Article in English | MEDLINE | ID: mdl-18276535

ABSTRACT

A highly sensitive piezoelectric ultrasonic micro-sensor with a grooved multilayer membrane was developed by a Si-based MEMS technique. The groove was located at one-quarter of the distance away from the edge of the membrane and opened into piezoelectric layer. The piezoelectric layer Pb(Zr,Ti)O(3) (PZT) was 2.2 microm thick and was prepared by a sol-gel method. The prepared PZT film was pure perovskite and showed a highly (100) textured structure. The sensitivity of the fabricated piezoelectric ultrasonic sensor without the groove structure was 100 microV/Pa. In comparison, the sensitivity of the ultrasonic sensor with the groove structure was about 500 microV/Pa, which is 5 times that without the groove structure. The diaphragm having grooves showed a corrugate-like structure that was formed by residual stress. The high sensitivity of the membrane with the grooved diaphragm is considered to relate to the corrugate-like structure.


Subject(s)
Lead/chemistry , Membranes, Artificial , Microelectrodes , Titanium/chemistry , Transducers , Ultrasonography/instrumentation , Zirconium/chemistry , Electrochemistry/instrumentation , Electrochemistry/methods , Equipment Design , Equipment Failure Analysis , Image Enhancement/instrumentation , Image Enhancement/methods , Materials Testing , Reproducibility of Results , Sensitivity and Specificity , Ultrasonography/methods
SELECTION OF CITATIONS