Igaku Butsuri ; 20(4): 151-158, 2000.
Article in English | MEDLINE | ID: mdl-12764240

ABSTRACT

Although low energy X-rays have been utilized for mammography, their safety in medical use is a matter of concern. Characteristics of the radiophotoluminescence glass dosimeter, GD-403, consisting of a glass element and filters, were investigated with respect to monochromatic X-rays obtained from a synchrotron radiation for personal monitoring of low energy photons. We focused on low energy X-rays ranging from 8 to 20 keV to study the linearity of the GD-403 response between photon fluence and dose equivalent. The GD-403 was placed on a tough water phantom and irradiated using an 11-15 mm x 0.1-7 mm beam for modulation of the photon fluence. The tough water phantom could be moved through a distance of 110-150 mm with a stepping motor. For the dose equivalent at 1cm depth (H1), 3mm (H3) and 70 &mgr;m (H70), the GD-403 showed sufficient linearities against the photon fluences in the energy regions of 8 to 20 keV, 13 to 20 keV and 13 to 20 keV, respectively. However, H3 and H70 did not provide sufficient linearities in the energy region of 8 to 12 keV. Moreover, we compared the result in this experiment with the value calculated from the absorbed dose of air using the mass absorption coefficient for the X-ray energy ranging from 10 to 20 keV.