J Biomed Opt ; 9(6): 1363-8, 2004.
Article in English | MEDLINE | ID: mdl-15568959

ABSTRACT

We develop the Pediatric Vision Screener (PVS) to automatically detect ocular misalignment (strabismus) and defocus in human subjects. The PVS utilizes binocular retinal birefringence scanning to determine when both eyes are aligned, with a theoretical accuracy of <1 deg. The device employs an autoconjugate, bull's-eye detector-based system to detect focus. The focus and alignment pathways are separated by both wavelength and data acquisition timing. Binocular focus and alignment are detected in rapid alternating sequence, measuring both parameters in both eyes in <0.5 sec. In this work, the theory and design of the PVS are described in detail. With objective, automated measurement of both alignment and focus, the PVS represents a new approach to screening children for treatable eye disease such as amblyopia.


Subject(s)
Amblyopia/diagnosis , Diagnosis, Computer-Assisted/methods , Optics and Photonics/instrumentation , Pediatrics/instrumentation , Strabismus/diagnosis , Vision Tests/instrumentation , Adult , Amblyopia/etiology , Equipment Design , Equipment Failure Analysis , Feasibility Studies , Female , Humans , Pediatrics/methods , Reproducibility of Results , Sensitivity and Specificity , Strabismus/complications , User-Computer Interface , Vision Tests/methods