Appl Radiat Isot ; 103: 25-30, 2015 Sep.
Article in English | MEDLINE | ID: mdl-26046520

ABSTRACT

To measure X-ray spectra with high count rates, we developed a detector consisting of a Lu2(SiO4)O [LSO] crystal with a decay time of 40 ns and a multipixel photon counter (MPPC). The photocurrents flowing through the MPPC are converted into voltages and amplified by a high-speed current-voltage amplifier, and event pulses from the amplifier are sent to a multichannel analyzer to measure spectra. We used three MPPCs of 100, 400 and 1600 pixels/mm(2), and the MPPCs were driven under pre-Geiger mode at a temperature of 20 °C. At a tube voltage of 100 kV and a tube current of 5.0 µA, the maximum count rate was 12.8 kilo-counts per second. The event-pulse widths were 200 ns, and the energy resolution was 53% at 59.5 keV using a 100-pixel MPPC.


Subject(s)
Amplifiers, Electronic , Lutetium/radiation effects , Photometry/instrumentation , Radiometry/instrumentation , Signal Processing, Computer-Assisted/instrumentation , X-Rays , Equipment Design , Equipment Failure Analysis , Photons , Radiation Dosage , Reproducibility of Results , Sensitivity and Specificity , Silicon Dioxide/radiation effects , Transducers