Respiration ; : 1-5, 2024 Jun 17.
Article in English | MEDLINE | ID: mdl-38885628

ABSTRACT

INTRODUCTION: Endobronchial radiofrequency ablation (RFA) is a novel minimally invasive approach to management of peripheral non-small-cell lung cancer (NSCLC) in medically inoperable patients. Minimally invasive ablative techniques are generally delivered with cone-beam computed tomography (CBCT) guidance. CBCT requires a significant number of two dimensional imaging projections to be acquired which is then reconstructed as a three-dimensional cone-beam image. The objective of this study was to determine the radiation dosimetry consequent to use of CBCT guidance for bronchoscopic RFA. METHODS: Post hoc analysis of data following bronchoscopic RFA of stage I biopsy-confirmed NSCLC performed with CBCT. Effective dose estimates for these patients were calculated using PCXMC2.0 software. RESULTS: Ten patients underwent bronchoscopic RFA, with a median 3 (range 2-4) CBCT spins per procedure. Mean dose area product (DAP) per procedure was 7,778 µGy.m2 (±4,743) with an effective dose of 11.6 mSv (±7.4). The DAP per spin for these 10 patients varied from 83.8 to 8,625.6 µGy.m2 (effective dose range 0.15-13.81 mSv). CONCLUSION: This is the first study to report radiation dosimetry consequent to CT guidance for bronchoscopic RFA procedures. Effective doses appear comparable to other CT fluoroscopic procedures.