Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 13 de 13
Filter
1.
Alzheimers Dement ; 2024 Sep 11.
Article in English | MEDLINE | ID: mdl-39258539

ABSTRACT

The magnetic resonance imaging (MRI) Core has been operating since Alzheimer's Disease Neuroimaging Initiative's (ADNI) inception, providing 20 years of data including reliable, multi-platform standardized protocols, carefully curated image data, and quantitative measures provided by expert investigators. The overarching purposes of the MRI Core include: (1) optimizing and standardizing MRI acquisition methods, which have been adopted by many multicenter studies and trials worldwide and (2) providing curated images and numeric summary values from relevant MRI sequences/contrasts to the scientific community. Over time, ADNI MRI has become increasingly complex. To remain technically current, the ADNI MRI protocol has changed substantially over the past two decades. The ADNI 4 protocol contains nine different imaging types (e.g., three dimensional [3D] T1-weighted and fluid-attenuated inversion recovery [FLAIR]). Our view is that the ADNI MRI data are a greatly underutilized resource. The purpose of this paper is to educate the scientific community on ADNI MRI methods and content to promote greater awareness, accessibility, and use. HIGHLIGHTS: The MRI Core provides multi-platform standardized protocols, carefully curated image data, and quantitative analysis by expert groups. The ADNI MRI protocol has undergone major changes over the past two decades to remain technically current. As of April 25, 2024, the following numbers of image series are available: 17,141 3D T1w; 6877 FLAIR; 3140 T2/PD; 6623 GRE; 3237 dMRI; 2846 ASL; 2968 TF-fMRI; and 2861 HighResHippo (see Table 1 for abbreviations). As of April 25, 2024, the following numbers of quantitative analyses are available: FreeSurfer 10,997; BSI 6120; tensor based morphometry (TBM) and TBM-SYN 12,019; WMH 9944; dMRI 1913; ASL 925; TF-fMRI NFQ 2992; and medial temporal subregion volumes 2726 (see Table 4 for abbreviations). ADNI MRI is an underutilized resource that could be more useful to the research community.

2.
Parkinsonism Relat Disord ; 103: 60-68, 2022 10.
Article in English | MEDLINE | ID: mdl-36063706

ABSTRACT

OBJECTIVE: To systematically evaluate structural MRI and diffusion MRI features for cross-sectional discrimination and tracking of longitudinal disease progression in early multiple system atrophy (MSA). METHODS: In a prospective, longitudinal study of synucleinopathies with imaging on 14 controls and 29 MSA patients recruited at an early disease stage (15 predominant cerebellar ataxia subtype or MSA-C and 14 predominant parkinsonism subtype or MSA-P), we computed regional morphometric and diffusion MRI features. We identified morphometric features by ranking them based on their ability to distinguish MSA-C from controls and MSA-P from controls and evaluated diffusion changes in these regions. For the top performing regions, we evaluated their utility for tracking longitudinal disease progression using imaging from 12-month follow-up and computed sample size estimates for a hypothetical clinical trial in MSA. We also computed these selected morphometric features in an independent validation dataset. RESULTS: We found that morphometric changes in the cerebellar white matter, brainstem, and pons can separate early MSA-C patients from controls both cross-sectionally and longitudinally (p < 0.01). The putamen and striatum, though useful for separating early MSA-P patients from control subjects at baseline, were not useful for tracking MSA disease progression. Cerebellum white matter diffusion changes aided in capturing early disease related degeneration in MSA. INTERPRETATION: Regardless of clinically predominant features at the time of MSA assessment, brainstem and cerebellar pathways progressively deteriorate with disease progression. Quantitative measurements of these regions are promising biomarkers for MSA diagnosis in early disease stage and potential surrogate markers for future MSA clinical trials.


