J Artif Organs ; 24(3): 412-416, 2021 Sep.
Article in English | MEDLINE | ID: mdl-33200301

ABSTRACT

One of the often-used methods for in vitro evaluation of the blood compatibility of hemodialysis membranes is the circulation of human blood through a miniaturized hemodialyzer. The use of a rather small amount of human blood in its evaluation is one advantage of this method. However, because it is manufactured by a different process than actual ones, a miniaturized hemodialyzer membrane cannot always preserve the properties of actual hemodialyzers. To address this problem, we established a new experimental method that uses a relatively small amount of human blood and actual dialyzers. In this method, a test hemodialyzer and a control hemodialyzer filled with human blood obtained from the same donor is slowly rotated to prevent spontaneous blood cell sedimentation for 4 h at 37 °C. By use of this method, we were able to compare blood compatibility between a polysulfone (PS) membrane and a vitamin E (VE)-bonded PS membrane in terms of their relative antithrombotic, antioxidative, and anti-inflammatory properties. Consistent with many previous reports, the results clearly showed that compared with the PS membrane, VE-bonded PS membrane is more blood compatible. These findings suggest that our method is applicable, at least to in vitro blood compatibility evaluation of PS type dialysis membranes.


Subject(s)
Kidneys, Artificial , Antioxidants , Humans , Membranes, Artificial , Renal Dialysis , Vitamin E