1.
Cardiovasc Intervent Radiol ; 19(4): 275-7, 1996.
Article in English | MEDLINE | ID: mdl-8755083

ABSTRACT

Two cases are presented in which a pigtail catheter was entrapped by the chordae tendineae of the tricuspid valve during pulmonary arteriography. A technique for removal of the catheter from its entanglement by the chordae tendineae is described. Caution must be taken when advancing through the right ventricle a catheter that appears to be entrapped by the chordae tendineae. When such an entanglement occurs, measures to reduce the risk of rupturing a papillary muscle must be taken.


Subject(s)
Angiography/adverse effects , Angiography/instrumentation , Catheterization/instrumentation , Chordae Tendineae , Pulmonary Artery/diagnostic imaging , Tricuspid Valve , Adult , Female , Humans , Male , Middle Aged