Subject(s)
Multiple System Atrophy , Humans , Multiple System Atrophy/diagnostic imaging , Prospective Studies , Longitudinal Studies , Cross-Sectional Studies , Magnetic Resonance Imaging/methods , Cerebellum/diagnostic imaging , Disease Progression , Biomarkers , Diagnosis, Differential
3.
Mayo Clin Proc ; 95(6): 1195-1205, 2020 06.
Article in English | MEDLINE | ID: mdl-32498775

ABSTRACT

OBJECTIVE: To report population age-specific prevalence of core cerebrovascular disease lesions (infarctions, cerebral microbleeds, and white-matter hyperintensities detected with magnetic resonance imaging); estimate cut points for white-matter hyperintensity positivity; investigate sex differences in prevalence; and estimate prevalence of any core cerebrovascular disease features. PATIENTS AND METHODS: Participants in the population-based Mayo Clinic Study of Aging aged 50 to 89 years underwent fluid-attenuated inversion recovery and T2* gradient-recalled echo magnetic resonance imaging to assess cerebrovascular disease between October 10, 2011, and September 29, 2017. We characterized each participant as having infarct, normal versus abnormal white-matter hyperintensity, cerebral microbleed, or a combination of lesions. Prevalence of cerebrovascular disease biomarkers was derived through adjustment for nonparticipation and standardization to the population of Olmsted County, Minnesota. RESULTS: Among 1462 participants without dementia (median [range] age, 68 [50 to 89] y; men, 52.7%), core cerebrovascular disease features increased with age. Prevalence (95% CI) of cerebral microbleeds was 13.6% (11.6%-15.6%); infarcts, 11.7% (9.7%-13.8%); and abnormal white-matter hyperintensity, 10.7% (8.7%-12.6%). Infarcts and cerebral microbleeds were more common among men. In contrast, abnormal white-matter hyperintensity was more common among women ages 60 to 79 y and men, ages 80 y and older. Prevalence of any core cerebrovascular disease feature determined by presence of at least one cerebrovascular disease feature increased from 9.5% (ages 50 to 59 y) to 73.8% (ages 80 to 89 y). CONCLUSION: Whereas this study focused on participants without dementia, the high prevalence of cerebrovascular disease imaging lesions in elderly persons makes assignment of clinical relevance to cognition and other downstream manifestations more probabilistic than deterministic.


Subject(s)
Aging/physiology , Cerebrovascular Disorders/diagnostic imaging , Aged , Aged, 80 and over , Cerebrovascular Disorders/epidemiology , Cerebrovascular Disorders/pathology , Cohort Studies , Female , Humans , Magnetic Resonance Imaging/methods , Male , Middle Aged , Prevalence , Sex Distribution , White Matter/diagnostic imaging , White Matter/pathology
4.
Neuroimage Clin ; 11: 802-812, 2016.
Article in English | MEDLINE | ID: mdl-28050342

ABSTRACT

Alzheimer's disease (AD) researchers commonly use MRI as a quantitative measure of disease severity. Historically, hippocampal volume has been favored. Recently, "AD signature" measurements of gray matter (GM) volumes or cortical thicknesses have gained attention. Here, we systematically evaluate multiple thickness- and volume-based candidate-methods side-by-side, built using the popular FreeSurfer, SPM, and ANTs packages, according to the following criteria: (a) ability to separate clinically normal individuals from those with AD; (b) (extent of) correlation with head size, a nuisance covariatel (c) reliability on repeated scans; and (d) correlation with Braak neurofibrillary tangle stage in a group with autopsy. We show that volume- and thickness-based measures generally perform similarly for separating clinically normal from AD populations, and in correlation with Braak neurofibrillary tangle stage at autopsy. Volume-based measures are generally more reliable than thickness measures. As expected, volume measures are highly correlated with head size, while thickness measures are generally not. Because approaches to statistically correcting volumes for head size vary and may be inadequate to deal with this underlying confound, and because our goal is to determine a measure which can be used to examine age and sex effects in a cohort across a large age range, we thus recommend thickness-based measures. Ultimately, based on these criteria and additional practical considerations of run-time and failure rates, we recommend an AD signature measure formed from a composite of thickness measurements in the entorhinal, fusiform, parahippocampal, mid-temporal, inferior-temporal, and angular gyrus ROIs using ANTs with input segmentations from SPM12.


