Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Biochem Biophys Methods ; 11(4-5): 277-86, 1985 Oct.
Article in English | MEDLINE | ID: mdl-4067175

ABSTRACT

A slit-type mixer specially designed for stopped-flow X-ray scattering was constructed. It mixes two solutions at a time, allowing a large-sized window to be used immediately after mixing. Dead time of the mixer was 13 ms with 3.5 kg/cm2 of pressure at the piston head. It has been successfully used for studying the tetramer-dimer interconversion of phosphorylase a. The applicability of this mixer to other detection methods, far-UV circular dichroism, light scattering, EXAFS and fluorescence is also indicated.


Subject(s)
Scattering, Radiation/instrumentation , 2,6-Dichloroindophenol , Air , Ascorbic Acid , Equipment Design , Oxidation-Reduction , Pressure , X-Rays