Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Biosens Bioelectron ; 10(6-7): 527-32, 1995.
Article in English | MEDLINE | ID: mdl-7612205

ABSTRACT

Integrated thin film biosensors were developed for the simultaneous measurement of L-glutamine and L-glutamate in a mu-flow cell. Due to a novel glutaminase with an activity optimum in the neutral pH range, direct monitoring of glutamine in a mammalian cell culture medium could be performed. The glutamine bienzyme sensor was prepared by co-immobilization of glutaminase with glutamate oxidase within a photopatterned poly(2-hydroxyethyl methacrylate) (pHEMA) hydrogel membrane. The sensor response was linear in the concentration range of 50 mumol to 10 mmol glutamine/l. Additionally, a glutamate biosensor was integrated on the sensor chip for difference measurement of possible glutamate interferences. The sensor-chip could be used for at least 300 measurements without any alteration in the performance of its sensors. A new sensor-chip with an integrated flow cell provided the possibility of simultaneous measurement of four different parameters at a cell volume of 1 microliter. In order to complete the microsystem, and in order to obtain a "lab on chip", a battery operated surface mounted device (SMD) potentiostat was developed.


Subject(s)
Biosensing Techniques , Glutamic Acid/analysis , Glutamine/analysis , Calibration , Hydrogen-Ion Concentration
SELECTION OF CITATIONS
SEARCH DETAIL
...