Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Med Dosim ; 39(4): 354-9, 2014.
Article in English | MEDLINE | ID: mdl-25151596

ABSTRACT

Use of a patient test dose before single-fraction total body irradiation (TBI) allows review of in vivo dosimetry and modification of the main treatment setup. However, use of computed tomography (CT) planning and online in vivo dosimetry may reduce the need for this additional step. Patients were treated using a supine CT-planned extended source-to-surface distance (SSD) technique with lead compensators and bolus. In vivo dosimetry was performed using thermoluminescent dosimeters (TLDs) and diodes at 10 representative anatomical locations, for both a 0.1-Gy test dose and the treatment dose. In total, 28 patients were treated between April 2007 and July 2013, with changes made in 10 cases (36%) following test dose results. Overall, 98.1% of measured in vivo treatment doses were within 10% of the prescribed dose, compared with 97.0% of test dose readings. Changes made following the test dose could have been applied during the single-fraction treatment itself, assuming that the dose was delivered in subportions and online in vivo dosimetry was available for all clinically important anatomical sites. This alleviates the need for a test dose, saving considerable time and resources.


Subject(s)
Dose Fractionation, Radiation , Neoplasms/radiotherapy , Patient Positioning/methods , Radiation Protection/methods , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Whole-Body Irradiation/methods , Humans , Online Systems , Radiometry/instrumentation
SELECTION OF CITATIONS
SEARCH DETAIL
...