Biosens Bioelectron ; 77: 805-11, 2016 Mar 15.
Article in English | MEDLINE | ID: mdl-26513287

ABSTRACT

The ability of a diagnostic test to detect multiple pathogens simultaneously is useful to obtain meaningful information for clinical treatment and preventive measures. We report a highly sensitive and specific electrochemical biosensor assay for simultaneous detection of three gene targets using quantum dots (QDs). The targets are novel non-protein coding RNA (npcRNA) sequences of Vibrio cholerae, Salmonella sp. and Shigella sp., which cause diarrheal diseases. QDs (PbS, CdS, ZnS) were synthesized and functionalized with DNA probes that were specific to each pathogen. Electrochemical detection of QDs was performed using square wave anodic stripping voltammetry (SWASV). The QDs gave distinct peaks at 0.5 V (PbS), 0.75 V (CdS) and 1.1 V (ZnS). There was no interference in signal response when all three QDs were mixed and detected simultaneously. The detection limits of single and multiplex assays with linear targets and PCR products were in the attomolar ranges. The high assay sensitivity, in combination with specific npcRNA sequences as novel diagnostic targets, makes it a viable tool for detecting pathogens from food, environment and clinical samples.


Subject(s)
Conductometry/instrumentation , Microchemistry/instrumentation , Quantum Dots , RNA, Bacterial/analysis , RNA, Bacterial/genetics , RNA, Untranslated/chemistry , Biosensing Techniques/instrumentation , Equipment Design , Equipment Failure Analysis , RNA, Untranslated/genetics , Reproducibility of Results , Sensitivity and Specificity , Staining and Labeling
SELECTION OF CITATIONS
SEARCH DETAIL