Subject(s)
Alzheimer Disease/pathology , Brain Mapping , Cerebral Cortex/pathology , Adult , Aged , Aged, 80 and over , Alzheimer Disease/diagnostic imaging , Cerebral Cortex/diagnostic imaging , Cognitive Dysfunction/pathology , Cohort Studies , Datasets as Topic , Female , Humans , Image Processing, Computer-Assisted , Magnetic Resonance Imaging , Male , Middle Aged , Psychiatric Status Rating Scales
5.
Neurobiol Aging ; 36 Suppl 1: S167-77, 2015 Jan.
Article in English | MEDLINE | ID: mdl-25444601

ABSTRACT

The Alzheimer's Disease Neuroimaging Initiative recently implemented accelerated T1-weighted structural imaging to reduce scan times. Faster scans may reduce study costs and patient attrition by accommodating people who cannot tolerate long scan sessions. However, little is known about how scan acceleration affects the power to detect longitudinal brain change. Using tensor-based morphometry, no significant difference was detected in numerical summaries of atrophy rates from accelerated and nonaccelerated scans in subgroups of patients with Alzheimer's disease, early or late mild cognitive impairment, or healthy controls over a 6- and 12-month scan interval. Whole-brain voxelwise mapping analyses revealed some apparent regional differences in 6-month atrophy rates when comparing all subjects irrespective of diagnosis (n = 345). No such whole-brain difference was detected for the 12-month scan interval (n = 156). Effect sizes for structural brain changes were not detectably different in accelerated versus nonaccelerated data. Scan acceleration may influence brain measures but has minimal effects on tensor-based morphometry-derived atrophy measures, at least over the 6- and 12-month intervals examined here.


Subject(s)
Alzheimer Disease/diagnosis , Alzheimer Disease/pathology , Brain/pathology , Diffusion Magnetic Resonance Imaging/methods , Neuroimaging/methods , Aged , Aged, 80 and over , Atrophy , Brain Mapping/methods , Cognitive Dysfunction/diagnosis , Cognitive Dysfunction/pathology , Female , Humans , Male , Time Factors
7.
J Magn Reson Imaging ; 36(5): 1234-40, 2012 Nov.
Article in English | MEDLINE | ID: mdl-22570196

ABSTRACT

PURPOSE: To evaluate the effects of recent advances in magnetic resonance imaging (MRI) radiofrequency (RF) coil and parallel imaging technology on brain volume measurement consistency. MATERIALS AND METHODS: In all, 103 whole-brain MRI volumes were acquired at a clinical 3T MRI, equipped with a 12- and 32-channel head coil, using the T1-weighted protocol as employed in the Alzheimer's Disease Neuroimaging Initiative study with parallel imaging accelerations ranging from 1 to 5. An experienced reader performed qualitative ratings of the images. For quantitative analysis, differences in composite width (CW, a measure of image similarity) and boundary shift integral (BSI, a measure of whole-brain atrophy) were calculated. RESULTS: Intra- and intersession comparisons of CW and BSI measures from scans with equal acceleration demonstrated excellent scan-rescan accuracy, even at the highest acceleration applied. Pairs-of-scans acquired with different accelerations exhibited poor scan-rescan consistency only when differences in the acceleration factor were maximized. A change in the coil hardware between compared scans was found to bias the BSI measure. CONCLUSION: The most important findings are that the accelerated acquisitions appear to be compatible with the assessment of high-quality quantitative information and that for highest scan-rescan accuracy in serial scans the acquisition protocol should be kept as consistent as possible over time.


Subject(s)
Algorithms , Brain/anatomy & histology , Brain/physiology , Image Interpretation, Computer-Assisted/methods , Imaging, Three-Dimensional/methods , Magnetic Resonance Imaging/methods , Adult , Female , Humans , Image Enhancement/methods , Male , Observer Variation , Organ Size/physiology , Reproducibility of Results , Sensitivity and Specificity
8.
Med Phys ; 36(6): 2193-205, 2009 Jun.
Article in English | MEDLINE | ID: mdl-19610308

