Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 9 de 9
Filter
Add more filters










Database
Language
Publication year range
1.
Phys Med Biol ; 63(22): 225007, 2018 11 09.
Article in English | MEDLINE | ID: mdl-30412476

ABSTRACT

By combining magnetic resonance imaging (MRI) scanners and radiotherapy treatment units the need arises for new radiation measurement equipment that can be used in the magnetic field of the MRI. This study describes the investigation of the influence of the 1.5 T magnetic field from an MRI linac on the STARCHECKMAXI MR, a large 2D ionization chamber detector panel. Measurements were performed on an MRI linac and a conventional linac to investigate the behaviour of the detector panel with and without the 1.5 T magnetic field. We measured reproducibility, linearity, warm-up effect, saturation/recombination and chamber orientation. A comparison with gafchromic film was performed and the effect of motion of the panel during measurements inside a magnetic field was investigated. The reproducibility, linearity, warm-up effect, saturation/recombination show no significant deviations with or without magnetic field. An absolute difference in reading of 2.1% was found between off-axis chambers on different axes. The comparison with film shows good agreement. Spurious readings are induced while the panel is undergoing a motion in the magnetic field during measurements. The STARCHECKMAXI MR is suited for use in a 1.5 T MRI linac. Care must be taken when comparing un-normalized profiles from different axes of the detector panel and when the panel is undergoing motion during measurements.


Subject(s)
Magnetic Resonance Imaging/methods , Particle Accelerators/instrumentation , Humans , Magnetic Fields , Magnetic Resonance Imaging/instrumentation , Radiometry/methods , Reproducibility of Results
2.
Med Phys ; 41(3): 031714, 2014 Mar.
Article in English | MEDLINE | ID: mdl-24593718

ABSTRACT

PURPOSE: To compare the dosimetric and geometric properties of a commercial x-ray based image-guided small animal irradiation system, installed at three institutions and to establish a complete and broadly accessible commissioning procedure. METHODS: The system consists of a 225 kVp x-ray tube with fixed field size collimators ranging from 1 to 44 mm equivalent diameter. The x-ray tube is mounted opposite a flat-panel imaging detector, on a C-arm gantry with 360° coplanar rotation. Each institution performed a full commissioning of their system, including half-value layer, absolute dosimetry, relative dosimetry (profiles, percent depth dose, and relative output factors), and characterization of the system geometry and mechanical flex of the x-ray tube and detector. Dosimetric measurements were made using Farmer-type ionization chambers, small volume air and liquid ionization chambers, and radiochromic film. The results between the three institutions were compared. RESULTS: At 225 kVp, with 0.3 mm Cu added filtration, the first half value layer ranged from 0.9 to 1.0 mm Cu. The dose-rate in-air for a 40 × 40 mm(2) field size, at a source-to-axis distance of 30 cm, ranged from 3.5 to 3.9 Gy/min between the three institutions. For field sizes between 2.5 mm diameter and 40 × 40 mm(2), the differences between percent depth dose curves up to depths of 3.5 cm were between 1% and 4% on average, with the maximum difference being 7%. The profiles agreed very well for fields >5 mm diameter. The relative output factors differed by up to 6% for fields larger than 10 mm diameter, but differed by up to 49% for fields ≤5 mm diameter. The mechanical characteristics of the system (source-to-axis and source-to-detector distances) were consistent between all three institutions. There were substantial differences in the flex of each system. CONCLUSIONS: With the exception of the half-value layer, and mechanical properties, there were significant differences between the dosimetric and geometric properties of the three systems. This underscores the need for careful commissioning of each individual system for use in radiobiological experiments.


Subject(s)
Radiometry/methods , Radiotherapy, Image-Guided/methods , Animals , Calibration , Equipment Design , Humans , Particle Accelerators , Phantoms, Imaging , Radiometry/instrumentation , Radiotherapy, Conformal/methods , Reproducibility of Results , Software , X-Rays
3.
Phys Med Biol ; 43(9): 2665-75, 1998 Sep.
Article in English | MEDLINE | ID: mdl-9755953

