Med Phys ; 35(12): 5558-64, 2008 Dec.
Article in English | MEDLINE | ID: mdl-19175113

ABSTRACT

High dose rate brachytherapy (HDR-BT) is one of the many modalities for prostate cancer treatment. Due to the nature of HDR-BT, in vivo dosimetry is feasible and can be used to verify consistent dose delivery. In order to validate a dose verification system for HDR-BT prostate cancer treatment, a radiophotoluminescent glass dosimeter (RPLGD) was used and the measurements were compared with those from a thermoluminescent dosimeter. The RPLGD shows many advantages in HDR-BT dose measurement, such as repeatability, stability, and small effective size. These advantages make the RPLGD a superior option for use as a dosimeter in HDR-BT. The results described here show that the difference between the measured dose and the treatment planned dose is less than 5%. A Monte Carlo simulation for the dose was performed using Monte Carlo N -particle to investigate position error. This study concludes that the RPLGD is a promising and reliable dosimeter for HDR-BT in vivo dosimetry with clinically acceptable accuracy.


Subject(s)
Brachytherapy/methods , Glass , Prostatic Neoplasms/radiotherapy , Thermoluminescent Dosimetry/methods , Dose-Response Relationship, Radiation , Equipment Design , Humans , Male , Monte Carlo Method , Phantoms, Imaging , Radiation Dosage , Radiometry/methods , Radiotherapy Dosage , Reproducibility of Results , Water