1.
J Biochem Biophys Methods
; 11(4-5): 277-86, 1985 Oct.
Article
in English
| MEDLINE
| ID: mdl-4067175
ABSTRACT
A slit-type mixer specially designed for stopped-flow X-ray scattering was constructed. It mixes two solutions at a time, allowing a large-sized window to be used immediately after mixing. Dead time of the mixer was 13 ms with 3.5 kg/cm2 of pressure at the piston head. It has been successfully used for studying the tetramer-dimer interconversion of phosphorylase a. The applicability of this mixer to other detection methods, far-UV circular dichroism, light scattering, EXAFS and fluorescence is also indicated.