Talanta ; 84(1): 187-91, 2011 Mar 15.
Article in English | MEDLINE | ID: mdl-21315918

ABSTRACT

A novel long drop time mercury electrode has been constructed from common fused-silica capillary (50 µm I.D., 360 µm E.D.). Proposed device provides reproducible mercury drops with typical lifetime 40-120 s. The electrode was used for a set of electrocapillary measurements aimed at determination of critical micelle concentration of anionic surfactants by a convection controlled drop-time technique. A critical micelle concentration of sodium dodecyl sulfate 5.6 ± 0.4 mmol L(-1) and 4.3 ± 0.4 mmol L(-1) were obtained in 1 mmol L(-1) and 5 mmol L(-1) phosphate buffer (pH 7.0), respectively. The values were comparable to those obtained from conductometric measurements under the same conditions (7.0 ± 0.1 mmol L(-1) and 5.2 ± 0.1 mmol L(-1), respectively) and the difference was explained in accordance with theory of hemi-micelle formation.


Subject(s)
Electrowetting/methods , Mercury/chemistry , Micelles , Silicon Dioxide/chemistry , Conductometry , Electrodes , Electrolytes/chemistry , Sodium Dodecyl Sulfate/chemistry , Surface-Active Agents/chemistry , Time Factors