Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Rev Sci Instrum ; 88(2): 025112, 2017 Feb.
Article in English | MEDLINE | ID: mdl-28249496

ABSTRACT

A design for a Quartz Crystal Microbalance (QCM) setup for use with viscous liquids at temperatures of up to 300 °C is reported. The system response for iron and gold coated QCM crystals to two common lubricant base oils, polyalphaolefin and halocarbon, is reported, yielding results that are consistent with theoretical predictions that incorporate electrode nanoscale surface roughness into their analysis.

SELECTION OF CITATIONS
SEARCH DETAIL
...