Appl Radiat Isot ; 141: 246-249, 2018 Nov.
Article in English | MEDLINE | ID: mdl-29843989

ABSTRACT

Our research group has developed a thermoluminescence dosimeter (TLD) based on pellets of LiF:Mg,Ti mixed with polytetrafluoroethylene (LiF:Mg,Ti +PTFE). This TLD can be used as a personal dosimeter. Extensive type testing, carried out with reference to the International Electrotechnical Commission (IEC) Standard, were performed for the purpose of accepting the LiF:Mg,Ti+PTFE as a personal TL dosimeter. Tests performed include repeatability, batch homogeneity, linearity, detection threshold, and light sensitivity. Results showed that locally made LiF:Mg,Ti+PTFE TLDs met all the standard requirements.


Subject(s)
Thermoluminescent Dosimetry/instrumentation , Fluorides , Humans , Lithium Compounds , Magnesium , Polytetrafluoroethylene , Radiation Dosimeters/standards , Radiation Dosimeters/statistics & numerical data , Reproducibility of Results , Thermoluminescent Dosimetry/standards , Thermoluminescent Dosimetry/statistics & numerical data , Titanium