J AAPOS ; 7(4): 291-2, 2003 Aug.
Article in English | MEDLINE | ID: mdl-12917618

ABSTRACT

PURPOSE: To compare scleral Tono-Pen (Medtronic Solan, Jacksonville, FL) readings to corneal Tono-Pen readings. METHODS: Intraocular pressure (IOP) was measured prospectively in 72 eyes of 37 adult patients and in 10 eyes of 5 children. Measurements were taken on the central cornea and on the sclera. Recorded measurements were within 5% confidence levels by Tono-Pen. RESULTS: Corneal IOP ranged from 10 to 28 mm Hg (mean +/- standard deviation, 17.0 +/- 3.8 mm Hg). Scleral measurements ranged from 4 to 99 mm Hg (40.4 +/- 23.0 mm Hg). Scleral measurements ranged from 11 mm Hg lower to 76 mm Hg higher than corneal measurements. CONCLUSIONS: Tono-Pen readings obtained from sclera are of no clinical value and should not be used to approximate corneal IOP.


Subject(s)
Cornea/physiology , Intraocular Pressure/physiology , Sclera/physiology , Tonometry, Ocular/methods , Adolescent , Adult , Aged , Aged, 80 and over , Child , Child, Preschool , Humans , Middle Aged , Prospective Studies , Reproducibility of Results