Anal Sci ; 36(6): 739-743, 2020 Jun 10.
Article in English | MEDLINE | ID: mdl-31932522

ABSTRACT

A clear and positive correlation between the CO2 concentration and the blood-sugar level has been observed via a non-invasive and time-dependent monitoring of CO2 concentration from human breath, which is carried out by using a home-made gas chromatography (GC)/milli-whistle compact analyzer. The time-dependent sampling of the CO2 concentration correlated between 5.0 to 5.6% (1% = 104 ppm) in accordance with blood-sugar level variations of 80 to 110 mg/dL. The analytical method results in a rapid, continuous and non-invasive determination of blood-sugar level via measurement of the CO2 concentration exhaled from the lungs.


Subject(s)
Breath Tests , Carbon Dioxide/blood , Sugars/blood , Chromatography, Gas/instrumentation , Humans , Time Factors