ABSTRACT

The objectives of this study are as follows: to describe practical implementation challenges of multisite, multivendor quantitative studies; to describe the MRI phantom and analysis software used in the Alzheimer's Disease Neuroimaging Initiative (ADNI) study, illustrate the utility of the system for measuring scanner performance, the ability to assess gradient field nonlinearity corrections: and to recover human brain images without geometric scaling errors in multisite studies. ADNI is a large multicenter study with each center having its own copy of the phantom. The design of the phantom and analysis software are presented as results from predistribution systematics studies and results from field experience with the phantom at 58 enrolling ADNI sites over a 3 year period. The estimated coefficients of variation intrinsic to measurements of geometry in a single phantom are in the range of 3-5 parts in 10(4). Phantom measurements accurately detect linear and nonlinear scaling in images. Gradient unwarping methods are readily assessed by phantom nonlinearity measurements. Phantom-based scaling correction reduces observed geometric drift in human images by one-third or more. Repair or replacement of phantoms between scans, however, is a confounding factor. The ADNI phantom can be used to assess both scanner performance and the validity of postprocessing image corrections in order to reduce systematic errors in human images. Reduced measurement errors should decrease measurement bias and increase statistical power for measurements of rates of change in the brain structure in AD treatment trials. Perhaps the greatest practical value of incorporating ADNI phantom measurements in a multisite study is to identify scanner errors through central monitoring. This approach has resulted in identification of system errors including sites misidentification of their own gradient hardware and the disabling of autoshim, and a miscalibrated laser alignment light. If undetected, these errors would have contributed to imprecision in quantitative metrics at over 25% of all enrolling ADNI sites.


Subject(s)
Alzheimer Disease/diagnosis , Brain/pathology , Image Enhancement/instrumentation , Image Interpretation, Computer-Assisted/instrumentation , Magnetic Resonance Imaging/instrumentation , Phantoms, Imaging , Equipment Design , Equipment Failure Analysis/methods , Humans , Image Enhancement/methods , Magnetic Resonance Imaging/methods , Reproducibility of Results , Sensitivity and Specificity
9.
Neuroimage ; 45(3): 645-55, 2009 Apr 15.
Article in English | MEDLINE | ID: mdl-19280686

ABSTRACT

Tensor-based morphometry can recover three-dimensional longitudinal brain changes over time by nonlinearly registering baseline to follow-up MRI scans of the same subject. Here, we compared the anatomical distribution of longitudinal brain structural changes, over 12 months, using a subset of the ADNI dataset consisting of 20 patients with Alzheimer's disease (AD), 40 healthy elderly controls, and 40 individuals with mild cognitive impairment (MCI). Each individual longitudinal change map (Jacobian map) was created using an unbiased registration technique, and spatially normalized to a geometrically-centered average image based on healthy controls. Voxelwise statistical analyses revealed regional differences in atrophy rates, and these differences were correlated with clinical measures and biomarkers. Consistent with prior studies, we detected widespread cerebral atrophy in AD, and a more restricted atrophic pattern in MCI. In MCI, temporal lobe atrophy rates were correlated with changes in mini-mental state exam (MMSE) scores, clinical dementia rating (CDR), and logical/verbal learning memory scores. In AD, temporal atrophy rates were correlated with several biomarker indices, including a higher CSF level of p-tau protein, and a greater CSF tau/beta amyloid 1-42 (ABeta42) ratio. Temporal lobe atrophy was significantly faster in MCI subjects who converted to AD than in non-converters. Serial MRI scans can therefore be analyzed with nonlinear image registration to relate ongoing neurodegeneration to a variety of pathological biomarkers, cognitive changes, and conversion from MCI to AD, tracking disease progression in 3-dimensional detail.


