstrated that > 96% of biographer glucose values determined in the clinic or home setting were in the clinically acceptable A and B regions. CONCLUSIONS: This study confirms the accuracy and precision of glucose values as determined using the GlucoWatch biographer in clinic and home settings.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Blood Glucose Self-Monitoring/methods , Diabetes Mellitus, Type 1/blood , Adult , Capillaries , Equipment Design , Female , Fingers/blood supply , Humans , Male , Regression Analysis , Reproducibility of Results
13.
Noshuyo Byori ; 13(2): 149-55, 1996 Nov.
Article in English | MEDLINE | ID: mdl-8958522

ABSTRACT

A 33-year-old woman had an undifferentiated tumor originating in the cerebral dysgenetic lesion resembling fibrous cartilage. She had a headache, vomiting, late-onset epilepsy and left hemiparesis. The lesion was located in the right temporal lobe on computed tomographic (CT) scan. It was totally resected and only local irradiation was performed postoperatively. Normal cortical architectures were lost in the resected specimens. Straight or curved fasciculi composed of fine collagen fibers were distributed in parallel and perpendicular to the cortical surface in the mildly eosinophilic hyaline matrix. Hypertrophic astrocytes were scattered with low cellularity in these abnormal cortices. Clusters of tumor cells were observed in a few areas. The tumor cells, having oval and relatively vesicular nuclei with a few prominent nucleoli and basophilic well-circumscribed narrow cytoplasm, had proliferated diffusely with a cobblestone appearance. Immunohistochemical and electron microscopic investigations demonstrated no evidence of specific differentiation, either. There were 14.5 mitotic figures/high power field on average and most nuclei of the tumor cells were strongly positive for proliferating cell nuclear antigen (PCNA). Moreover, subarachnoid dissemination of the tumor cells were apparent in a few areas. Nevertheless the patient returned to work and no recurrence was observed for 10 years postoperatively. We concluded that neoplastic transformation occurred de novo in the dysgenetic cortex.


Subject(s)
Brain Neoplasms/pathology , Temporal Lobe/pathology , Adult , Brain Neoplasms/diagnosis , Brain Neoplasms/surgery , Cartilage/pathology , Cell Nucleus/ultrastructure , Cytoplasm/ultrastructure , Diagnosis, Differential , Female , Gadolinium DTPA , Humans , Immunohistochemistry , Magnetic Resonance Imaging , Microscopy, Electron , Organometallic Compounds , Pentetic Acid/analogs & derivatives , Proliferating Cell Nuclear Antigen/analysis , Tomography, X-Ray Computed
14.
Pharm Res ; 12(12): 1869-73, 1995 Dec.
Article in English | MEDLINE | ID: mdl-8786958

ABSTRACT

PURPOSE: To demonstrate that "reverse iontophoresis" can be used to noninvasively obtain information about systemic glucose levels in vivo in humans. METHODS: The passage of current across the skin in vivo drives ions into the tissue, from the electrode chambers positioned on the skin surface, and simultaneously pulls ions from the body in the opposite direction. Because of the net negative charge on the skin, under normal conditions, the membrane is permselective to cations, and a potential gradient also results, therefore, in electroosmotic convection of solvent in the direction of counterion flow (i.e., from anode to cathode). Thus, it is also possible to enhance the transport of polar, yet uncharged, species using iontophoresis. In an earlier study, the in vitro extraction of glucose, by "reverse iontophoresis" was established, and extension of the approach to an in vivo model was indicated. The idea has therefore been further explored in vivo in humans. RESULTS: Using small, simple, prototypical electrode chambers, attached to the ventral forearm surface, direct current iontophoresis at 0.25 mA/cm2 for periods of up to 1 hour, and a sensitive analytical procedure to measure the quantities of glucose extracted, it has been shown that iontophoretic sampling of glucose is feasible. However, the shorter periods (15 minutes or less) of extraction considered yield results which are "contaminated" (it is believed) by glucose that is a product of lipid metabolism within the skin. While this material is expected to complicate the initial calibration of the approach, the problem is effectively resolved within one hour, by which time the glucose arriving in the electrode chambers on the skin surface is expected to directly reflect the subcutaneous tissue concentration. CONCLUSIONS: Based upon these initial observations, further investigation can now be directed towards optimization of electroosmotic flow and sampling time, improved reproducibility and the development of a practical assay methodology.