ABSTRACT

The use of intensity modulation with multiple static fields has been suggested by many authors as a way to achieve highly conformal fields in radiotherapy. However, quality assurance of linear accelerators is generally done only for beam segments of 100 MU or higher, and by measuring beam profiles once the beam has stabilized. We propose a set of measurements to check the stability of dose delivery in small segments, and present measured data from three radiotherapy centres. The dose delivered per monitor unit, MU, was measured for various numbers of MU segments. The field flatness and symmetry were measured using either photographic films that are subsequently scanned by a densitometer, or by using a diode array. We performed the set of measurements at the three radiotherapy centres on a set of five different Philips SL accelerators with energies of 6 MV, 8 MV, 10 MV and 18 MV. The dose per monitor unit over the range of 1 to 100 MU was found to be accurate to within +/-5% of the nominal dose per monitor unit as defined for the delivery of 100 MU for all the energies. For four out of the five accelerators the dose per monitor unit over the same range was even found to be accurate to within +/-2%. The flatness and symmetry were in some cases found to be larger for small segments by a maximum of 9% of the flatness/symmetry for large segments. The result of this study provides the dosimetric evidence that the delivery of small segment doses as top-up fields for beam intensity modulation is feasible. However, it should be stressed that linear accelerators have different characteristics for the delivery of small segments, hence this type of measurement should be performed for each machine before the delivery of small dose segments is approved. In some cases it may be advisable to use a low pulse repetition frequency (PRF) to obtain more accurate dose delivery of small segments.


Subject(s)
Particle Accelerators/standards , Radiotherapy, High-Energy/methods , Radiotherapy, High-Energy/standards , Biophysical Phenomena , Biophysics , Humans , Particle Accelerators/statistics & numerical data , Quality Control , Radiometry/instrumentation , Radiometry/statistics & numerical data , Radiotherapy Dosage , Radiotherapy, High-Energy/statistics & numerical data , Technology, Radiologic
4.
Radiother Oncol ; 27(2): 156-63, 1993 May.
Article in English | MEDLINE | ID: mdl-8356226

ABSTRACT

Dosimetric characteristics of a number of clinically applied electron beams were analyzed as part of a dosimetry intercomparison program performed at the radiotherapy centers in The Netherlands. Absorbed dose values, determined under reference conditions, were compared during site visits with stated dose values. The mean deviation was 0.2% with a standard deviation (SD) of 2.5%. The maximum deviation was 5.7%. The largest differences were due to differences in the calibration procedures and differences in the numerical values of conversion factors adopted from different dosimetry protocols. In addition, a number of clinically relevant parameters of the dose distribution on the central beam axis were analyzed including the depth of the 85% and 50% relative dose, the dose reduction at the depth of the 50% relative dose and the surface dose. The average difference between the stated and measured therapeutic depth (85% dose level) was -0.4 mm with an SD = 1.2 mm. Due to this dosimetric uncertainty observed a safety margin of about 3 mm at the therapeutic depth is recommended. The maximum difference between stated and observed mean energy of the electron beams had only a small influence, < 1%, on the absorbed dose determination. The normalized dose gradient is not an adequate parameter to describe the dose reduction beyond the therapeutic depth. The depth of a low dose level is a better parameter. The relative dose at the surface showed differences up to 10% between scanning electron beams and beams from accelerators with a single scattering foil and closed wall collimating system.


Subject(s)
Health Facilities , Radiotherapy Dosage , Electrons , Humans , Netherlands , Radiometry
5.
Radiother Oncol ; 11(4): 405-14, 1988 Apr.
Article in English | MEDLINE | ID: mdl-3131846

ABSTRACT

As part of a quality assurance program in The Netherlands, the performance of computer planning systems was tested. Relative dose values, determined with an ionization chamber, were compared with dose values obtained from locally applied computer planning systems. Several clinically relevant situations were investigated: perpendicular incident beams, wedged beams, oblique incident beams, variable source-surface distances (SSD) and off-axis planes. The mean value of the ratios of calculated to measured dose values is 0.994, with an uncertainty of 2.4% (1 S.D.) and a maximum deviation of 9%, for all combinations of energies, planning systems and geometries investigated. The uncertainty for each situation separately was less than 2% (1 S.D.), except for the wedged beams and off-axis plane, which showed uncertainties of 2.6% (1 S.D.). Part of the additional uncertainty for the wedged beams originates from the value chosen for the wedge factor. Systematic deviations between calculated and measured dose values were investigated for three commercially available planning systems, separately. The mean deviation was smaller than 1% (1 S.D.), for most situations. Only for the wedged beams, larger deviations, up to a mean deviation of 2.6%, were observed.


