Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 3 de 3
Filter
Add more filters










Database
Language
Publication year range
1.
Pacing Clin Electrophysiol ; 9(3): 316-21, 1986 May.
Article in English | MEDLINE | ID: mdl-2423973

ABSTRACT

A case is described of pacemaker pseudomalfunction caused by intermittent displacement of the ventricular lead. The displacement was shown to be caused by critically timed atrial contractions, dislodging the looped ventricular lead, and resulting in intermittent failure to stimulate the heart.


Subject(s)
Heart Block/therapy , Myocardial Contraction , Pacemaker, Artificial , Aged , Electrocardiography , Equipment Failure , Heart Atria , Heart Block/physiopathology , Humans , Male
SELECTION OF CITATIONS
SEARCH DETAIL
...