Subject(s)
Blood Glucose/metabolism , Glucose/metabolism , Iontophoresis/methods , Humans , Kinetics
15.
Nat Med ; 1(11): 1198-201, 1995 Nov.
Article in English | MEDLINE | ID: mdl-7584995

ABSTRACT

Results from the Diabetes Care and Complications Trial show that tight blood glucose control significantly reduces the long-term complications of diabetes mellitus. In that study, frequent self-testing of glucose and insulin administration resulted in a significant reduction in long-term complications. This protocol, however, also resulted in a threefold increase in the frequency of hypoglycaemic incidents. Currently, self-testing requires a drop of blood for each measurement. The pain and inconvenience of self-testing, along with the fear and danger of hypoglycaemia has led to poor patient acceptance of a tight control regimen, despite the clear long-term advantages. A continuously worn, noninvasive method to periodically measure glucose would provide a convenient and comfortable means of frequent self-testing. A continuously worn device could also alert the user of low glucose levels, thereby reducing the incidence of hypoglycaemia. Guy et al. demonstrated a noninvasive method to transport glucose through the skin using low-level electrical current. To provide a quantitative measurement, the flux of glucose extracted across the skin must correlate with serum glucose in a predictive manner. The results presented here show a quantitative relationship between serum and transdermally extracted glucose in diabetics.


Subject(s)
Blood Glucose Self-Monitoring/methods , Blood Glucose/analysis , Diabetes Mellitus/blood , Iontophoresis/methods , Adult , Diabetes Mellitus, Type 1/blood , Diabetes Mellitus, Type 2/blood , Female , Humans , Hyperglycemia/prevention & control , Hypoglycemia/prevention & control , Male , Middle Aged , Reproducibility of Results , Skin/metabolism
16.
J Biomed Mater Res ; 28(12): 1465-75, 1994 Dec.
Article in English | MEDLINE | ID: mdl-7876286

ABSTRACT

Studies investigating the erosion mechanism of the newly developed poly (fatty acid dimer: sebacic acid) polyanhydride (p:[FAD:SA]) are described. The overall erosion of different monomer compositions of p(FAD:SA) copolymers was examined to determine whether and to what extent copolymer properties affected polymer erosion. Increasing the hydrophobic monomer (FAD) content up to 50 wt% in the copolymer resulted in longer erosion, whereas further increases up to 70 wt% decreased the erosion period. Polymer crystallinity depended on copolymer FAD content. Copolymer degradation was studied by examining anhydride bond hydrolysis using infrared spectroscopy. Much faster hydrolysis was found in p(FAD:SA) 70:30 compared with more crystalline copolymers of higher SA content. Light microscopy indicates the presence of an erosion zone, a distinct area where mass loss occurs. This erosion zone moves from the outside toward the interior of the polymer matrix. It plays an important role in erosion because any water or monomer must diffuse through this eroded layer.


Subject(s)
Decanoic Acids/chemistry , Polymers/chemistry , Stearic Acids/chemistry , Biodegradation, Environmental , Crystallization , Decanoic Acids/metabolism , Hydrolysis , Polymers/metabolism , Stearic Acids/metabolism , Time Factors , Water/metabolism
17.
Pharm Res ; 11(12): 1809-14, 1994 Dec.
Article in English | MEDLINE | ID: mdl-7899247

ABSTRACT

Electroporation, the creation of transient, enhanced membrane permeability using short duration (microseconds to millisecond) electrical pulses, can be used to increase transdermal drug delivery. The effect of an (electroporative) electric pulse (1000 V, tau = 5 msec) on the iontophoretic transport of LHRH through human skin was studied in vitro. Fluxes achieved with and without a pulse under different current densities (0- 4 mA/cm2) were compared. The results indicated that the application of a single pulse prior to iontophoresis consistently yielded higher fluxes (5-10 times the corresponding iontophoretic flux). For example, at 0.5 mA/cm2 fluxes were 0.27 +/- 0.08 and 1.62 +/- 0.05 micrograms/hr/cm2 without and with the pulse, respectively. At each current density studied, the LHRH flux decreased after iontophoresis, approaching pre-treatment values. The results show that electroporation can significantly and reversibly increase the flux of LHRH through human skin. These results also indicate the therapeutic utility of using electroporation for enhanced transdermal transport.


