Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 5 de 5
Filter
Add more filters










Database
Language
Publication year range
1.
Klin Lab Diagn ; (3): 9-10, 2002 Mar.
Article in Russian | MEDLINE | ID: mdl-11980150

ABSTRACT

Miniature analyzer Glucophot BP-"M", intended for express measurements of glucose concentrations in capillary blood, was tested under clinical conditions. The reference method was the hexokinase method for glucose measurements on a Spectrum 11 biochemical analyzer (Abbott); in addition, glucose measurements were carried out on a One Touch-11 glucometer (Johnson-Johnson). The results evidence that the reproducibility and correctness of glucose analysis by the Glucophot BP-"M" analyzer both in health and disease met the requirements to laboratory measurements performed by "dry" chemical methods and demonstrated its high analytical reliability.


Subject(s)
Blood Glucose/analysis , Diabetes Mellitus/diagnosis , Humans , Photometry , Reagent Kits, Diagnostic , Sensitivity and Specificity