1.
Rep Pract Oncol Radiother ; 23(4): 242-250, 2018.
Article in English | MEDLINE | ID: mdl-29991928

ABSTRACT

AIM: This study aims at examining absolute dose verification of step-and-shoot intensity modulated radiation treatment (IMRT) of prostate and brain patients by use of ion chambers of two different volumes and thermoluminescent detectors (TLD). BACKGROUND: The volume of the ion chamber (IC) is very important for absolute dose verification of IMRT plans since the IC has a volume average effect. With TLD detectors absolute dose verification can be done measuring the dose of multiple points simultaneously. MATERIALS AND METHODS: Ion chambers FC65-P of volume 0.65 cc and semiflex of volume 0.125 cc as well as TLDs were used to measure the central axis absolute dose of IMRT quality assurance (QA) plans. The results were compared with doses calculated by a treatment planning system (TPS). The absolute doses of off axis points located 2 cm and 4 cm away from the isocenter were measured with TLDs. RESULTS: The measurements of the 0.125 cc ion chamber were found to be closer to TPS calculations compared to the 0.65 cc ion chamber, for both patient groups. For both groups the root mean square (RMS) differences between doses of the TPS and the TLD detectors are within 3.0% for the central axis and points 2 cm away from the isocenter of each axis. Larger deviations were found at the field edges, which have steep dose gradient. CONCLUSIONS: The 0.125 cc ion chamber measures the absolute dose of the isocenter more accurately compared to the 0.65 cc chamber. TLDs have good accuracy (within 3.0%) for absolute dose measurements of in-field points.

SELECTION OF CITATIONS
SEARCH DETAIL
...