Subject(s)
Radiotherapy Planning, Computer-Assisted , Radiotherapy, Computer-Assisted , Radiotherapy, High-Energy/standards , Humans , Models, Structural , Netherlands , Quality Assurance, Health Care , Radiation , Radiometry/instrumentation , Radiotherapy Dosage
6.
Radiother Oncol ; 9(1): 33-44, 1987 May.
Article in English | MEDLINE | ID: mdl-3602428

ABSTRACT

In 1985, a dosimetry intercomparison was performed at all 20 radiotherapy centres in The Netherlands. Absorbed dose was determined with an ionization chamber under reference conditions in a water phantom for cobalt-60 gamma-ray and megavoltage X-ray beams. The mean difference between measured and stated dose values was 0.5% with a standard deviation of 1.9%, but up to 6% at maximum. As soon as all institutes apply a common dosimetry protocol, this maximum difference will reduce to about 2%. In addition, an anthropomorphic phantom was irradiated to simulate the treatment of a prostatic cancer. The dose, determined with an ionization chamber at the isocentre and thermoluminescent dosimeters (TLD) powder at several points situated in the target volume, the bladder and the rectum, was compared with the stated dose calculated with the local planning system. Only small differences were found between the measured and stated dose at the isocentre: on the average 1.5%, with a standard deviation of 1.5%. The difference between stated and measured dose at several points situated in the target volume was on the average 0.4%, with a standard deviation of 5.2%. Almost the same result was found for a point situated in the bladder. In the rectum, the average difference was about 4%, however, with a large standard deviation, 18%, due to the relatively steep dose gradient at these points.


Subject(s)
Prostatic Neoplasms/radiotherapy , Radiotherapy Dosage , Radiotherapy/standards , Gamma Rays , Male , Netherlands , Quality Control , X-Rays
7.
Radiother Oncol ; 8(1): 49-56, 1987 Jan.
Article in English | MEDLINE | ID: mdl-3809601

ABSTRACT

In a recently published code of practice for the dosimetry of high-energy photon beams, the absorbed dose to water is determined using an ionization chamber having an air kerma calibration factor and applying the air kerma to absorbed dose conversion factor Cw,u. The consistency of these Cw,u values has been determined for four commonly employed types of ionization chambers in photon beams with quality varying between 60Co gamma-rays and 25 MV X-rays. Using a graphite calorimeter, Cw,u has been determined for a graphite-walled ionization chamber (NE 2561) for the same qualities. The values of Cw,u determined with the calorimeter are within the experimental uncertainty equal to Cw,u values determined according to any of the recent dosimetry protocols.


Subject(s)
Radiation Dosage , Calorimetry , Radiometry
8.
Radiother Oncol ; 7(4): 371-84, 1986 Dec.
Article in English | MEDLINE | ID: mdl-3544087

ABSTRACT

Recent revision of exposure and air kerma standards in Standards Laboratories require a simultaneous change in physical parameters at other positions in the dosimetry chain. Adoption of new data, recommended by international organizations, will introduce changes in absorbed dose determinations in high-energy photon beams using ionization chambers. A new code of practice has therefore been drafted using a consistent set of data. In this code of practice, single conversion factors are given to convert ionization chamber reading to absorbed dose to water for some types of reference ionization chamber as a function of radiation quality. Equations and recommended numerical data for the physical parameters and correction and conversion factors will be provided.


Subject(s)
Radiation Dosage , Calibration , Water
9.
Phys Med Biol ; 31(4): 407-16, 1986 Apr.
Article in English | MEDLINE | ID: mdl-3090570

ABSTRACT

Absorbed dose values were determined under the reference conditions in a phantom irradiated by high-energy photon beams with quality varying between 60Co gamma rays and 25 MV X-rays, using four commonly employed types of ionisation chamber. The ionisation chamber readings were converted to absorbed dose values applying the recent NACP, AAPM and SEFM Protocols and the revised HPA Code of Practice. The AAPM and SEFM Protocols gave consistent results for the four types of chamber whereas the NACP Protocol should be adapted to take the differences in chamber wall material and chamber dimensions into account. Absorbed dose values determined with the standard chamber and procedure recommended in the HPA Code of Practice show good agreement, within 0.8%, with absorbed dose values obtained using the AAPM and SEFM Protocols.


Subject(s)
Radiotherapy Dosage/standards , Radiotherapy, High-Energy/standards , Cobalt Radioisotopes/therapeutic use , Gamma Rays , Humans , Quality Assurance, Health Care
SELECTION OF CITATIONS
SEARCH DETAIL
...