Subject(s)
Alzheimer Disease/pathology , Biomarkers/cerebrospinal fluid , Brain/pathology , Image Interpretation, Computer-Assisted/methods , Nerve Degeneration/pathology , Aged , Aged, 80 and over , Alzheimer Disease/cerebrospinal fluid , Amyloid beta-Peptides/cerebrospinal fluid , Cognition , Disease Progression , Female , Follow-Up Studies , Humans , Magnetic Resonance Imaging , Male , Middle Aged , Neuropsychological Tests , Positron-Emission Tomography , tau Proteins/cerebrospinal fluid
10.
Neuroimage ; 41(1): 19-34, 2008 May 15.
Article in English | MEDLINE | ID: mdl-18378167

ABSTRACT

Tensor-based morphometry (TBM) creates three-dimensional maps of disease-related differences in brain structure, based on nonlinearly registering brain MRI scans to a common image template. Using two different TBM designs (averaging individual differences versus aligning group average templates), we compared the anatomical distribution of brain atrophy in 40 patients with Alzheimer's disease (AD), 40 healthy elderly controls, and 40 individuals with amnestic mild cognitive impairment (aMCI), a condition conferring increased risk for AD. We created an unbiased geometrical average image template for each of the three groups, which were matched for sex and age (mean age: 76.1 years+/-7.7 SD). We warped each individual brain image (N=120) to the control group average template to create Jacobian maps, which show the local expansion or compression factor at each point in the image, reflecting individual volumetric differences. Statistical maps of group differences revealed widespread medial temporal and limbic atrophy in AD, with a lesser, more restricted distribution in MCI. Atrophy and CSF space expansion both correlated strongly with Mini-Mental State Exam (MMSE) scores and Clinical Dementia Rating (CDR). Using cumulative p-value plots, we investigated how detection sensitivity was influenced by the sample size, the choice of search region (whole brain, temporal lobe, hippocampus), the initial linear registration method (9- versus 12-parameter), and the type of TBM design. In the future, TBM may help to (1) identify factors that resist or accelerate the disease process, and (2) measure disease burden in treatment trials.


Subject(s)
Alzheimer Disease/pathology , Brain/pathology , Cognition Disorders/pathology , Aged , Aged, 80 and over , Algorithms , Alzheimer Disease/psychology , Atrophy , Brain Mapping , Cognition Disorders/psychology , Female , Humans , Image Processing, Computer-Assisted , Magnetic Resonance Imaging , Male , Middle Aged , Neuropsychological Tests , Psychiatric Status Rating Scales
11.
Neurobiol Aging ; 29(9): 1285-95, 2008 Sep.
Article in English | MEDLINE | ID: mdl-17452062

ABSTRACT

The vitamin E and donepezil trial for the treatment of amnestic mild cognitive impairment (MCI) was conducted at 69 centers in North America; 24 centers participated in an MRI sub study. The objective of this study was to evaluate the effect of treatment on MRI atrophy rates; and validate rate measures from serial MRI as indicators of disease progression in multi center therapeutic trials for MCI. Annual percent change (APC) from baseline to follow-up was measured for hippocampus, entorhinal cortex, whole brain, and ventricle in the 131 subjects who remained in the treatment study and completed technically satisfactory baseline and follow-up scans. Although a non-significant trend toward slowing of hippocampal atrophy rates was seen in APOE is an element of 4 carriers treated with donepezil; no treatment effect was confirmed for any MRI measure in either treatment group. For each of the four brain atrophy rate measures, APCs were greater in subjects who converted to AD than non-converters, and were greater in APOE is an element of 4 carriers than non-carriers. MRI APCs and changes in cognitive test performance were uniformly correlated in the expected direction (all p<0.000). Results of this study support the feasibility of using MRI as an outcome measure of disease progression in multi center therapeutic trials for MCI.


Subject(s)
Cognition Disorders/drug therapy , Cognition Disorders/pathology , Hippocampus/pathology , Indans/administration & dosage , Piperidines/administration & dosage , Vitamin E/administration & dosage , Aged , Aged, 80 and over , Antioxidants/administration & dosage , Atrophy/drug therapy , Atrophy/epidemiology , Atrophy/pathology , Cholinesterase Inhibitors/administration & dosage , Cognition Disorders/epidemiology , Donepezil , Female , Hippocampus/drug effects , Humans , Incidence , Longitudinal Studies , Male , Middle Aged , Treatment Outcome , United States/epidemiology
12.
Neuroimage ; 31(2): 627-40, 2006 Jun.
Article in English | MEDLINE | ID: mdl-16480900

