Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 10 de 10
Filter
Add more filters










Publication year range
1.
PLoS One ; 19(2): e0296919, 2024.
Article in English | MEDLINE | ID: mdl-38421998

ABSTRACT

Uncemented acetabular shell primary stability is essential for optimal clinical outcomes. Push-out testing, rotation testing, and lever-out testing are major evaluation methods of primary stability between the shell and bone. However, these test methods do not consider shell loads during daily activity and shell installation angle. This study proposes a novel evaluation method of acetabular shell primary stability considering load during level walking and acetabular installation angles such as inclination and anteversion. To achieve this, a novel primary stability test apparatus was designed with a shell position of 40° acetabular inclination and 20° anteversion. The vertical load, corresponding to walking load, was set to 3 kN according to ISO 14242-1, which is the wear test standard for artificial hip joints. The vertical load was applied by an air cylinder controlled by a pressure-type electro-pneumatic proportional valve, with the vertical load value monitored by a load cell. Torque was measured when angular displacement was applied in the direction of extension during the application of vertical load. For comparison, we also measured torque using the traditional lever-out test. The novel primary stability test yielded significantly higher primary stabilities; 5.4 times greater than the lever-out test results. The novel primary stability test failure mode was more similar to the clinical failure than the traditional lever-out test. It is suggested that this novel primary stability test method, applying physiological walking loads and extension motions to the acetabular shell, better reflects in vivo primary stability than the traditional lever-out test.


Subject(s)
Hip Prosthesis , Walking , Humans , Activities of Daily Living , Acetabulum , Catheters
2.
Biomed Mater Eng ; 33(4): 315-324, 2022.
Article in English | MEDLINE | ID: mdl-35180102

ABSTRACT

BACKGROUND: There is currently no methodology for evaluating the accuracy of ablation in ablation therapy, and thus normal cells in the surrounding area can be damaged, possibly leading to complications. OBJECTIVE: The aim of this study was to distinguish heat-treated dead cells from viable cells using the electrical impedance-to-frequency ratio as an evaluation index. METHODS: Rat heart striated myocytes were cultured in a monolayer on collagen-coated microelectrodes placed in the center of an electrode-loaded chamber. The cells in the chamber were killed by heat treatment for 5 minutes at 50 °C, and the frequency response of the cell impedance was measured before and after heat treatment. The frequency of the input current was varied from 10 to 100 kHz. The measured electrical impedance at each frequency was divided by the value at 100 kHz, and we refer to the resulting values as the impedance ratio. RESULTS: The impedance ratio was high at low frequencies and low at high frequencies. Furthermore, the impedance ratio was lower at lower frequencies after heat treatment than before heat treatment. CONCLUSIONS: The electrical impedance ratio can be used to distinguish viable and dead cells after heat treatment.


Subject(s)
Hot Temperature , Electric Impedance , Microelectrodes
3.
Proc Inst Mech Eng H ; 235(1): 65-72, 2021 Jan.
Article in English | MEDLINE | ID: mdl-32996400

ABSTRACT

Excellent primary stability of uncemented acetabular shells is essential to obtain successful clinical outcomes. However, in the case of developmental dysplasia of the hip (DDH), aseptic loosening may be induced by instability due to a decrease of the contact area between the acetabular shell and host bone. The aim of this study was to assess the primary stability of two commercially-available acetabular shells, hemispherical and hemielliptical, in normal and DDH models. Synthetic bone was reamed using appropriate surgical reamers for each reaming condition (normal acetabular model). The normal acetabular model was also cut diagonally at 40° to create a dysplasia model. Stability of the acetabular components was evaluated by the lever-out test. In the normal acetabular model conditions, the maximum primary stabilities of hemispherical and hemielliptical shells were observed in the 1-mm under- and 1-mm over-reamed conditions, respectively, and the resulting stabilities were comparable. The lateral defect in the dysplasia model had an adverse effect on the primary stabilities of the two designs. The lever-out moment of the hemielliptical acetabular shell was 1.4 times greater than that of the hemispherical acetabular shell in the dysplasia model. The hemispherical shell is useful for the normal acetabular condition, and the hemielliptical shell for the severe dysplasia condition, in the context of primary stability.


