Microscopy (Oxf) ; 73(4): 358-366, 2024 Jul 30.
Article in English | MEDLINE | ID: mdl-38412273

ABSTRACT

This paper describes the development of a gas chromatography-quadrupole mass spectrometry system attached to a differential-pumping-type environmental cell of the reaction science high-voltage electron microscopy instrument at Nagoya University to distinguish unambiguously between different gas species with the same mass-to-charge ratio. Several model experiments were used to verify the efficacy of the newly proposed system, confirming its ability to analyse the atomic-level structural changes during heterogeneous catalysts and the associated gas-reaction kinetics simultaneously, providing new insights into operando measurements in the field of environmental transmission electron microscopy. Graphical Abstract.