ABSTRACT

In the present paper, the long noses of Pinocchio, Tengu and Trolls are used to measure, respectively, radon, thoron, and their decay products both by track-etch detectors and by Geiger-Müller (G-M) counters. Just recently, four new passive samplers (termed quatrefoil) have been developed which greatly simplify the detection of all airborne radionuclides by using either passive or real-time detectors. In particular, surface-deposited radon (thoron) decay products are sampled by films with large area and small surface density (0.1-1 mg cm(-2)). Once exposed, these films are stacked together for their detection by a pancake G-M counter. For the measurements of radon and thoron in soil, 25-cm-long tubes with sampling films along their internal surfaces can be successfully used. Once exposed, these films can be counted by a pancake G-M for the selective measurement of radon and thoron.


Subject(s)
Air Pollution, Indoor/analysis , Radiation Monitoring/instrumentation , Radon Daughters/analysis , Radon/analysis , Air Pollutants, Radioactive , Environmental Exposure , Equipment Design , Housing , Humans , Italy , Japan , Models, Statistical , Radiation Monitoring/methods , Radiometry , Risk