Subject(s)
Arthroplasty, Replacement, Hip , Developmental Dysplasia of the Hip , Hip Prosthesis , Acetabulum/surgery , Humans , Reoperation
4.
J Orthop Surg Res ; 12(1): 34, 2017 Feb 23.
Article in English | MEDLINE | ID: mdl-28228100

ABSTRACT

BACKGROUND: For internal fixation of AO classification Type B lateral malleolar fracture, insertion of lag screws into the fracture plane and fixation with a one-third tubular plate as a neutralization plate are the standard treatment procedures. The one-third tubular plate is processed to a hook shape and hung on the distal end of the fibula. In this study, to compare the function of the hook and lag screws of a one-third tubular plate and LCP for osteosynthesis of lateral malleolar fracture, mechanical indices of internal fixation were compared among the one-third tubular plates with lag screws with and without the hook and a locking compression plate. METHODS: As mechanical tests, a compression test was performed in which compression in the bone axis direction produced by supporting the body weight was simulated, and a torsion test was performed in which external rotation of the bone axis caused by plantar flexion of the ankle joint was simulated. Muscle strength during walking and the force and torque acting on the ankle and knee joints were determined using inverse dynamic analysis. Finite element analysis was performed to analyze the function of hooks and lag screws. The joint reaction force determined by inverse dynamic analysis was adopted as the loading condition of finite element analysis. RESULTS: A stiffness equivalent to that of healthy bone could be achieved by all three internal fixations. It was clarified that the presence of the hook does not make a difference in stiffness. Displacement of the one-third tubular plate was small regardless of the presence or absence of the hook compared with those of locking compression plates. CONCLUSIONS: The presence of the hook did not make any difference in stiffness, suggesting that active preparation of the hook is unnecessary. We also clarified that lag screws inhibit displacement.


Subject(s)
Ankle Fractures/surgery , Bone Plates , Bone Screws , Fracture Fixation, Internal/instrumentation , Ankle Fractures/physiopathology , Ankle Joint/physiopathology , Biomechanical Phenomena , Compressive Strength , Ergonomics/methods , Finite Element Analysis , Fracture Fixation, Internal/methods , Humans , Materials Testing/methods , Muscle Strength/physiology
5.
ScientificWorldJournal ; 2014: 109876, 2014.
Article in English | MEDLINE | ID: mdl-25506061

ABSTRACT

Allogenic osteochondral tissue (OCT) is used for the treatment of large cartilage defects. Typically, OCTs collected during the disease-screening period are preserved at 4°C; however, the gradual reduction in cell viability during cold preservation adversely affects transplantation outcomes. Therefore, improved storage methods that maintain the cell viability of OCTs are needed to increase the availability of high-quality OCTs and improve treatment outcomes. Here, we evaluated whether long-term hydrogen delivery to preservation solution improved the viability of rat OCTs during cold preservation. Hydrogen-supplemented Dulbecco's Modified Eagles Medium (DMEM) and University of Wisconsin (UW) solution both significantly improved the cell viability of OCTs during preservation at 4°C for 21 days compared to nonsupplemented media. However, the long-term cold preservation of OCTs in DMEM containing hydrogen was associated with the most optimal maintenance of chondrocytes with respect to viability and morphology. Our findings demonstrate that OCTs preserved in DMEM supplemented with hydrogen are a promising material for the repair of large cartilage defects in the clinical setting.