ABSTRACT

Measures of brain change can be computed from sequential MRI scans, providing valuable information on disease progression, e.g., for patient monitoring and drug trials. Tensor-based morphometry (TBM) creates maps of these brain changes, visualizing the 3D profile and rates of tissue growth or atrophy, but its sensitivity depends on the contrast and geometric stability of the images. As part of the Alzheimer's Disease Neuroimaging Initiative (ADNI), 17 normal elderly subjects were scanned twice (at a 2-week interval) with several 3D 1.5 T MRI pulse sequences: high and low flip angle SPGR/FLASH (from which Synthetic T1 images were generated), MP-RAGE, IR-SPGR (N = 10) and MEDIC (N = 7) scans. For each subject and scan type, a 3D deformation map aligned baseline and follow-up scans, computed with a nonlinear, inverse-consistent elastic registration algorithm. Voxelwise statistics, in ICBM stereotaxic space, visualized the profile of mean absolute change and its cross-subject variance; these maps were then compared using permutation testing. Image stability depended on: (1) the pulse sequence; (2) the transmit/receive coil type (birdcage versus phased array); (3) spatial distortion corrections (using MEDIC sequence information); (4) B1-field intensity inhomogeneity correction (using N3). SPGR/FLASH images acquired using a birdcage coil had least overall deviation. N3 correction reduced coil type and pulse sequence differences and improved scan reproducibility, except for Synthetic T1 images (which were intrinsically corrected for B1-inhomogeneity). No strong evidence favored B0 correction. Although SPGR/FLASH images showed least deviation here, pulse sequence selection for the ADNI project was based on multiple additional image analyses, to be reported elsewhere.


Subject(s)
Brain Mapping/methods , Brain/anatomy & histology , Aged , Analysis of Variance , Female , Humans , Image Processing, Computer-Assisted , Magnetic Resonance Imaging/methods , Male , Middle Aged , Reference Values , Reproducibility of Results
13.
Neuroimage ; 30(4): 1196-202, 2006 May 01.
Article in English | MEDLINE | ID: mdl-16380273

ABSTRACT

Measuring rates of brain atrophy from serial magnetic resonance imaging (MRI) studies is an attractive way to assess disease progression in neurodegenerative disorders, particularly Alzheimer's disease (AD). A widely recognized approach is the boundary shift integral (BSI). The objective of this study was to evaluate how several common scan non-idealities affect the output of the BSI algorithm. We created three types of image non-idealities between the image volumes in a serial pair used to measure between-scan change: inconsistent image contrast between serial scans, head motion, and poor signal-to-noise (SNR). In theory the BSI volume difference measured between each pair of images should be zero and any deviation from zero should represent corruption of the BSI measurement by some non-ideality intentionally introduced into the second scan in the pair. Two different BSI measures were evaluated, whole brain and ventricle. As the severity of motion, noise, and non-congruent image contrast increased in the second scan, the calculated BSI values deviated progressively more from the expected value of zero. This study illustrates the magnitude of the error in measures of change in brain and ventricle volume across serial MRI scans that can result from commonly encountered deviations from ideal image quality. The magnitudes of some of the measurement errors seen in this study exceed the disease effect in AD shown in various publications, which range from 1% to 2.78% per year for whole brain atrophy and 5.4% to 13.8% per year for ventricle expansion (Table 1). For example, measurement error may exceed 100% if image contrast properties dramatically differ between the two scans in a measurement pair. Methods to maximize consistency of image quality over time are an essential component of any quantitative serial MRI study.


Subject(s)
Alzheimer Disease/diagnosis , Brain/pathology , Image Processing, Computer-Assisted/methods , Magnetic Resonance Imaging/methods , Adult , Aged , Alzheimer Disease/pathology , Artifacts , Atrophy , Cephalometry/statistics & numerical data , Cerebral Ventricles/pathology , Computer Simulation , Disease Progression , Female , Humans , Male , Mathematical Computing , Middle Aged , Neurodegenerative Diseases , Reference Values , Sensitivity and Specificity , Software
SELECTION OF CITATIONS
SEARCH DETAIL