Subject(s)
Electroporation , Gonadotropin-Releasing Hormone/administration & dosage , Iontophoresis , Skin/metabolism , Amino Acid Sequence , Gonadotropin-Releasing Hormone/pharmacokinetics , Humans , Molecular Sequence Data
18.
Nihon Kyobu Geka Gakkai Zasshi ; 42(10): 1944-8, 1994 Oct.
Article in Japanese | MEDLINE | ID: mdl-7798713

ABSTRACT

A 73-year-old man with a right aortic arch was admitted because of stridor and dyspnea. We diagnosed a cervico-mediastinal cyst associated with bilateral recurrent laryngeal nerve palsy. Median sternotomy revealed a cystic tumor located in the esophageal wall. Microscopic examination of the resected cyst showed cartilage, smooth muscle and mucous gland, so the final diagnosis was bronchogenic cyst. The bilateral recurrent laryngeal nerve palsy was caused by chronic pressure from the cyst. Mediastinal cysts should be found and treated as early as possible, because these cysts may sometimes cause fatal complications.


Subject(s)
Bronchogenic Cyst/surgery , Esophageal Diseases/surgery , Aged , Bronchogenic Cyst/diagnosis , Esophageal Diseases/diagnosis , Humans , Male
19.
J Biomed Mater Res ; 28(3): 387-95, 1994 Mar.
Article in English | MEDLINE | ID: mdl-8077254

ABSTRACT

Intracranial studies to analyze the degradation kinetics of the bioerodible polymer poly[bis(p-carboxyphenoxy)propane-sebacic acid] [p(CPP-SA) 20:80] copolymer wafers were conducted in a rat model. Rats were separated into four groups: those receiving 1) polymer, 2) polymer loaded with the chemotherapeutic agent BCNU, 3) drug-loaded polymer with previous tumor implantation, and 4) polymer and an absorbable hemostatic material. A polymer wafer was surgically implanted into the brain of each animal. Residual polymer was harvested at varying times for chromatographic analysis. In vitro effects of pH, mixing, and water availability on degradation were also studied. The results of in vitro and in vivo studies were compared to understand the behavior of polymers in a clinical setting. We found that degradation of p(CPP-SA) initially occurred more slowly in vivo than in vitro. The presence of BCNU, tumor, and absorbable hemostatic material did not affect the ultimate time of polymer degradation in vivo, and the intrinsic polymer degradation time of 1 mm thick p(CPP-SA) 20:80 disks in vivo was 6-8 weeks.


Subject(s)
Brain Neoplasms/drug therapy , Decanoic Acids/pharmacokinetics , Drug Carriers/pharmacokinetics , Gliosarcoma/drug therapy , Polyesters/pharmacokinetics , Animals , Biodegradation, Environmental , Carmustine/administration & dosage , Carmustine/therapeutic use , Cellulose, Oxidized , Delayed-Action Preparations , Drug Evaluation, Preclinical , Drug Implants , Foreign-Body Reaction/etiology , Inflammation , Male , Rats , Rats, Inbred F344
20.
Acta Neurochir Suppl ; 62: 43-6, 1994.
Article in English | MEDLINE | ID: mdl-7717134

ABSTRACT

Clinical characteristics of radiosensitive craniopharyngiomas and histologically identical tumours were re-evaluated from among 53 patients. There were 9 squamous cell type and 3 mixed type tumours. Early effects of radiosurgery for two recent cases are reported. Radiosurgery may have an important role to play in the treatment of craniopharyngiomas, especially of the squamous cell type.


Subject(s)
Brain Neoplasms/surgery , Brain/surgery , Craniopharyngioma/surgery , Radiosurgery , Adolescent , Adult , Aged , Brain/pathology , Brain Neoplasms/pathology , Child , Child, Preschool , Craniopharyngioma/pathology , Female , Humans , Male , Neoplasm Recurrence, Local , Radiation Dosage
SELECTION OF CITATIONS
SEARCH DETAIL
...