Subject(s)
Cartilage/physiology , Cartilage/transplantation , Hydrogen/pharmacology , Organ Preservation Solutions/pharmacology , Tissue Survival/drug effects , Animals , Cartilage/drug effects , Chondrocytes/cytology , Chondrocytes/drug effects , Chondrocytes/metabolism , Cryopreservation , Male , Rats, Sprague-Dawley , Real-Time Polymerase Chain Reaction , Staining and Labeling
6.
Biomed Mater Eng ; 24(2): 1497-506, 2014.
Article in English | MEDLINE | ID: mdl-24642976

ABSTRACT

Reduction of cellular damage in densely cultured cell monolayers after cryopreservation by pre-incubation with hyaluronan (HA) was investigated. Monolayers of human dermal fibroblasts were cultured for 24 h at a density of 0.5×104 or 5×104 cells/cm2. The following two experimental conditions were compared: cells incubated with or without 0.5% w/w HA solution for 6 h. Samples were frozen from 4 to -80°C at 0.3 or 3°C/min in a cryoprotectant solution containing 10% w/w dimethyl sulfoxide, cooled down below -185°C, and then thawed. Post-thaw cell viability was evaluated by the fluorescent double-staining technique using a fluorescence microscope, and cellular uptake of the fluorescein-isothiocyanate-labeled HA after pre-incubation was also observed. Cell viability decreased with increasing cell density at both cooling rates without preliminary HA incubation. However, cell viability did not decrease at either cooling rate with preliminary HA incubation. Cellular HA uptake was observed. Pre-incubation with HA reduces cellular damage in densely cultured cell monolayers.


Subject(s)
Cryopreservation/methods , Cryoprotective Agents/metabolism , Fibroblasts/cytology , Hyaluronic Acid/metabolism , Cell Line , Cell Survival , Cryopreservation/instrumentation , Equipment Design , Fibroblasts/metabolism , Humans
7.
Cryo Letters ; 31(2): 101-11, 2010.
Article in English | MEDLINE | ID: mdl-20687452

ABSTRACT

The purpose of this study was to assess the availability of low-molecular-weight (low-MW) hyaluronan (HA) as a cryoprotectant for cellular cryopreservation. To clarify whether low-MW HA is cryoprotective, we evaluated the effect of HA concentration (0-5% w/w) in a cryoprotectant solution on cell membrane integrity after freeze-thaw. A test sample was created using human dermal fibroblast monolayers incubated in a culture dish for 24 h (37 degrees C, 5% CO2). Sodium hyaluronate (MW 3 x 10(4)-5 x 10(4)) dissolved in medium served as the cryoprotectant solution. Samples were immersed in the solution for 2 h at 0-4 degrees C. They were frozen at a cooling rate of 3 degrees C/min from 4 to -80 degrees C, cooled further to below -185 degrees C, and then thawed. Cell membrane integrity after thawing was evaluated using a trypan blue exclusion assay. The sample and freezing procedures were repeated in subsequent experiments, while the conditions of the solution immersion with respect to the sample varied. Next, to clarify whether the cryoprotective action of HA is intra- or extracellular, we performed three experiments. The first studied the dependence of membrane integrity after freeze-thaw on preliminary incubation time (0.75-24 h at 37 degrees C) with a sample immersed in the solution (5% w/w HA). In the second, membrane integrity of thawed samples that were initially frozen in a medium instead of solution, by removing extracellular HA following a preliminary 6-h incubation period, were evaluated. Thirdly, we investigated cellular uptake of fluorescein isothiocyanate-labeled HA (MW 10(5), 1% w/w) after a preliminary 6-h incubation period under fluorescent microscopy (without freeze-thaw). The results show that HA had a cryoprotective effect, and that this cryoprotective action was intracellular. Therefore, low- MW HA proves to be a promising cellular cryoprotectant.


Subject(s)
Cryopreservation , Cryoprotective Agents , Fibroblasts/cytology , Hyaluronic Acid , Skin/cytology , Cells, Cultured , Humans , Molecular Weight
8.
Cryo Letters ; 28(5): 337-46, 2007.
Article in English | MEDLINE | ID: mdl-18075703

ABSTRACT

The influence of cell density of cells cryopreserved inside a collagen matrix at various cooling rates was investigated. Human fibroblasts were three-dimensionally cultured for 2 days in a collagen sponge (20 mm in diameter and 1 mm in thickness) as an extracellular matrix to imitate biological tissue (artificial tissue). Different cell densities for the artificial tissue were used, from 10(5) to 10(7) cells/cm(3). Four artificial tissues were first stacked in a test chamber, frozen at a cooling rate of 0.3 to 50 degrees C/min in a solution of Dulbecco's Modified Eagle Medium, 20% fetal bovine serum and 10% dimethylsulfoxide, kept frozen below -185 degrees C for 2 hours, and then finally thawed. Membrane integrity of fibroblasts using a trypan blue exclusion assay was evaluated as an index for post-thaw cellular viability. Results show that with increasing cell density, the post-thaw membrane integrity decreased. Therefore, in the cryopreservation of biological tissue, it seems high cell density is one factor which causes a decline in viability.


