1.
Pain Med ; 16(2): 266-73, 2015 Feb.
Article in English | MEDLINE | ID: mdl-25393059

ABSTRACT

OBJECTIVE: This study is a pilot study to assess the clinical outcomes of percutaneous disc decompression using the L'DISQ in patients with lumbar discogenic pain. STUDY DESIGN: An institutional, prospective clinical data analysis. METHODS: We ablated the torn annulus using L'DISQ on 20 patients with axial low back pain for at least 3 months (average 29 months) unresponsive to conservative management. Before the therapeutic procedure, all the patients had been diagnosed with lumbar discogenic pain through provocation discography, which had confirmed the level of painful discs. The torn annulus was identified through lumbosacral magnetic resonance image and computed tomographic discogram. Baseline data were prospectively gathered before the procedure and at 1, 4, 12, 24, and 48 weeks post-procedure. Data included pain intensity (visual analog scale [VAS]), measure of disability (Oswestry Disability Index [ODI] and Rolando-Morris Disability Questionnaire [RM]), and health-related quality of life (Bodily Pain Scale of Short Form-36 version 2 [SF-36 BP]). RESULTS: At 48 weeks, the VAS fell from 7.55 ± 1.28 to 3.60 ± 2.28 scores, the ODI and RM had decreased significantly, and the SF-36 BP showed significant improvement (P < 0.05). The success rates of procedure were 55.0% at 48 weeks. There were no complications with the exception of a minor venous bleeding at the site of needle puncture. CONCLUSIONS: The L'DISQ device is specifically designed to ablate adjacent disc tissue using a wand that can be navigated into a torn annulus. Following ablation, we measured clinically significant pain improvement and decreased disability for patients with axial low back pain.


Subject(s)
Decompression, Surgical/instrumentation , Intervertebral Disc Displacement/surgery , Adult , Female , Humans , Lumbar Vertebrae , Male , Middle Aged , Pain Measurement , Pilot Projects , Young Adult
2.
Ann Rehabil Med ; 37(3): 433-7, 2013 Jun.
Article in English | MEDLINE | ID: mdl-23869344

ABSTRACT

A 22-year-old woman visited our clinic with a history of radiofrequency volumetric reduction for bilateral masseter muscles at a local medical clinic. Six days after the radiofrequency procedure, she noticed a facial asymmetry during smiling. Physical examination revealed immobility of the mouth drawing upward and laterally on the left. Routine nerve conduction studies and needle electromyography (EMG) in facial muscles did not suggest electrodiagnostic abnormalities. We assumed that the cause of facial asymmetry could be due to an injury of zygomaticus muscles, however, since defining the muscles through surface anatomy was difficult and it was not possible to identify the muscles with conventional electromyographic methods. Sono-guided needle EMG for zygomaticus muscle revealed spontaneous activities at rest and small amplitude motor unit potentials with reduced recruitment patterns on volition. Sono-guided needle EMG may be an optimal approach in focal facial nerve branch injury for the specific localization of the injury lesion.

3.
Pain Med ; 13(8): 1000-3, 2012 Aug.
Article in English | MEDLINE | ID: mdl-22882411

ABSTRACT

STUDY DESIGN: This is an in vitro experimental study of the technical capability and safety study of a navigable percutaneous disc decompression device named L'DISQ. OBJECTIVES: The objectives of this study were to determine if L'DISQ could adequately reach certain target zones in the disc and to measure the distribution of rises in temperature in the surrounding tissue when the device was used to ablate the disc. METHODS: Placement of the wand of L'DISQ was attempted into the posterior annulus of the discs of four fresh human cadavers. During disc ablation, thermocouple probes were used to measure the temperature within the nucleus pulposus and annulus fibrosus, on the surface of the annulus, and on the posterior longitudinal ligament. Tissues harvested from around the disc were examined histologically. RESULTS: The tip of the wand could be successfully navigated to the posterolateral or posterocentral annulus at all levels above L5-S1 using a lateral approach. Rises in temperature did not exceed 13.25 ± 0.84°C within the disc, and did not exceed 1°C on the surface of the disc. Histology demonstrated no thermal damage to the surrounding neural tissues. CONCLUSION: L'DISQ can be successfully navigated to the target zones, and disc tissue ablated without thermal or structural damage to the adjacent neural tissues.


Subject(s)
Decompression, Surgical/instrumentation , Diskectomy, Percutaneous/instrumentation , Intervertebral Disc Displacement/surgery , Intervertebral Disc/pathology , Intervertebral Disc/surgery , Adult , Cadaver , Decompression, Surgical/methods , Diskectomy, Percutaneous/methods , Hot Temperature , Humans , Intervertebral Disc Displacement/pathology , Models, Anatomic , Thermography/methods
4.
Pain Med ; 12(3): 370-6, 2011 Mar.
Article in English | MEDLINE | ID: mdl-21332936

ABSTRACT

STUDY DESIGN: An institutional, prospective clinical data analysis. OBJECTIVE: To evaluate the safety and efficacy of a new navigable percutaneous disc decompression device (L'DISQ) in patients with lumbar disc herniation with radicular pain. METHODS AND OUTCOME MEASURES: We performed disc decompressions using L'DISQ on 27 patients with persistent disabling back and leg pain for 1 month or longer (average 6.48 months) due to a herniated lumbar intervertebral disc. Baseline data were prospectively gathered before the index procedure and at 1, 4, 12, and 24 weeks post-procedure. Data included pain intensity (visual analog scale [VAS]), measure of disability (Oswestry Disability Index [ODI] and Rolando-Morris Questionnaire [RM]), health-related quality of life (Bodily Pain Scale of Short Form-36 version 2 [SF-36 BP]), and passive straight leg raising test (SLR). RESULTS: The VAS fell from 7.08±1.22 to 1.84±0.99 scores at 24 weeks post-procedure. At 24 weeks, the ODI had fallen from 41.88±10.61 to 16.66±8.55% and the RM from 11.52±3.91 to 2.68±1.97 points. The SF-36 BP dropped significant improvement from 32.89±5.83 to 49.57±4.96 scales. In the SLR test, the angular change of 24 weeks showed considerable improvement from 60.20±20.02 to 83.00±14.29 degrees. No major complication occurred, although two cases developed a disc reherniation 1 month post-procedure. CONCLUSIONS: The L'DISQ device is specifically designed to remove herniated disc using a wand that can be navigated into a disc protrusion or extrusion. Following decompression, we measured clinically significant pain improvement and decreased disability for patients with both radicular and axial pain caused by protruded and extruded discs.


Subject(s)
Decompression, Surgical/instrumentation , Intervertebral Disc Displacement/complications , Intervertebral Disc Displacement/surgery , Pain/etiology , Adolescent , Adult , Aged , Decompression, Surgical/methods , Female , Humans , Intervertebral Disc Displacement/pathology , Male , Middle Aged , Pain/surgery , Pain Measurement , Pilot Projects , Prospective Studies , Surveys and Questionnaires , Treatment Outcome , Young Adult