Subject(s)
Cryopreservation/methods , Fibroblasts/cytology , Tissue Preservation/methods , Artificial Organs , Cell Count , Cell Survival , Collagen , Humans
9.
J Orthop Sci ; 10(1): 77-83, 2005.
Article in English | MEDLINE | ID: mdl-15666127

ABSTRACT

The purpose of this study was to develop a disinfection method using a microwave apparatus to treat large bone allografts. Heating of a bone allograft is an effective method for the disinfection of bacteria or inactivation of viruses. However, the size of the bone we can treat is limited, and following the popular method of using a bathtub is a lengthy process. The experimental system described here was designed using a microwave oven, an optical-fiber thermometer, and a power regulator. Large and small specimens, a femoral head, and a metatarsal were harvested from a bovine femur. The influence of size and the electrical or thermal characteristics of the specimens were assessed regarding temperature distribution after microwave irradiation. The effects of humidity or hot-air supply were also assessed. The average temperature of the bovine femoral head became 80 degrees C throughout the 15 min of microwave irradiation, although the temperature in the metatarsal did not attain uniformity. Microwave irradiation with a hot-air supply realized a uniform distribution of temperature at 83.0 degrees +/- 0.4 degrees C in the metatarsal within 15 min. Use of microwave irradiation enables quick heating for disinfection of large allograft bones when a hot-air supply was used as well.


Subject(s)
Bone and Bones , Hot Temperature , Microwaves , Sterilization/methods , Animals , Bone and Bones/microbiology , Cattle
10.
Clin Biomech (Bristol, Avon) ; 19(4): 362-9, 2004 May.
Article in English | MEDLINE | ID: mdl-15109756

ABSTRACT

OBJECTIVE: Estimation of appropriate radial clearance of ceramic-on-ceramic total hip prostheses to realize squeeze-film lubrication. BACKGROUND: Some clinical results show that severe wear occurs at ceramic-on-ceramic interfaces of total hip prostheses if the design parameters are improper. Appropriate design is required to realize the optimum lubrication of joint prostheses. DESIGN: Squeeze-film thickness of total hip prostheses with different radial clearances was numerically estimated under the physiological conditions of normal walking. METHODS: The changes in the fluid film thickness between ceramic balls and a ceramic cups with radial clearances of various values from 10 to 80 microm were numerically estimated under the assumption of normal walking conditions by means of elastohydrodynamic squeeze-film lubrication theory including the three-dimensional theory of elasticity. RESULTS: The minimum film thickness remained over 0.02 microm during a few walking steps with a radial clearance under 20 microm and a head diameter of 28 mm. On the other hand, the fluid film was squeezed out during a few steps when the radial clearance was greater than 30 microm. CONCLUSIONS: Squeeze-film lubrication can be realized in a ceramic-on-ceramic total hip prosthesis if the radial clearance is smaller than 20 microm on a head of 28 mm diameter. RELEVANCE: Fluid film lubrication is necessary to prevent the severe wear of ceramic joint prostheses. The present investigation shows which radial clearance of ceramic-on-ceramic hip prostheses is appropriate for squeeze-film lubrication under physiological conditions in daily action.


Subject(s)
Ceramics , Computer-Aided Design , Hip Joint/physiopathology , Hip Prosthesis , Locomotion , Microfluidics/methods , Prosthesis Design/methods , Synovial Fluid , Computer Simulation , Equipment Failure Analysis/methods , Friction , Hip Joint/surgery , Humans , Lubrication , Models, Biological , Pressure , Weight-Bearing
SELECTION OF CITATIONS
SEARCH DETAIL
...